| Literature DB >> 3425272 |
P Aspenberg1, L Kolmert, L Persson, R Onnerfält.
Abstract
Twenty-five patients underwent total hip arthroplasty using a copy of a well-known prosthesis during the period 1974-1976. Five of the prostheses fractured through the neck after 5-12 years. There were no prodromal symptoms and no trauma preceded the fractures. Technical analysis revealed that inadequate welds were responsible.Entities:
Mesh:
Year: 1987 PMID: 3425272 DOI: 10.3109/17453678709146382
Source DB: PubMed Journal: Acta Orthop Scand ISSN: 0001-6470