Literature DB >> 34277107

Quantification of image quality of intra-fractional cone-beam computed tomography for arc irradiation with various imaging condition.

Masahide Saito1, Yuki Shibata1, Koji Ueda1, Hidekazu Suzuki1, Takafumi Komiyama1, Kan Marino1, Shinichi Aoki1, Yoshiyasu Maehata1, Naoki Sano1, Hiroshi Onishi1.   

Abstract

BACKGROUND: 3-dimensional intra-cone beam computed tomography (intra-CBCT ) could be a potentially powerful tool for use with arc irradiation such as volumetric modulated arc therapy. The aim of the study was to evaluate the image quality of intra-cone beam computed tomography (intra-CBCT ) for arc irradiation with various imaging condition.
MATERIALS AND METHODS: Two types of intra-CBCT imaging techniques were evaluated - intra-fractional CBCT with flattening filtered (FF) beam (intra-FF CBCT ) and that with flattening filter free (FFF) beam (intra-FFF CBCT ). For the intra-MV beams, four different field sizes (2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, and 20 cm × 20 cm) were used with dose rates of 500 MU/min and 1600 MU/min, for 6 MV FF and 6 MV FFF, respectively. For all image acquisitions, two rotation angles (full-arc and half-arc) were investigated. Thereafter, the linearity, contrast-to-noise ratio (CNR), and uniformity index (UI) of intra-CBCT image were compared with those of conventional CBCT image.
RESULTS: All acquisition conditions had good linearity of the CT value (R2 > 0.99). For CNR, the change rates from conventional CBCT ranged from 0.6-33.7% for a 2 cm × 2 cm beam, whereas that for a 20 cm × 20 cm beam ranged from 62.7-82.3%. Similarly, the UI increased from 1.5% to 7.0% as the field size increased.
CONCLUSION: Quality of intra-CBCT image was affected by the field size and acquisition angle. Image quality of intra-CBCT was worse than that of conventional CBCT, but it was better under a smaller field and wider correction angle and would be acceptable for clinical use.
© 2021 Greater Poland Cancer Centre.

Entities:  

Keywords:  MV simultaneous exposure; cone-beam computed tomography; image quality; volumetric modulated arc therapy

Year:  2021        PMID: 34277107      PMCID: PMC8281903          DOI: 10.5603/RPOR.a2021.0066

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


Introduction

In high conformal radiotherapy for a tumor with respiratory motion, monitoring the intrafractional tumor motion is important to confirm the accuracy of irradiation and reduce internal margin. Institutes globally have used multiple techniques to monitor tumor motion during irradiation. One of the techniques involves the use of an electronic portal imaging device (EPID), which enables to monitor the target localization using the treatment beam on cine mode EPID image [1]. Another technique is a 4-dimensional (4-D) kV-cone-beam computed tomography (CBCT). This technique enables to verify the tumor position and other anatomical positions during treatment to generate the respiratory correlated CBCT by projection phase sorting [2]. Some studies have investigated the effectiveness and improvement of 4D-kV CBCT to assess the respiratory motion of tumor during high conformal treatments, such as volumetric modulated arc therapy (VMAT) [3-6]. In this study, we investigated the effectiveness of intra-fractional CBCT, called intra-CBCT, a potentially powerful tool for use with arc irradiation such as volumetric modulated arc therapy. Although the quality of intra-CBCT images is worse due to the ray scattering derived from MV beam irradiation and the acquisition of them gives an extra dose, no study has evaluated the details of image quality during arc irradiation.

