AIMS: This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.
METHODS: This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors' institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.
RESULTS: Adjacent to the nail's telescoping junction, osteolytic changes were observed in bi-planar radiographs of 20/57 segments (35%) after a mean of 9.5 months (95% confidence interval 7.2 to 11.9) after surgery. A total of 8/20 patients with osseous alterations (40%) reported rest and ambulation pain of the lengthened segment during consolidation. So far, 24 Stryde nails were retrieved and in 20 (83%) macroscopic corrosion was observed at the nail's telescoping junction. Before implant removal 11/20 radiographs (55%) of lengthened segments with these 20 nails revealed osteolysis. Implant retrieval analysis by means of SEM showed pitting and crevice corrosion. EDX detected chromium as the main metallic element of corrosion.
CONCLUSION: Patients are exposed to the risk of implant-related osteolysis of unclear short- and long-term clinical consequences. The authors advocate in favour of an early implant removal after osseous consolidation. Cite this article: Bone Joint Res 2021;10(7):425-436.

Entities:  

Keywords:  Intramedullary lengthening nail; Osteolysis; Stryde

Year:  2021        PMID: 34269599     DOI: 10.1302/2046-3758.107.BJR-2021-0146.R1

Source DB:  PubMed          Journal:  Bone Joint Res        ISSN: 2046-3758            Impact factor:   5.853


  6 in total

1.  Letter to the editor on "STRYDE versus PRECICE magnetic internal lengthening nail for femur lengthening".

Authors:  Björn Vogt; Jan Duedal Rölfing; Robert Roedl; Adrien Frommer
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-22       Impact factor: 3.067

2.  Limb reconstruction in a resource-limited environment.

Authors:  Nando Ferreira; Sanjeev Sabharwal; Gamal Ahmed Hosny; Hemant Sharma; Ashok Johari; Vasudevan P Nandalan; Mauro Vivas; Mangal Parihar; Selvadurai Nayagam; David Ferguson; Jan Duedal Rölfing
Journal:  SICOT J       Date:  2021-12-31

3.  Technical considerations and early results of magnetic compressive intramedullary nailing for humeral shaft delayed unions and nonunions.

Authors:  Daniel J Cognetti; Abdullah Ghali; Jose M Gutierrez-Naranjo; Jordan Handcox; Ravi Karia; Anil K Dutta
Journal:  JSES Int       Date:  2022-01-17

4.  Does Focal Osteolysis in a PRECICE Stryde Intramedullary Lengthening Nail Resolve after Explantation?

Authors:  Oliver C Sax; Larysa P Hlukha; Kyle A Kowalewski; John E Herzenberg; Philip K McClure
Journal:  Children (Basel)       Date:  2022-06-09

5.  The reconstruction of critical bone loss : the holy grail of orthopaedics.

Authors:  Shao-Ting J Tsang; Nando Ferreira; A H R W Simpson
Journal:  Bone Joint Res       Date:  2022-06       Impact factor: 4.410

6.  Correction of post-traumatic lower-limb discrepancy with Precice intramedullary lengthening nails: a review of 34 adults with an average follow-up of 2 years.

Authors:  Christof Radler; Gabriel T Mindler; Alexandra Stauffer; Carina Weiß; Rudolf Ganger
Journal:  Acta Orthop       Date:  2022-09-02       Impact factor: 3.925

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.