| Literature DB >> 34291407 |
Folker Spitzenberger1, Jaimin Patel2, Inga Gebuhr3, Klaus Kruttwig2,4, Abdulrahim Safi2, Christian Meisel3.
Abstract
PURPOSE: This study aimed at the development of a regulatory strategy for compliance of laboratory-developed tests (LDTs) with requirements of the Regulation (EU) 2017/746 ("EU-IVDR") under consideration of international requirements for LDTs as established in major regulatory regions. Furthermore, it was analysed in how far elements of current LDT regulation could qualify for an internationally harmonised concept ensuring quality, safety and performance of LDTs.Entities:
Keywords: Conformity assessment; IVDR; In-house IVD; LDT; Laboratory-developed tests; Validation
Mesh:
Substances:
Year: 2021 PMID: 34291407 PMCID: PMC8294224 DOI: 10.1007/s43441-021-00323-7
Source DB: PubMed Journal: Ther Innov Regul Sci ISSN: 2168-4790 Impact factor: 1.778
Figure 1The four major steps of a regulatory strategy for medical devices: (1) qualification and demarcation of the considered product on the basis of applicable definitions, (2) risk-based classification of the device, (3) identification and fulfilment of basic elements of conformity assessment (CA), and (4) choice and conduct of the applicable regulatory approval procedure.
The Definition of IVD According to the EU IVDR is Largely Convergent with the International Consensus as Originally Defined by GHTF and Adopted by IMDRF.
| Regulation (EU) 2017/746 (IVDR) | GHTF/SG1/N071:2012 (IMDRF/IVD WG/N64FINAL:2021) |
|---|---|
| ‘In Vitro diagnostic medical device’ means any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, piece of equipment, software or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information on one or more of the following: (a) concerning a physiological or pathological process or state; (b) concerning congenital physical or mental impairments; (c) concerning the predisposition to a medical condition or a disease; (d) to determine the safety and compatibility with potential recipients; (e) to predict treatment response or reactions; (f) to define or monitoring therapeutic measures. Specimen receptacles shall also be deemed to be in vitro diagnostic medical devices; | ‘In Vitro Diagnostic (IVD) medical device’ means a medical device, whether used alone or in combination, intended by the manufacturer for the in-vitro examination of specimens derived from the human body solely or principally to provide information for diagnostic, monitoring or compatibility purposes Note 1: IVD medical devices include reagents, calibrators, control materials, specimen receptacles, software, and related instruments or apparatus or other articles and are used, for example, for the following test purposes: diagnosis, aid to diagnosis, screening, monitoring, predisposition, prognosis, prediction, determination of physiological status Note 2: In some jurisdictions, certain IVD medical devices may be covered by other regulations |
Figure 2Although the IVDR does not include a definition of LDTs, there are four major scenarios for the qualification and demarcation of “devices manufactured and used only within health institutions” based on the laboratory’s intention that the product or product combination is intended to be used for in vitro diagnostic examination purposes.
Determination of the Intended Purpose of LDTs Based on the Major Criteria for the Definition of IVD.
Regulatory Demarcation of LDTs from Other Potential Health Products and as IVD Based on the Definition According to the EU IVDR.
| Question 1 | Is the device used in vitro for the examination of specimens derived from the human body? |
| Question 2 | Is the device used for providing one or more of the following information? a. concerning a physiological or pathological process or state b. concerning congenital physical or mental impairments c. concerning the predisposition to a medical condition or a disease d. to determine the safety and compatibility with potential recipients e. to predict treatment response or reactions f. to define or monitoring therapeutic measures |
| Question 3 | Is the device used as specimen receptacle? Note: Specimen receptacle is a device used for the primary containment and preservation of specimens derived from the human body for the purpose of in vitro diagnostic examination |
| Conditions: | Results |
| If both the question 1 and 2 are correct | The LDT is an IVD device |
| If either question 1 or question 2, and question 3 are not correct | The LDT is not an IVD device |
| If question 3 is correct | The LDT is an IVD device |
IVD Devices in Service at Labor Berlin per Department (per December 2020).
| Laboratory Department of Labor Berlin | IVD devices in service at Labor Berlin | |
|---|---|---|
| Commercially available IVDs | LDTs | |
| Toxicology | 39 | 209 |
| Hematology/Oncology | 15 | 168 |
| Human Genetics | 1 | 145 |
| Laboratory Medicine | 274 | 42 |
| Immunology | 20 | 39 |
| Virology | 42 | 31 |
| Endocrinology | 73 | 24 |
| Microbiology | 107 | 1 |
| Tumor Cytogenetics | 10 | 0 |
| Allergy Diagnostics | 3 | 0 |
| Autoimmune Diagnostics | 93 | 0 |
| Total | 677 | 659 |
Figure 3Search for an equivalent CE-marked IVD and justification for use according to Article 5 (5) of the EU IVDR.
