| Literature DB >> 34259563 |
Ojas Kaduskar1, Vaishali Bhatt1, Christine Prosperi2, Kyla Hayford2, Alvira Z Hasan2, Gururaj Rao Deshpande1, Bipin Tilekar1, Jeromie Wesley Vivian Thangaraj3, Muthusamy Santhosh Kumar3, Nivedita Gupta4, Manoj V Murhekar3, William J Moss2,5,6, Sanjay M Mehendale4, Lucky Sangal7, Gajanan Sapkal1.
Abstract
Blood collection using dried blood spots (DBS) provides an easier alternative to venipuncture for sample collection, transport, and storage but requires additional processing that can cause variability in results. Whole-blood samples spotted on four DBS devices and respective paired serum samples were tested for antimeasles and antirubella IgG antibody concentrations by enzyme immunoassay. Elution protocols for DBS devices were optimized for comparability relative to serum samples using 12 adult volunteers. Stability of DBS collected on HemaSpot HF was assessed under various temperature conditions (+4, 22 to 25, and 45°C) at six time points (0, 7, 15, 30, 60, and 90 days) in a controlled laboratory setting using six adult volunteers. Devices were shipped and stored for 30 days at four settings with variable temperature and humidity conditions to assess the impact on antibody concentrations. Three DBS devices demonstrated comparable antibody concentrations with paired sera following optimization. Antibodies recovered from DBS were stable for at least 90 days at 4°C and for 30 days at ambient temperature (22 to 25°C) using the HemaSpot HF device. A drastic decline in antibody concentrations was observed at 45°C, resulting in quantitative and qualitative discrepancies by day 7. HemaSpot HF devices shipped to field sites and stored at ambient temperature and humidity resulted in quantitative, but not qualitative, variability. Measurement of antimeasles and antirubella IgG antibodies with DBS devices is an accurate alternative to testing serum, provided elution protocols are optimized. Stability of HemaSpot HF devices at ambient temperature enables broader use in surveys when serum processing and cold storage are not feasible. IMPORTANCE Dried blood spot (DBS) collection offers various advantages over conventional methods of blood collection, especially when collecting and transporting samples for a serosurvey. Yet use of DBS requires additional processing steps in the laboratory that can add to variability in results. We optimized a protocol to elute IgG antibodies against measles and rubella viruses in four DBS devices, demonstrating high concordance with paired venous sera for most devices. Extensive stability studies with various temperature and storage conditions in the laboratory and in the field were conducted using HemaSpot HF DBS devices prior to its use in one of the largest community-based measles and rubella serological surveys in the world.Entities:
Keywords: ELISA; HemaSpot; dried blood spot; elution buffer; elution protocol; laboratory and field; measles; optimization; quantitative IgG titers; rubella; stability; temperature and humidity
Mesh:
Substances:
Year: 2021 PMID: 34259563 PMCID: PMC8386427 DOI: 10.1128/mSphere.00490-21
Source DB: PubMed Journal: mSphere ISSN: 2379-5042 Impact factor: 4.389
FIG 1Box plot showing absolute differences in measles (A) and rubella (B) antibody concentration by DBS collection device. The DBS collection devices were Whatman 903, HemaSpot HF, HemaSpot SE, and TropBio. Twelve participants were included in the experiment. Triplicate samples were run on three separate days for a total of nine results per participant for serum sample and for each device. Mean antibody concentrations were calculated for each participant across the replicates for each device or serum sample. The plot depicts the distribution of absolute differences between the mean serum value and mean DBS device values across the 12 participants (gray circles overlaid on boxplot). The diamond represents the group mean, and the horizontal line inside the box is the median.