Materials and methods

Equipment for intra-CBCT

X-ray volume imaging (XVI®) system ver. 5.0 (Elekta AB, Crawley, UK) attached to the Elekta Synergy was used in this study. It utilizes an amorphous silicon/cesium iodide flat-panel detector with a 1024 × 1024 array of 0.4-mm elements. The X-ray source is mounted on a retractable arm with a fixed source to an isocenter distance of 100 cm and offers several different interchangeable collimators and kV filter combinations based on the desired scanning volume. For acquiring intra-CBCT images, the 3D Volume ViewTM mode was used, which are reconstructed automatically after the acquisition of projection images. For reconstruction of intra-CBCT images, beam delivery has to be accompanied by a rotational treatment, such as conformal arc radiotherapy and VMAT. In this study, three types of CBCT images were compared, namely: 1) conventional CBCT image, 2) intra-CBCT image acquired using a 6 MV flattening filter (FF) beam (intra-FF CBCT), and 3) intra-CBCT image acquired using a 6 MV with flattening filter free (FFF) beam (intra-FFF CBCT). Two collection angles (full-arc: CW, −180° to 180° and half-arc: CW, −180°to 20°) were investigated for image acquisition. All images were acquired with a 120-kV, 20-mA with the following imaging parameters: S10 collimator (27 cm of axial field-of-view (FOV), 12 cm of nominal axial length); image size, 540 × 540 pixels; pixel size, 0.5 mm; and slice thickness, 0.5 mm. For the intra-MV beams, the monitor unit (MU) was set as 1000, and four different field sizes (2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, and 20 cm × 20 cm) were used with dose rates of 500 MU/min and 1600 MU/min, for 6 MV FF and 6 MV FFF, respectively.

Evaluation of image quality

The Catphan® 503 (The Phantom Laboratory, Salem, NY, US) phantom with a 20-cm diameter was used for the evaluation of intra-CBCT image quality. It has the following different inserts for evaluating the geometric accuracy with defined distances in the axial and longitudinal directions: 1) a section with uniform material and a density equivalent to that of water for determining spatial uniformity; 2) a section featuring a cylindrically arranged resolution pattern in an axial plane (1 through 21 line pairs per centimeter) and two point markers (tungsten-carbide bead; diameter, 0.28 mm) embedded into a uniform material; and 3) a section with inserts of different densities such as polystyrene, polymethylpentene (PMP), low-density polyethylene (LDPE), acrylic, air, Delrin, and Teflon implemented for conversion testing of electron density to the CT number. The phantom was set in the treatment couch top and the laser was aligned as shown in Figure 1A.
Figure 1

Setup of the Catphan®503 phantom in the head of treatment couch (A), and the region of interest (ROI) for each image analysis (B, C); B — the image slice used for evaluating the linearity and contrast-to-noise ratio (CNR), which has nine ROIs with each material; C — the slice for the evaluation of the uniformity index (UI) which has five homogeneity ROIs

Image quality evaluation was performed using the following three indexes: 1) linearity between nominal and measured CT values; 2) contrast-to-noise ratio (CNR); and 3) uniformity index (UI). For the linearity, the average CT value was measured with nine regions of interest (ROIs), where each material existed (two materials had water). For the CNR, the value was calculated by following equation (1): where, S is the mean pixel value of each material; S is the mean pixel value of the background (water), and σ is the standard deviation of the background (water). Figure 1B shows the image slice used for evaluation of the linearity and CNR. To assess the UI, the average CT value was measured with five ROIs at regions with uniformity. Then, the UI was calculated with the following equation (2): where, is the mean HU that is the largest value among the peripheral four ROIs; is the mean HU of the center of the ROI. Figure 1C shows the image slice used for the evaluation of UI. All analyses were performed using MATLAB R2016a (Mathworks Inc., Natick, MA).

