| Literature DB >> 34249751 |
Jiayao Sun1,2,3, Lin Kong2,3,4, Zhi Chen1,2,3, Dan You2,3,5, Jingfang Mao2,3,4, Xiyin Guan2,3,6, Xiaodong Wu1,2,3, Yinxiangzi Sheng1,2,3.
Abstract
PURPOSE: To verify the practicality and safety of a treatment chair with six degrees of freedom (6DTC) through demonstrating the efficacy of the workflow in clinical settings and analyzing the obtained technical data, including intra-fraction patient movement during the use of the 6DTC.Entities:
Keywords: carbon-ion radiotherapy (C-ion RT); intra-fraction movement; setup error; treatment chair; upright posture
Year: 2021 PMID: 34249751 PMCID: PMC8260974 DOI: 10.3389/fonc.2021.694749
Source DB: PubMed Journal: Front Oncol ISSN: 2234-943X Impact factor: 6.244
Figure 1The clinical treatment workflow for this clinical trial.
Figure 2The patient setup on the 6DTC.
Characteristics of patients.
| Characteristics | No. of patients |
|---|---|
|
| |
| Median (Range) | 48 (21-83) |
|
| |
| Median (Range) | 61 (49-105) |
|
| |
| Median (Range) | 168 (146-177) |
|
| |
| Male | 9 |
| Female | 6 |
|
| |
| 100 | 7 |
| 90 | 8 |
|
| |
| Nasopharyngeal carcinoma (NPC) (Nasopharyngeal) | 8 |
| Meningioma (Cerebellopontine Angle) | 1 |
| Adenoid cystic carcinoma (Maxillary sinus) | 1 |
| Squamous cell carcinoma (External auditory canal) | 1 |
| Adipose-derived tumor (Orbit) | 1 |
| Osteosarcoma (Orbit) | 1 |
| Atypical carcinoid tumor (Slope) | 1 |
| Chordoma (Skull base) | 1 |
|
| |
| 5* | 9 |
| 8 | 1 |
| 10 | 1 |
| 20 | 3 |
| 27 | 1 |
|
| |
| 2 | 14 |
| 3 | 1 |
*For these nine patients, the prescriptions were 56 Gy(RBE) delivered by either protons or x rays in 28 fractions using a lying posture plus 17.5Gy(RBE) by carbon ions in five fractions using a sitting posture.
Figure 3The deviation between PE-B1 and PE-B2. The x-axis represents the fifteen individual patients; the y-axis represents the mean value of the deviation of the specific patient in 6 degrees of freedom. Error bars represent for the standard deviation.
Frequency of deviation [%] between PE-B1 and PE-B2 by thresholds.
| Translational | Rotational | ||||||
|---|---|---|---|---|---|---|---|
| x | y | z | u | v | w | ||
| >1mm | 41(27.3)* | 50(33.3) | 40(26.7) | >1° | 2(1.3) | 4(2.7) | 3(2.0) |
| >2mm | 1(0.7) | 10(6.7) | 8(5.3) | >2° | 0(0.0) | 0(0.0) | 0(0.0) |
| >3mm | 0(0.0) | 2(1.3) | 1(0.7) | >3° | 0(0.0) | 0(0.0) | 0(0.0) |
| >4mm | 0(0.0) | 2(1.3) | 0(0.0) | >4° | 0(0.0) | 0(0.0) | 0(0.0) |
| >5mm | 0(0.0) | 0(0.0) | 0(0.0) | >5° | 0(0.0) | 0(0.0) | 0(0.0) |
*Data in parentheses are percentages.
Figure 4Intra-fraction patient movements. The x-axis represents the three translational directions and three rotational directions, while the y-axis represents the amplitude of the intra-fraction patient movements. The error bars represent the standard deviation.