Literature DB >> 34276894

Patent Ductus Arteriosus Device Closure in Interrupted Inferior Vena Cava: Challenges Overcome and Lessons Learnt: A Case Series.

Palanivel Rajan1, Parag Barwad1, Pankaj Aggarwal2, Manoj Kumar Rohit1.   

Abstract

Interrupted inferior vena cava (IVC) with azygous continuation provides technical challenge in many percutaneous cardiac interventions. They are performed via the femoral venous access route. We describe four such cases in whom, patent ductus arteriosus (PDA) device closure was done in interrupted IVC. All the four cases were done using the femoral route. Two cases were done via retrograde approach through femoral artery access, in which one was closed with a muscular VSD device and the other with a Gianturco coil. The introduction of newer closure devices helps in the retrograde approach. Two cases were done via anterograde approach through the femoral venous route, with a loop through the azygous vein. One child had transient bradycardia due to cardiac stretch which normalized after the device deployment and introducer system removal. Thus, we describe different femoral approaches in interrupted IVC patients for PDA closure, with the difficulties faced and the complications managed. Copyright:
© 2021 Heart Views.

Entities:  

Keywords:  Device embolization; interrupted inferior vena cava; patent ductus arteriosus device; retrograde approach

Year:  2021        PMID: 34276894      PMCID: PMC8254156          DOI: 10.4103/HEARTVIEWS.HEARTVIEWS_180_20

Source DB:  PubMed          Journal:  Heart Views        ISSN: 1995-705X


INTRODUCTION

The percutaneous closure of Patent Ductus Arteriosus (PDA) has become the treatment of choice in all cases wherever it is feasible. It is a fairly simple procedure in patients with normal venous anatomy. However, PDA device closure is challenging in case of interrupted Inferior Vena Cava (IVC) and azygous continuation of IVC where multiple technical challenges can be encountered and need experienced operator for successful completion of the procedure.

CASE PRESENTATIONS

The authors describe challenges encountered during PDA device closure in four cases [Table 1] of azygous continuation of IVC, two of which were closed via azygous route, and the rest were done via arterial route.
Table 1

Details of the patients along with the procedure

NumberAge (years)SexWeight (kg)PDA size (mm)RouteApproachDevice
Case 13Male144Femoral veinAnterograde8 mm × 6 mm lifetech PDA device
Case 22Female94Femoral arteryRetrograde6 mm cocoon muscular VSD device
Case 33Male113Femoral veinAnterograde8 mm × 6 mm ADO I
Case 46Female202Femoral arteryRetrograde5 mm × 4 mm Gianturco coil

ADO: Amplatzer duct occlude, PDA: Patent ductus arteriosus, VSD: Ventricular septal defect

Details of the patients along with the procedure ADO: Amplatzer duct occlude, PDA: Patent ductus arteriosus, VSD: Ventricular septal defect

Case 1

A 3-year-old male child weighing 14 kg with 4 mm PDA [Figure 1a] was taken up for device closure. 5Fr Femoral Arterial and Venous Access were taken. Venous angiogram via the femoral route showed the azygous continuation of IVC [Figure 1b]. The femoral sheath was exchanged with FlexorR Balkin guiding sheath (Cook Medical, USA) which was parked in the mid-right atrium. A 5Fr Multipurpose catheter (Cook Medical, USA) with 0.035” straight tip Terumo GlidewireR (Terumo, Japan) was used to cross PDA. The wire was exchanged with AmplatzerR extra-stiff 0.035” wire (AGA Medical Corporation, USA) which was parked in descending aorta [Figure 1c]. This caused transient bradycardia and hypotension due to the stretch, which resolved on withdrawal of Amplatzer wire. An 8 mm × 6 mm Cera™ duct occluder device (Lifetech Scientific Inc., China) was deployed successfully [Figure 1d] over a 6-Fr SteerEase™ sheath (Lifetech Scientific Inc., China). The post deployment angiogram showed no residual flow across the device.
Figure 1

Case 1: (a) Aortogram showing 4 mm patent ductus arteriosus; (b) venogram showing interrupted inferior vena cava; (c) patent ductus arteriosus crossed via azygous route from the femoral venous access; (d) lifetech 8 mm × 6 mm patent ductus arteriosus device deployed

Case 1: (a) Aortogram showing 4 mm patent ductus arteriosus; (b) venogram showing interrupted inferior vena cava; (c) patent ductus arteriosus crossed via azygous route from the femoral venous access; (d) lifetech 8 mm × 6 mm patent ductus arteriosus device deployed