Summary of four dried blood spot (DBS) collection devices selected for initial evaluation
| Characteristic | Whatman 903 Protein Saver | HemaSpot HF | HemaSpot SE | TropBio |
|---|---|---|---|---|
| Device description | Folded card with five half-inch circles printed on filter paper | Self-contained device with wedge-shaped filter paper, built-in desiccant, and a plastic cover | Self-contained device with spiral shape filter paper to separate serum, built-in desiccant, and a plastic cover | Circular card with six protrusions of filter paper, 4 cards packed in 1 sheet |
| Units (e.g., spots) and whole blood volume per device | 5 spots∼375 μl/device∼75 μl/spot | 8 wedges∼150 μl/device | Blood components separate out onto four sections in spiral form (whole blood, plasma, serum); no. of punched spots depend on blood component of interest∼150 μl/device | 6 protrusions∼60 μl/device∼10 μl/protrusion ( |
| Collection process | Fill circles independently; potential for blood overlay | Blood flows through a center hole in device to fill all wedges simultaneously; excess blood flows to outer membrane | Blood flows through a center hole in device to spread out through filter paper | Fill protrusions independently; potential for blood overlay |
| Serum separation | No | No | Yes | No |
| Integrated collection and storage | No. Card dried >4 h in clean dry space before storage in self-sealing bag with desiccants | Yes. Device shut closed for storage soon after sample collection; built-in desiccant | Yes. Device shut closed for storage soon after sample collection; built-in desiccant | No. Card dried >4 h in clean dry space before storage in self-sealing bag with desiccants |
| Potential for contamination with other samples | Possible. Spots are covered by a thin flap and require extended drying with flap open | Less likely. Device closed immediately after sample collection | Less likely. Device closed immediately after sample collection | Possible. Spots protrude out from center with no barrier and requires extended drying |
| Excision device | Punching device cleaned with alcohol between participants | Scissors and forceps for detaching wedges cleaned with alcohol between participants | Punching device cleaned with alcohol between participants | Scissors and forceps for detaching protrusion cleaned with alcohol between participants |
| Image |
|
|
|
|
| Strengths | Established card used in studies and many newborn screening programs. | Integrated collection device facilitates ease of collection. | Integrated collection device facilitates ease of collection. | Does not require punching, reduces variability in blood volume per protrusion. |
| Existing evidence on comparison between card and serum samples. | Self-contained device and built-in desiccant reduce drying steps and protects from humidity. | Built-in serum separation. | Requires very small blood volume per protrusions which are independently filled. | |
| Requires small blood volume per spot, independently filled | Devices include barcode for sample identification. | Self-contained device and built-in desiccant reduce drying steps and protects from humidity. | ||
| Devices include barcode for sample identification. | ||||
| Weaknesses | Requires more extensive drying procedures. | Limited data on sensitivity and stability. | Limited data on sensitivity and stability. | Requires more extensive drying procedures. |
| Elevated risk of contamination during drying or punching. | Device to be used within 30 min after opening. | Device to be used within 30 min after opening. | Elevated risk of contamination during drying. | |
| Potential for variability in volume between punches. | Elevated risk of undersaturation due to larger volume requirement and nonindependence of filling. | Four sections of the device are not clearly defined and concentration in each section dependent on the volume of blood collected. |
Merck. Whatman protein saver cards 903 protein saver card (US) (package of 100 each; Sigma-Aldrich).
Laboratory observed volume required to fully saturate device and all wedges.
Volume per spot punched from HemaSpot SE device could not be estimated due to diffusive nature of the serum separation paper (24; also Protein Extraction Methods from HemaSpot HF or SE Devices Standard Curve Preparation).
HemaSpot devices may be placed into a self-sealing bag with a desiccant to further prevent exposure to humidity, although this is not required (4).
For this study, TropBio cards were packed in petri plates with desiccants out of an abundance of caution to avoid damage to the protrusion.
HemaSpot images from manufacturer’s website (HemaSpot HF; Spot On Sciences [https://www.spotonsciences.com/hemaspot-hf]; HemaSpot SE; Spot On Sciences [https://www.spotonsciences.com/hemaspot-se]).
FIG 2Effect of storage duration and temperature on measles (A) and rubella (B) IgG antibodies obtained from HemaSpot HF devices stored at the central laboratory. The ambient temperature in the laboratory was an average temperature of 24.8°C (range, 24 to 26.7) and average humidity of 73.8% (range 63 to 80%). The average relative humidity (RH) at 4°C was 92.1% (maximum RH, 100%). The average RH at 45°C was 16.7% (maximum RH, 20.1%). Day 0 is the same in all three temperature graphs, reflecting day 0 at ambient temperature where nine replicates were available for each participant. For all other time points, the samples were run in triplicates. Solid circles represent the average of the replicates, and open circles represent the values for individual replicates. The dashed black line indicates 200 mIU/ml for measles and 8 IU/ml for rubella (threshold between negative and equivocal; values above the dashed black line were categorized as positive in this analysis).