Figure 4LDTs in service and potential alternatives from commercially available IVD (CE-marked) at Labor Berlin per department.
Figure 5Overview on the general obligations of manufacturers of CE-marked IVD according to Article 10 of the EU IVDR.
Overview on the Requirements Related to LDT Use According to Article 5 (5) of the EU IVDR, Proposed Implementation Methods and Comparison with the General Obligations of Commercial IVD Manufacturers According to Article 10 of the EU IVDR.
| No | Requirement for LDT use according to Article 5 (5) of the IVDR | Proposed implementation method for LDTs | General requirement for CE-marked IVD as per Article 10, IVDR |
|---|---|---|---|
| I | (a)the devices are not transferred to another legal entity | LDTs shall be manufactured and used only within a single health institution | Not applicable |
| II | (b) manufacture and use of the devices occur under appropriate quality management systems | QMS according to ISO 15189 and applicable requirements of ISO 13485 | QMS according to EN ISO 13485 and QM requirements according to Article 10, IVDR |
| III | (c) the laboratory of the health institution is compliant with standard EN ISO 15189 or where applicable national provisions, including national provisions regarding accreditation | Compliance with ISO 15189 (as shown by accreditation) or with national provisions, for example the RiLiBÄK (Germany) | Not applicable |
| IV | d) the health institution justifies in its documentation that the target patient group's specific needs cannot be met, or cannot be met at the appropriate level of performance by an equivalent device available on the market | QM documentation of the justification for use of LDTs. Compare Fig. | Not applicable |
| V | (e) the health institution provides information upon request on the use of such devices to its competent authority, which shall include a justification of their manufacturing, modification and use | QM documentation on the use of LDTs | Technical documentation according to Annex II and III, IVDR |
| VI | (f) the health institution draws up a declaration which it shall make publicly available, including: (i) the name and address of the manufacturing health institution, (ii) the details necessary to identify the devices, (iii) a declaration that the devices meet the general safety and performance requirements set out in Annex I to this Regulation and, where applicable, information on which requirements are not fully met with a reasoned justification therefore | Documentation and publication of a formal declaration of conformity with Annex I, IVDR | Draw up of an EU declaration of conformity in compliance with Article 17 and Annex IV, IVDR |
| VII | (g) as regards class D devices in accordance with the rules set out in Annex VIII, the health institution draws up documentation that makes it possible to have an understanding of the manufacturing facility, the manufacturing process, the design and performance data of the devices, including the intended purpose, and that is sufficiently detailed to enable the competent authority to ascertain that the general safety and performance requirements set out in Annex I to this Regulation are met. Member States may apply this provision also to class A, B or C devices in accordance with the rules set out in Annex VIII | Documentation of the design, performance evaluation data and manufacturing process within the laboratory’s QM system | Technical documentation according to Annex II, IVDR |
| VIII | (h) the health institution takes all necessary measures to ensure that all devices are manufactured in accordance with the documentation referred to in point (g) | Continuous documentation of the data related to manufactured LDTs | Technical documentation according to Annex II, IVDR |
| IX | (i) the health institution reviews experience gained from clinical use of the devices and takes all necessary corrective actions | Implementation of an LDT review system including CAPA procedures | Technical documentation according to Annex III, IVDR |
| X | Member States may require that such health institutions submit to the competent authority any further relevant information about such devices which have been manufactured and used on their territory. Member States shall retain the right to restrict the manufacture and use of any specific type of such devices and shall be permitted access to inspect the activities of the health institutions | Control of any documents and records of the QM system related to LDTs | Concepts and methods of market surveillance by competent authorities according to Articles 88ff. IVDR |
| XI | This paragraph shall not apply to devices that are manufactured on an industrial scale | Laboratories shall not offer their service in the frame of an industrial production chain and environment | Not applicable |
Cross-References Between Regulatory Obligations for Manufacture of LDTs According to the EU IVDR and Requirements According to ISO 15189.