Case 2

A 2-year-old female child weighing 11 kg with 4 mm PDA [Figure 2a] was taken up for device closure. The child was found to have the azygous continuation of IVC during the procedure. PDA was crossed via the femoral venous route and attempt was made with 8 mm × 6 mm Amplatzer Duct Occluder (ADO) I (AGA Medical Corporation, USA) device [Figure 2b]. As soon as the cable was released, the device embolised into the right pulmonary artery (PA) probably because of the rebound force of the extra loop-tension of the cable device [Figure 2c]. The device was snared and retrieved from the right PA [Figure 2d]. It could be pulled into the superior vena cava (SVC) but could not be retrieved into the azygous vein [Figure 2e]. Hence a second snare was introduced from the internal jugular vein (IJV) access and the device was fed to the second snare from the first snare in the SVC [Figure 2e and f]. The device was pulled back into IJV [Figure 2g] and retrieved by surgical cut down on the table. In the same sitting, a 6 mm Cocoon muscular VSD device (Vascular Innovations, Thailand) was used successfully to close from the aortic end via the femoral route [Figure 2h].
Figure 2

Case 2: (a) Aortogram showing 4 mm patent ductus arteriosus; (b) Amplatzer Duct Occluder I device 8 mm × 6 mm via femoral venous route; (c) device (White arrow) embolized into right pulmonary artery; (d) embolized device snared; (e) Snared device stuck at superior vena cava-azygous junction; (f) device retrieved via 2nd snare via internal jugular vein access; (g) device in internal jugular vein; (h) cocoon 6 mm muscular device closure of patent ductus arteriosus via femoral arterial route

Case 2: (a) Aortogram showing 4 mm patent ductus arteriosus; (b) Amplatzer Duct Occluder I device 8 mm × 6 mm via femoral venous route; (c) device (White arrow) embolized into right pulmonary artery; (d) embolized device snared; (e) Snared device stuck at superior vena cava-azygous junction; (f) device retrieved via 2nd snare via internal jugular vein access; (g) device in internal jugular vein; (h) cocoon 6 mm muscular device closure of patent ductus arteriosus via femoral arterial route

Case 3

A 2-year-old girl child weighing 11 kg with 3 mm PDA with the azygous continuation of IVC was successfully closed with ADO I device from the femoral venous route without much difficulty [Figure 3a].
Figure 3

(a) Case 3: 3 mm patent ductus arteriosus closed with 8 mm × 6 mm Amplatzer Duct Occluder I device via femoral venous route; (b) Case 4: 2 mm patent ductus arteriosus closed with 5 mm × 4 mm Gianturco coil via femoral arterial route

(a) Case 3: 3 mm patent ductus arteriosus closed with 8 mm × 6 mm Amplatzer Duct Occluder I device via femoral venous route; (b) Case 4: 2 mm patent ductus arteriosus closed with 5 mm × 4 mm Gianturco coil via femoral arterial route

Case 4

A 6-year-old girl child with tiny PDA and azygous continuation of IVC was successfully closed with 5 mm × 4 mm Gianturco coil (Cook Medical, USA) from the arterial side [Figure 3b].

DISCUSSION

Patent Ductus Arteriosus (PDA) is one of the common congenital heart diseases accounting for 5%–10% of congenital heart defects.[12] Porstmann[3] was the first to close PDA by the percutaneous route in 1966. PDA devices have undergone a sea of change since then, with the basic technique being the same. Interruption of IVC is rare with an incidence of 0.1%.[4] It is more commonly associated with left isomerism and complex congenital heart defects while relatively rare with other conditions. However, if it is associated, percutaneous access to the cardiac chambers for device closure via the femoral venous route is difficult. The options for percutaneous closure of PDA in interrupted cases are Difficulty in tracking the femoral venous sheath, kinking of catheters and delivery sheath, hemodynamic instability due to stretch caused by the system in the heart are the problems faced while using this access. Snaring the wire from the arterial end and forming an arterio-venous loop might be necessary when difficulty is encountered in advancing the system. With the newer devices like ADO II and muscular VSD device, the closure has been made possible from the arterial end. Alternatively, coils can be deployed in case of small PDA. The need for technical expertise and the availability of devices at the procedure time are essential. Larger bore access at the arterial end in a younger child can be a deterrent in this approach. IJV access is used for the closure. Smaller children might need intubation and mechanical ventilation for undisturbed access. The technical difficulty is facing 2 U-turns in the route: First at the SVC-Right Atrium-Right ventricle-PA track and second at PA-PDA-aorta track. Various case reports [Table 2] have been made in this particular aspect with different technical challenges faced. In the first case, difficulty in tracking was overcome by using a long FlexorR Balkin sheath. Transient hemodynamic instability was faced due to extra-stiff wire in both case 1 and case 2 (during the initial attempt via femoral venous route). Caution is advised due to the rebound torque of the delivery system while unscrewing the device, due to which we faced a device embolization. We did not face the necessity to form an arterio-venous loop in any of our cases for tracking. In selective cases, device or coil deployment through arterial route can be undertaken (case 2 and case 4). The newer devices like ADO II are particularly useful in this regard.
Table 2

Cases compilation of patent ductus arteriosus device closure with interrupted inferior vena cava from literature

