| Literature DB >> 34266516 |
Kristy L Buising1, Robyn Schofield2, Louis Irving3, Melita Keywood4, Ashley Stevens5, Nick Keogh5, Grant Skidmore6, Imogen Wadlow7, Kevin Kevin7, Behzad Rismanchi8, Amanda J Wheeler9, Ruhi S Humphries4, Marion Kainer10, Jason Monty6, Forbes McGain11, Caroline Marshall12.
Abstract
OBJECTIVE: To study the airflow, transmission, and clearance of aerosols in the clinical spaces of a hospital ward that had been used to care for patients with coronavirus disease 2019 (COVID-19) and to examine the impact of portable air cleaners on aerosol clearance.Entities:
Mesh:
Year: 2021 PMID: 34266516 PMCID: PMC8314194 DOI: 10.1017/ice.2021.284
Source DB: PubMed Journal: Infect Control Hosp Epidemiol ISSN: 0899-823X Impact factor: 6.520
Fig. 1.The effect of no air cleaners versus 2 air cleaners on aerosol clearance and transmission of aerosols within a patient room with the door closed. The left image shows the values normalized to the saturation value of the sensor whereas right shows the measured value. Note. The grey solid line indicates measures taken within the standard patient room. Black solid line indicates measures taken at nurses’ station. The grey dotted line indicates measures taken within the patient room with 2 air cleaners running. The black dotted line indicates measures taken at the nurses’ station when the 2 air cleaners were in the patient room.
Fig. 2.The effect of open vs closed door to patient room on aerosol clearance and transmission of aerosols. Left image shows the values normalized to the saturation value of the sensor whereas right shows the measured value. The measurements were taken with 2 air cleaners in the patient room. Note. The grey solid line indicates measures taken within the patient room with door to the corridor closed. The black solid line indicates measures taken at the nurses’ station (NS) when the door was closed. The grey dotted line indicates measures taken within the patient room with door to the corridor open. Black dotted line indicates measures taken at the nurses’ station (NS) when the door was open.
Fig. 3.Comparing different interventions at the nurses’ station on aerosol clearance and transmission of aerosols. Left image shows the values normalized to the saturation value of the sensor whereas right shows the measured value. The measurements were taken with 2 air cleaners in the patient room. Note. The black solid line indicates measures taken within the patient room. The grey solid line indicates measures taken at the nurses’ station (NS) with Zipwall present. The grey dashed line indicates measures taken at NS without barrier. The grey dotted line indicates measures taken at NS with air cleaner barrier present (3 air cleaners in front of desk).
Fig. 4.Rate of clearance of aerosolized smoke particles in the corridor with differing numbers of air cleaners. Some sources also called these “portable HEPA filters.” Line color goes from light to dark for cases with progressively more air cleaners, with 0, 4, 6, 8, 10, and 12 evenly spaced air cleaners considered in the corridor.