| Applicable regulatory obligations according to the IVDR | EU IVDR references to the requirements | Compliance status | Reference to ISO 15189 | |
|---|---|---|---|---|
| Related article | Related annex | |||
| General safety and performance requirements | Article 5 (5) | Annex I | Partial | Sections 5.4, 5.5, 5.6 |
| Quality management system | Article 5 (5) (b) & (c) | Section 4.2 | ||
| Risk management system | Annex I (2, 3, 4 & 5) | Partial | Section 4.14.6 | |
| Performance evaluation | Annex I (9) | Partial | Section 5.5 | |
| Justification | Article 5 (5) (d) | No | ||
| Documentation for compliance | Article 5 (5) (g) | Sections 4.2.2 and 4.13 | ||
| Publicly available declaration | Article 5 (5) (f) | No | ||
| Surveillance system (review of experience) | Article 5 (5) (i) | Sections 4.8, 4.9 and 4.14, 5.3 | ||
| Surveillance system (corrective actions) | Article 5 (5) (i) | Sections 4.10 and 4.11, 5.3 | ||
| Transfer of LDTs | Article 5 (5) (a) | No | ||
| Scale of production | Article 5 (5) | No | ||
Structure of the GSPR Checklist to be Filled for Evidence of Compliance with Annex I of the EU IVDR.
| No | General safety and performance requirement according to Annex I, IVDR | A/NA | Method used to demonstrate conformity | Method reference | Reference to the identity of controlled documents within the QMS | Justification for applicability/non-applicability |
|---|---|---|---|---|---|---|
| 1 | Devices shall achieve the performance intended by their manufacturer and shall be designed and manufactured in such a way that, during normal conditions of use, they are suitable for their intended purpose. They shall be safe and effective and shall not compromise the clinical condition or the safety of patients, or the safety and health of users or, where applicable, other persons, provided that any risks which may be associated with their use constitute acceptable risks when weighed against the benefits to the patient and are compatible with a high level of protection of health and safety, taking into account the generally acknowledged state of the art | A | Compliance with international standards | ISO 15189 (whole standard); ISO 22367, 5.4, 5.5, Annex A.4, Annex D; ISO 13485, 7.3 | Quality manual, documented procedures for the design and development process for LDTs including validation and verification procedures | Requirement is fully applicable for LDTs, because performance and safety are essential throughout the lifecycle of each LDT |
| … | … | … | … | … | … | … |
Only few IVDR requirements can be fully covered (italic) by compliance with ISO 15189
The checklist is applied to the General Requirement No. 1 of Annex I of the EU IVDR. A method used to demonstrate conformity could be, for example, the application of an international standard, specifications etc. The method reference is, for example, the title of a standard. The reference to the QMS could be the identifier of an SOP, for example
A applicable requirement, NA not applicable
Phases of Performance Evaluation of LDTs as
Adapted from Annex XIII of the EU IVDR.
| Phase of the performance evaluation | Title of the phase | Performance evaluation activities |
|---|---|---|
| Phase I | Planning of performance evaluation | Determination of the scope of the performance evaluation; description of the device under evaluation; assignment of responsibilities for the performance evaluation; selection of relevant performance criteria to generate the necessary clinical evidence for the LDT |
| Phase II | Execution of performance evaluation | Identification of the scientific validity and relevant analytical and clinical performance data |
| Generation of data through performance studies, if applicable | ||
| Appraisal of relevant performance data | ||
| Assessment of relevant performance data | ||
| Phase III | Documentation of performance evaluation | Documentation of the results of performance evaluation |
| Phase IV | Update of performance evaluation | Post-market performance follow-up, if applicable |
Key Characteristics as Central Elements for LDT Regulation in Different Advanced Regulatory Markets.
| No | Element | EU IVDR | U.S | Australia | Canada |
|---|---|---|---|---|---|
| 1 | Quality management system | Yes (Section 4) | |||
| 2 | Risk-based approach for classification/categorization | No | |||
| 3 | Risk management system | Yes (Section 5.4) | |||
| 4 | Evaluation and documentation related to essential requirements for quality, safety, performance | Yes (Sections 5, 6, 7) | |||
| 5 | Product monitoring and surveillance | Yes (Section 7.4) | |||
| 6 | Register | No | No | Not appplicable | |
| 7 | Justification for use | No | No | Not appplicable | |
| 8 | Notification requirement | No | Not appplicable | ||
| 9 | Regulatory oversight mechanism | Not appplicable |
For Canada, only the sections of the standard Z316.8-18 are considered. Elements that involve interactions with a regulatory body (e. g. registration, notification etc.) are therefore deemed to be not applicable for Canada. The boxes italic indicate common elements