FIG 3Measles (A) and rubella (B) IgG antibody concentrations obtained from HemaSpot HF devices stored for 30 days at sites with various temperature and humidity conditions. Each dot represents the value for one of the nine replicates for each storage site and participant. The dashed horizontal line represents the mean antibody concentration from the DBS day 0 (D0) samples. Temperature and humidity conditions reflect when the experiment was conducted (July 2019).
Difference in mean IgG antibody concentration obtained from dried blood spots, comparing baseline (day 0 samples stored at central laboratory) with samples stored for 30 days at sites
| Antibody and ID | Day 0 | Dibrugarh | Chennai | Jaipur | Ghatampur | ||||
|---|---|---|---|---|---|---|---|---|---|
| Mean | Mean | Mean | Mean | ||||||
| Measles | |||||||||
| 1 | 95.7 | −3.8 | 0.59 | −3.7 | 0.61 | −10.8 | 0.14 | −13.6 | 0.10 |
| 2 | 345.8 | −2.0 | 0.87 | −26.3 | 0.06 | −24.5 | 0.04 | −24.5 | 0.07 |
| 3 | 437.5 | 153.1 |
| 2.9 | 0.80 | −95.0 |
| −92.2 |
|
| 4 | 3,986.9 | 182.8 | 0.31 | 40.3 | 0.80 | −647.6 |
| −641.5 |
|
| 5 | 1,917.0 | −62.3 | 0.47 | −123.9 | 0.12 | −715.18 |
| −560.6 |
|
| 6 | 3,445.9 | −104.6 | 0.33 | −150.8 | 0.12 | −297.3 | 0.03 | −229.6 | 0.05 |
| Rubella | |||||||||
| 1 | 64.6 | 1.3 | 0.86 | −16.0 | 0.02 | −13.3 | 0.06 | −19.2 | 0.01 |
| 2 | 5.3 | −0.7 | 0.03 | −0.9 |
| −0.2 | 0.46 | −0.0 | 0.92 |
| 3 | 46.7 | 16.4 |
| −4.6 | 0.02 | −11.9 |
| −12 |
|
| 4 | 161.0 | 4.2 | 0.27 | 5.0 | 0.23 | −28.5 |
| −26.7 |
|
| 5 | 19.8 | 5.0 |
| −2.2 | 0.05 | −6.1 |
| −6.1 |
|
| 6 | 169.1 | −18.4 |
| −18.1 |
| −31.3 |
| −23.8 |
|
ID, identifier.
D0, day 0.
Mean difference calculated as day 30 site value minus day 0 value calculated as the mean across the nine replicates for each individual.
P values obtained from one-way analysis of variance test. Boldface values show P < 0.002 (Bonferroni-corrected P value accounting for 24 pairwise comparisons).
Temperature and humidity conditions at storage sites
| Characteristic | Dibrugarh | Chennai | Jaipur | Ghatampur |
|---|---|---|---|---|
| Temperature | ||||
| Mean (SD) (sites) | 28.6 (1.7) | 30.2 (1.1) | 30.0 (1.3) | 32.5 (1.3) |
| Mean (SD) (weather station) | 27.7 (2.3) | 31.6 (1.6) | 30.5 (2.5) | 29.7 (2.1) |
| Humidity, mean (SD) (sites) | 95.4 (5.0) | 60.0 (6.3) | 61.7 (10.8) | 69.5 (5.0) |
Temperature and humidity reflect the mean of the afternoon measurements obtained by staff. Temperature and humidity conditions of the four sites (July 2019) are as follows: Dibrugarh, moderate temperature, high humidity; Chennai, moderate temperature, moderate humidity; Jaipur, high temperature, moderate humidity; Ghatampur, high temperature, higher humidity.
Temperature from weather station reflect outside temperature obtained from Global Historical Climatology Network - Daily (GHCN-Daily) data set (21, 22). For Ghatampur, the nearest weather station with available data is Lucknow (approximately 96 km northeast).