AuthorYearAge (years)SexWeight (kg)PDA size (mm)RouteApproachDevice
Al-Hamash[5]20069F244Femoral veinAnterogradeADO I, 8 mm × 6 mm
3M154Femoral VeinAnterogradeADO I, 8 mm × 6 mm
Sivakumar and Francis[6]2007N/AN/AN/A8Femoral ArteryRetrogradeADO I, 12 mm × 10 mm (reversed)
Koh et al.[7]200937FN/A3.5Femoral ArteryRetrogradeADO II, 6 mm × 6 mm
Patel et al.[8]20099F215.1IJVAnterogradeADO I, 12 mm × 10 mm
Akhtar et al.[9]201014FN/A5IJVAnterogradeADO I, 12 mm × 10mm
Celebi et al.[10]2013¼F5.44.3Femoral ArteryRetrogradeLifetech, 6 mm × 8mm
Tefera and Bermudez-Cañete[11]201412F395Femoral VeinAnterogradeADO, 8 mm × 6 mm
Subramanian et al.[12]20141 1/2N/AN/A4.7IJVAnterogradeLifetech PDA Occluder, 10 mm × 12 mm
Aggarwal et al.[13]20153F83IJVAnterograde6 mm × 4 mm device
Jawahirani et al.[14]20185F86IJVAnterogradeCocoon duct occluder 10 mm × 8 mm

ADO: Amplatzer duct occluder, IJV: Internal jugular vein, N/A: Not available, PDA: Patent ductus arteriosus

Cases compilation of patent ductus arteriosus device closure with interrupted inferior vena cava from literature ADO: Amplatzer duct occluder, IJV: Internal jugular vein, N/A: Not available, PDA: Patent ductus arteriosus

CONCLUSION

PDA device closure in interrupted IVC via the femoral route is challenging as extra loop and tension of wire can cause transient hypotension or rebound device embolization during release. One can use coils, muscular devices, or newer devices like ADO II from the aortic end in appropriate cases or use an IJV route.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.14

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  13 in total

1.  Transcatheter closure of a patent ductus arteriosus in a patient with an anomalous inferior vena cava.

Authors:  S Akhtar; A Akhtar; S M Samad; M Atiq
Journal:  Pediatr Cardiol       Date:  2010-07-06       Impact factor: 1.655

2.  Interruption or congenital stenosis of the inferior vena cava: prevalence, imaging, and clinical findings.

Authors:  Zafer Koc; Levent Oguzkurt
Journal:  Eur J Radiol       Date:  2006-12-11       Impact factor: 3.528

3.  Transcatheter closure of patent ductus arteriosus and interruption of inferior vena cava with azygous continuation using an Amplatzer duct occluder.

Authors:  Sadiq Al-Hamash
Journal:  Pediatr Cardiol       Date:  2006-08-23       Impact factor: 1.655

4.  Anomalous inferior vena cava drainage to portal vein offers a challenge to transcatheter ductus arteriosus closure.

Authors:  K Sivakumar; E Francis
Journal:  Pediatr Cardiol       Date:  2007-08-02       Impact factor: 1.655

5.  Catheter closure of patent ductus arteriosus. 62 cases treated without thoracotomy.

Authors:  W Porstmann; L Wierny; H Warnke; G Gerstberger; P A Romaniuk
Journal:  Radiol Clin North Am       Date:  1971-08       Impact factor: 2.303

6.  Congenital heart disease in 56,109 births. Incidence and natural history.

Authors:  S C Mitchell; S B Korones; H W Berendes
Journal:  Circulation       Date:  1971-03       Impact factor: 29.690

7.  Transcatheter closure of patent ductus arteriosus and interruption of inferior vena cava with azygous continuation using an Amplatzer duct occluder II.

Authors:  Ghee Tiong Koh; Sharifah Ai Mokthar; Amir Hamzah; Jasvinder Kaur
Journal:  Ann Pediatr Cardiol       Date:  2009-07

8.  Congenital heart disease among 160 480 liveborn children in Liverpool 1960 to 1969. Implications for surgical treatment.

Authors:  D F Dickinson; R Arnold; J L Wilkinson
Journal:  Br Heart J       Date:  1981-07

9.  Percutaneous closure of patent ductus arteriosus via internal jugular vein in patient with interrupted inferior vena cava.

Authors:  Nehal H Patel; Tarun H Madan; Amar M Panchal; Bhavesh M Thakkar
Journal:  Ann Pediatr Cardiol       Date:  2009-07

10.  Pseudo interruption of the inferior vena cava complicating the device closure of patent ductus arteriosus: Case report and short review of venous system embryology.

Authors:  Venkateshwaran Subramanian; Krishnamoorthy Kavassery Mahadevan; Sivasankaran Sivasubramonian; Jaganmohan Tharakan
Journal:  Ann Pediatr Cardiol       Date:  2014-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.