Results

Figure 2 shows the comparison among three CBCT images (conventional, intra-FF, and intra-FFF CBCT images with a field size of 10 cm × 10 cm). The slice which has a 1 through 21-line pair per centimeter high-resolution test gauge is shown for each image. On comparison between conventional and intra-CBCT images, the latter contained more noise than the former because of MV beam scattering during the imaging of the phantom. Especially, intra-FFF CBCT contained more noise than intra-FF CBCT. Furthermore, on focusing on the collection angles, full-arc showed better image recognition than those acquired with half-arc. Figure 3 also shows the lateral profiles of each CBCT image with the same condition of Figure 2, indicating that intra-CBCT images contain noise from MV beam scattering. Figure 4 shows the comparison among four field sizes (intra-CBCT with the field size of 2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, and 20 cm × 20 cm) for two collection angles. The image quality was worse as the field size increased. Thereafter, all CBCT images were quantitatively analyzed. Table 1 shows the linearity between the nominal and measured CT values. Good linearity was observed for both conventional CBCT and intra-CBCT (FF and FFF) images (R2 > 0.99 for all acquisition conditions). Figure 5 shows the change in CNR values depending on each field size for seven materials. CNR values decreased as the field size increased for all conditions. Table 2 shows the average value of the CNR change rate (%) for all materials assessed using conventional CBCT images for each condition. The change rates from conventional CBCT ranged from 0.6–33.7% for a 2 cm × 2 cm beam, whereas that ranged from 62.7–82.3% for a 20 cm × 20 cm beam. With regard to the collection angles, full-arc showed better CNR than the half-arc. Similarly, for beam quality, intra-FF CBCT images showed better CNR than intra-FFF CBCT images. Figure 6 shows the changes in UI depending on each field size for the intra-FF and intra-FFF CBCT images with two collection angles. For all conditions, the UI increased as the field sizes increased (approximately, from −1.5 to 7.0).
Figure 2

Comparison of three CBCT images [(A) and (D), conventional CBCT images; (B) and (E), intra-FF CBCT images; and (C) and (F), intra-FFF CBCT images] at the slice which has a 1 through 21-line pair per centimeter high-resolution test gauge. Two collection angles (upper, full-arc; bottom, half-arc) are shown. The image size is 540 × 540 pixels, and the pixel size was 0.5 mm for all dimensions. Window width of 1,000 HU and window level of 10 HU are shown

Figure 3

Lateral profiles of three CBCT images (black line, conventional CBCT images; red line, intra-FF CBCT images; and blue line, intra-FFF CBCT images). Two collection angles (upper, full-arc; bottom, half-arc) are shown. The selected profile line is shown in right figure (red line)

Figure 4

Comparison among four field sizes [(A) and (E), 2 cm × 2 cm; (B) and (F), 5 cm × 5 cm; (C) and (G), 10 cm × 10 cm; and (D) and (H), 20 cm × 20 cm] at the slice which has a 1 through 21-line pair per centimeter high-resolution test gauge. The intra FF-CBCT images with two collection angles (upper, full-arc; bottom, half-arc) are shown. The image size is 540 × 540 pixels, and the pixel size was 0.5 mm for all dimensions. Window width of 1,000 HU and window level of 10 HU are shown

Table 1

Linearity (R2) between the nominal and measured CT values

Linearity (R2)
CBCT2 cm × 2 cm5 cm × 5 cm10 cm × 10 cm20 cm × 20 cm
FFFull-arc0.99990.99990.99990.99980.9996
Half-arc0.99950.99960.99970.99930.9996
FFFFull-arc0.99990.99980.99990.99990.9991
Half-arc0.99950.99940.99900.99950.9968
Figure 5

Changes in contrast-to-noise ratio depending on the field size (2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, 20 cm × 20 cm) for seven materials. (A) and (C) show the results of intra-FF CBCT images and (B) and (D) show the results of intra-FFF-CBCT images. Two collection angles (upper, full-arc; bottom, half-arc) are shown

Table 2

The average value of the change rate (%) of contrast-to-noise ratio for all materials as evaluated using conventional CBCT images for each condition

CNR change rate (%)
2 cm × 2 cm5 cm × 5 cm10 cm × 10 cm20 cm × 20 cm
Full-arcFF−23.9 ± 20.7−36.9 ± 32.8−62.1 ± 14.2−72.3 ± 20.5
FFF−27.1 ± 44.8−38.6 ± 32.3−66.3 ± 18.0−74.6 ± 26.7
Half-arcFF−0.6 ± 39.5−11.8 ± 25.7−37.1 ± 16.4−62.7 ± 10.3
FFF−33.7 ± 18.6−48.3 ± 39.4−73.2 ± 19.1−82.3 ± 17.0
Figure 6

Uniformity index changes depending on the field size (2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, 20 cm × 20 cm). The intra-FF CBCT (black line) and intra-FF CBCT (gray line) for two collection angles [full-arc (solid line) and half-arc (dash line)] are shown

