| Literature DB >> 34307038 |
Boudewijn Klop1, Rik Willems1,2, Joris Ector1,2, Peter Haemers1,2.
Abstract
Entities:
Keywords: Implantable cardioverter-defibrillator; PVARP; Pacemaker; Pacemaker-mediated tachycardia; Premature ventricular complex; Resynchronization; Retrograde conduction
Year: 2021 PMID: 34307038 PMCID: PMC8283539 DOI: 10.1016/j.hrcr.2021.04.007
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Device settings for bradytherapy
| Pacing mode | DDD |
|---|---|
| Lower rate | 50 |
| Upper rate | 105 |
| Atrial output (bipolar) | 3.5 V @ 0.4 ms |
| Ventricular output (bipolar) | 3.5 V @ 0.4 ms |
| Atrial sensitivity (bipolar) | 0.25 mV |
| Ventricular sensitivity (bipolar) | 0.6 mV |
| Paced AV delay | 210–300 ms |
| Sensed AV delay | 190–270 ms |
| PVAB | Smart |
| PVARP | 360–380 ms |
| PVARP after PVC | 400 ms |
| Ventricular blanking after atrial pace | 65 ms |
| VRP | 230–250 ms |
PVAB = postventricular atrial blanking; PVARP = postventricular atrial refractory period; PVC = premature ventricular complex; VRP = ventricular refractory period.
Figure 1A: The initial electrocardiogram during the outpatient visit showed sequential atrial-ventricular pacing and frequent premature ventricular beats. B: Suddenly an arrhythmia developed during the outpatient visit.
Figure 2A: The electrocardiogram shows an endless-loop, retrograde-dependent pacemaker-mediated rhythm of 80 beats/min with retrograde P waves (red arrows) with significantly prolonged retrograde ventriculoatrial conduction of 520 ms, which corresponds with a pacemaker-mediated tachycardia (PMT). B: The intracardiac electrogram shows the exact onset and etiology of the PMT. First the tracing shows sequential atrial-ventricular pacing. However, the sixth QRS complex is a premature ventricular beat (∗) with retrograde ventriculoatrial conduction of approximately 420 ms (blue arrow with solid line), which triggers the PMT. During the PMT the retrograde conduction exceeds the maximal PVARP of 500 ms (blue arrows with dotted dash line) and the rate is below the maximal tracking rate.