| Literature DB >> 34262309 |
Roberto W Dal Negro1, Paola Turco2, Massimiliano Povero3.
Abstract
BACKGROUND: The MiniBox+™ is an innovative technique for assessing lung volumes (LVs) and the diffusing capacity of the lung for carbon monoxide (DLco). Differently from the equipment needed for whole body plethysmography (WBP), the MiniBox+ is a small, transportable instrument, which derives total lung capacity (TLC) during tidal breathing by analyzing gas pressures and airflows immediately preceding and immediately following airway occlusions. AIM: To compare the consistency and the feasibility of LV and DLco measurements between the two instruments in different lung function disorders, and their cost of execution.Entities:
Keywords: CO diffusing capacity; MiniBox+™; economic impact; failure risk; lung volumes; whole body plethysmography
Year: 2021 PMID: 34262309 PMCID: PMC8275122 DOI: 10.2147/CEOR.S316529
Source DB: PubMed Journal: Clinicoecon Outcomes Res ISSN: 1178-6981
Baseline Characteristics of the Whole Sample and by Lung Disorder
| Variable | Whole | Asthma | COPD | Restrictive | |
|---|---|---|---|---|---|
| n | 164 | 43 (32.09%) | 47 (35.07%) | 44 (32.84%) | |
| Age (years) | 67.35±13.54 | 60.42±13.48 | 69.13±12.08 | 72.22±12.56 | 0.0001 |
| Sex (female) | 67 (50%) | 21 (48.84%) | 23 (48.94%) | 23 (52.27%) | 0.952 |
| BMI (kg/m2) | 25.88±4.57 | 26.78±4.92 | 25.49±4.58 | 25.43±4.17 | 0.2956 |
| Smoking status | 0.005 | ||||
| Never | 57 (42.54%) | 20 (46.51%) | 18 (38.30%) | 19 (43.18%) | |
| Former | 56 (41.79%) | 17 (39.53%) | 15 (31.91%) | 24 (54.55%) | |
| Smoker | 21 (15.67%) | 6 (13.95%) | 14 (29.79%) | 1 (2.27%) | |
| Workers | 73 (54.48%) | 31 (72.09%) | 25 (53.19%) | 17 (38.64%) | 0.007 |
| Schooling level | 0.062 | ||||
| Pre-primary school | 4 (2.99%) | 1 (2.33%) | 2 (4.26%) | 1 (2.27%) | |
| Primary school | 11 (8.21%) | 1 (2.33%) | 4 (8.51%) | 6 (13.64%) | |
| Lower secondary | 34 (25.37%) | 13 (30.23%) | 12 (25.53%) | 9 (20.45%) | |
| Upper secondary | 56 (41.79%) | 24 (55.81%) | 14 (29.79%) | 18 (40.91%) | |
| Degree | 29 (21.64%) | 4 (9.30%) | 15 (31.91%) | 10 (22.73%) |
Abbreviations: BMI, body mass index; COPD, chronic obstructive pulmonary disease.
Figure 1Overall failure risk with the MiniBox+™ and WBP.
Figure 2MiniBox+™ and WBP: analytic comparison of their failure risks.
MiniBox+™ and WBP: Comparison of Their Usability
| MiniBox+ | WBP | Δ (95% CI) | |
|---|---|---|---|
| Number of attempts | 1.44±0.62 | 2.81±1.35 | −1.20 (−1.38 to −1.01) |
| No failure | 1.37±0.56 | 2.57±1.18 | −1.19 (−1.39 to −1.00) |
| Failure | 2.27±0.65 | 3.81±1.58 | −1.48 (−1.96 to −1.00) |
| Total time (min) | 7.20±1.81 | 13.93±5.08 | −6.07 (−6.85 to −5.30) |
| No failure | 6.93±1.52 | 13.06±5.00 | −5.99 (−6.82 to −5.17) |
| Failure | 10.18±2.23 | 17.54±3.70 | −7.14 (−8.73 to −5.55) |
Note: Δ:, mean difference between MiniBoxPlusTM and WBP.
Abbreviations: CI, confidence interval; Δ, mean difference between MiniBox+ and WBP; WBP, whole body plethysmograph.
Comparison Among Parameters Measured by the MiniBox+™ and by WBP
| Parameters | MiniBox+ | WBP | Δ (95% CI) | Δ% (95% CI) |
|---|---|---|---|---|
| TLC (L) | 5.07±1.45 | 5.08±1.44 | −0.024 (−0.082 to 0.032) | −0.2% (−1.4 to 1.0) |
| FEV1 (L) | 2.32±1.11 | 2.27±1.07 | 0.043 (0.016 to 0.071) | 2.5% (1.1 to 4.0) |
| RV (L) | 2.02±0.84 | 2.10±0.82 | −0.129 (−0.196 to −0.061) | −5.0% (−7.9 to −2.1) |
| DLco (mL/min/mmHg) | 18.94±5.95 | 19.86±6.62 | −0.481 (−1.108 to 0.147) | 1.7% (−4.5 to 7.9) |
Abbreviations: CI, confidence interval; Δ, mean difference between MiniBox+ and WBP; DLco, diffusing capacity of the lung for carbon monoxide; FEV1, forced expiratory volume in 1 second; RV, residual volume; TLC, total lung capacity; WBP, whole body plethysmograph.
Figure 3MiniBox+™ vs WBP: comparison of their economic impact.