Impact of four kVp combinations available in a dual-source CT on the spectral performance of abdominal imaging: A task-based image quality assessment on phantom data.

Joël Greffier1, Julien Frandon1, Alexandre Sadate1, Philippe Akessoul1, Asmaa Belaouni1, Jean-Paul Beregi1, Djamel Dabli1.   

Abstract

PURPOSE: To compare the spectral performance of four combinations of kVp available in a third generation dual-source CT (DSCT) on abdominal imaging.
METHODS: An image-quality phantom was scanned with a DSCT using four kVp pairs (tube "A" voltage/tube "B" voltage): 100/Sn150 kVp, 90/Sn150 kVp, 80/Sn150 kVp, and 70/Sn150 kVp, classic parameters and dose level for abdomen examination (CTDIvol : 11 mGy). The noise power spectrum (NPS) and the task-based transfer function (TTF) of two inserts were computed on virtual monochromatic images (VMIs) at 40/50/60/70 keV and for mixed, low-, and high-kVp images. Detectability index (d') was computed on VMIs and mixed images to model the detection task of liver metastasis (LM) and hepatocellular carcinoma (HCC). Iodine quantification accuracy was assessed using the Root Mean Square Deviation (RMSDiodine ) and the iodine bias (IB).
RESULTS: Noise magnitude decreased by -55%± 0% between 40 and 70 keV for all kVp pairs. Compared to 70/Sn150 kVp, noise magnitude was increased by 9% ± 0% with 80/Sn150 kVp, by 16% ± 1% with 90/Sn150 kVp and by 24%± 1% with 100/Sn150 kVp. The average NPS spatial frequency (fav ) shifted toward higher frequencies as energy level increased for all kVp pairs. Lowest fav values were found for 70/Sn150 kVp and highest for 100/Sn150 kVp. The value of TTF at 50% (f50 ) shifted toward lower frequencies with increasing energy level. The highest f50  values occurred for 100/Sn150 kVp and the lowest for 80/Sn150 kVp. For both lesions, d' was highest for 70/Sn150 kVp and lowest for 100/Sn150 kVp. Compared to 70/Sn150 kVp, d' decreased by -6% ± 3% with 80/Sn150 kVp, by -11% ± 2% with 90/Sn150 kVp and by -13%± 2% with 100/Sn150 kVp. For all acquisitions, the RSMDiodine and IB were the lowest for 100/Sn150 kVp (0.29 ± 0.10 mg/ml and 0.88 ± 0.30 mg/ml, respectively) and increased when the tube "A" voltage decreased (2.34 ± 0.29 mg/ml for 70/Sn150 kVp and 7.42 ± 0.51 mg/ml respectively).
CONCLUSION: 70/Sn150 kVp presented the lowest image noise and highest detectability in VMIs of two small focal liver lesions. 100/Sn150 kVp presented the lowest image noise on mixed images and highest accuracy of iodine quantification in iodine images.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  dual-energy imaging; dual-source CT scan; spectral performance; task-based image quality assessment; virtual monoenergetic images

Year:  2021        PMID: 34312979     DOI: 10.1002/acm2.13369

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  2 in total

1.  Comparison of virtual monoenergetic imaging between a rapid kilovoltage switching dual-energy computed tomography with deep-learning and four dual-energy CTs with iterative reconstruction.

Authors:  Joël Greffier; Salim Si-Mohamed; Boris Guiu; Julien Frandon; Maeliss Loisy; Fabien de Oliveira; Philippe Douek; Jean-Paul Beregi; Djamel Dabli
Journal:  Quant Imaging Med Surg       Date:  2022-02

2.  Impact of an artificial intelligence deep-learning reconstruction algorithm for CT on image quality and potential dose reduction: A phantom study.

Authors:  Joël Greffier; Salim Si-Mohamed; Julien Frandon; Maeliss Loisy; Fabien de Oliveira; Jean Paul Beregi; Djamel Dabli
Journal:  Med Phys       Date:  2022-06-24       Impact factor: 4.506

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.