Discussion

Quality assurance of the kV-CBCT system is one of the important topics in radiotherapy and new research, such as image quality of dual energy CT, has been carried out [7]. However, in our knowledge, there has been no reports to investigate the image quality of intra-CBCT with various imaging conditions in detail. In our results, image quality depended on the field size and acquisition angle. With regard to field size, the scattered dose from the subject increased when the field size was large, which resulted in deteriorating image quality. With regard to the acquisition angle, full-arc imaging provided better image quality than half-arc because the former usually generates more projection images during treatment. Therefore, our study suggested that treatment with small fields, such as stereotactic body radiotherapy, might be preferred to use intra-CBCT. However, this study did not perform the evaluation with actual patient images and, therefore, additional study is required. Our study had some limitations. First, we investigated the image quality by only one vender (Elekta XVI). Our results might differ from those on other vendor machines. Second, our study only investigated images acquired with the following condition: 120 kV and 20 mA with S collimator. Therefore, different future studies should investigate image acquisition with different irradiation conditions. Third, in this study, we only evaluated the following two beam qualities: 6 MV FF and 6 MV FFF with the same MU. From our results, the quality of images acquired with FF seems better than acquired with FFF. However, the dose rate and gantry speed between them differed, indicating that they cannot be compared with each other. Therefore, further studies are needed to compare different beam qualities.

Conclusion

In this study, we evaluated intra CBCT at the various imaging conditions using the Catphan® 503 phantom. Quality of intra-CBCT image was affected by the field size and acquisition angle. Image quality of intra-CBCT was worse than that of conventional CBCT, but it was better under a smaller field and wider correction angle and would be acceptable for clinical use.
  7 in total

1.  In-treatment 4D cone-beam CT with image-based respiratory phase recognition.

Authors:  Satoshi Kida; Yoshitaka Masutani; Hideomi Yamashita; Toshikazu Imae; Taeko Matsuura; Naoya Saotome; Kuni Ohtomo; Keiichi Nakagawa; Akihiro Haga
Journal:  Radiol Phys Technol       Date:  2012-02-25

2.  Respiratory correlated cone beam CT.

Authors:  Jan-Jakob Sonke; Lambert Zijp; Peter Remeijer; Marcel van Herk
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

3.  Verification of in-treatment tumor position using kilovoltage cone-beam computed tomography: a preliminary study.

Authors:  Keiichi Nakagawa; Hideomi Yamashita; Kenshiro Shiraishi; Hiroshi Igaki; Atsuro Terahara; Naoki Nakamura; Kuni Ohtomo; Shigeki Saegusa; Takashi Shiraki; Takashi Oritate; Kiyoshi Yoda
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-15       Impact factor: 7.038

4.  First clinical cone-beam CT imaging during volumetric modulated arc therapy.

Authors:  Keiichi Nakagawa; Akihiro Haga; Kenshiro Shiraishi; Hideomi Yamashita; Hiroshi Igaki; Atsuro Terahara; Kuni Ohtomo; Shigeki Saegusa; Takashi Shiraki; Takashi Oritate; Kiyoshi Yoda
Journal:  Radiother Oncol       Date:  2008-12-04       Impact factor: 6.280

5.  Acquisition of MV-scatter-free kilovoltage CBCT images during RapidArc™ or VMAT.

Authors:  Clifton Ling; Pengpeng Zhang; Thanos Etmektzoglou; Josh Star-Lack; Mingshan Sun; Edward Shapiro; Margie Hunt
Journal:  Radiother Oncol       Date:  2011-08-05       Impact factor: 6.280

6.  Quality evaluation of monoenergetic images generated by dual-energy computed tomography for radiotherapy: A phantom study.

Authors:  Bartosz Pawałowski; Hubert Szweda; Alina Dudkowiak; Tomasz Piotrowski
Journal:  Phys Med       Date:  2019-05-29       Impact factor: 2.685

7.  Clinical feasibility of using an EPID in CINE mode for image-guided verification of stereotactic body radiotherapy.

Authors:  Ross I Berbeco; Fred Hacker; Dan Ionascu; Harvey J Mamon
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-01       Impact factor: 7.038

  7 in total