Acceptance procedure for the linear accelerator component of the 1.5 T MRI-linac.

Simon J Woodings1, J H Wilfred de Vries1, Jan M G Kok1, Sara L Hackett1, Bram van Asselen1, Johanna J Bluemink1, Helena M van Zijp1, Theo L van Soest1, David A Roberts2, Jan J W Lagendijk1, Bas W Raaymakers1, Jochem W H Wolthaus1.   

Abstract

PURPOSE: To develop and implement an acceptance procedure for the new Elekta Unity 1.5 T MRI-linac.
METHODS: Tests were adopted and, where necessary adapted, from AAPM TG106 and TG142, IEC 60976 and NCS 9 and NCS 22 guidelines. Adaptations were necessary because of the atypical maximum field size (57.4 × 22 cm), FFF beam, the non-rotating collimator, the absence of a light field, the presence of the 1.5 T magnetic field, restricted access to equipment within the bore, fixed vertical and lateral table position, and the need for MR image to MV treatment alignment. The performance specifications were set for stereotactic body radiotherapy (SBRT).
RESULTS: The new procedure was performed similarly to that of a conventional kilovoltage x-ray (kV) image guided radiation therapy (IGRT) linac. Results were acquired for the first Unity system.
CONCLUSIONS: A comprehensive set of tests was developed, described and implemented for the MRI-linac. The MRI-linac met safety requirements for patients and operators. The system delivered radiation very accurately with, for example a gantry rotation locus of isocenter of radius 0.38 mm and an average MLC absolute positional error of 0.29 mm, consistent with use for SBRT. Specifications for clinical introduction were met.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  MRI-linac; Unity; acceptance tests; dosimetry; magnetic field

Year:  2021        PMID: 34275176     DOI: 10.1002/acm2.13068

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  3 in total

1.  Commissioning measurements on an Elekta Unity MR-Linac.

Authors:  Marcus Powers; John Baines; Robert Crane; Chantelle Fisher; Stephen Gibson; Linda Marsh; Bronwyn Oar; Ariadne Shoobridge; Emily Simpson-Page; Marchant Van der Walt; Glenn de Vine
Journal:  Phys Eng Sci Med       Date:  2022-03-02

2.  Clinical Implementational and Site-Specific Workflows for a 1.5T MR-Linac.

Authors:  David A P Dunkerley; Daniel E Hyer; Jeffrey E Snyder; Joël J St-Aubin; Carryn M Anderson; Joseph M Caster; Mark C Smith; John M Buatti; Sridhar Yaddanapudi
Journal:  J Clin Med       Date:  2022-03-16       Impact factor: 4.241

3.  Longitudinal assessment of quality assurance measurements in a 1.5 T MR-linac: Part II-Magnetic resonance imaging.

Authors:  Ergys Subashi; Alex Dresner; Neelam Tyagi
Journal:  J Appl Clin Med Phys       Date:  2022-03-25       Impact factor: 2.243

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.