Godwin A Udourioh1, Moses M Solomon2, Emmanuel I Epelle3. 1. Analytical/Material Chemistry Laboratory, Department of Pure and Applied Chemistry, Faculty of Natural and Applied Sciences, Veritas University, Abuja, P.O.Box 6523, Garki, Abuja Nigeria. 2. Department of Chemistry, College of Science and Technology, Covenant University, Canaanland, Km10, Idiroko Road, Ota, Ogun State Nigeria. 3. Institute for Materials and Processes (IMP), School of Engineering, University of Edinburgh, The King's Buildings, Edinburgh, EH9 3FB UK.
As at 6.03 pm CEST, 10 March, 2021, there have been 118,684,343 COVID-19 cases, including 2,633,281 deaths confirmed globally by WHO, ECDC, NCDC and John Hopkins University.120 Since the first case of COVID-19 was detected in China’s Hubei province in late 2019, stringent measures such as lockdowns, travel bans, border closure etc. have been imposed in order to control or stop the spread of the outbreak, yet the global spread of the virus has continued to record significant increase with attendant global socio-economic burden.31,88 In fact, reports from WHO and John Hopkins University reveals that the second wave of the outbreak is more devastating thus has left people in great fear.93Sample testing is an essential first step to responding to any pandemic outbreak.31,45 Diagnosis plays a decisive role in making prompt decisions on detection, contact tracing, isolation, management and treatment of infected persons.45 However, in the ongoing COVID-19 outbreak, most countries are unable to meet up with the massive diagnostic testing order given by WHO.79,83,86 This has resulted in a continuous spread due to community transmission.79,86As suggested by WHO, the general benchmark for adequate testing for a positive rate is around 3–12% per 1000 persons.23 South Korea, Uruguay, Germany and Australia recorded a positive rate of 1% hence being considered as countries with lowest COVID-19 related deaths in the world.24,37 South Korea was able to achieve the feat through their intensive testing programs occasioned by “drive-through” and “phone booths” tests.81 On the other hand, countries like Mexico and Nigeria have positive rates of 20–50% (a case is found for every few tests conducted), indicating the unlikelihood of testing widely enough to find all cases.37 No wonder the number of new confirmed cases keep increasing daily.120Apart from the disparity in country’s political and policy frameworks which may hinder the control of the continuous spread of the pandemic, the biosensing technologies (technologies behind testing and diagnosis) are important factors to consider. A biosensor is a device used for the detection of biological and biochemical agents; employing a biologically derived or a biomimetic recognition element while either undergoing a biochemical reaction (for example, enzyme-based biosensors) or binding the target molecule in a highly specific way. Studies have shown that biosensing devices, materials or technologies for testing infectious diseases or at time of pandemic outbreak must rigorously satisfy requirements of accessibility and affordability, rapidity, high sensitivity and selectivity, robustness, flexibility and simplicity in usage, ability to be mass produced etc.15,81The present trend in the daily reports on the confirmed cases of the COVID-19 globally may be a pointer to the fact that the biosensing technologies currently in used for SARS-COV-2 testing are not satisfactory. It is pertinent to review the type of assays, strength and limitations of the commonly used biosensing methods since the pandemic outbreak and explore the potentials of other versatile biosensing materials and technologies such as metal organic frameworks (MOFs) for possibility of development and utilization in solving diagnostic problem associated with the current and future pandemic outbreaks.
Metal Organic Frameworks (MOFs) and Their Characteristics
Metal Organic Frameworks (MOFs) are advanced structures that are highly ordered, porous and customizable. They grow in a crystal form and are extremely flexible, especially when combined with nanoparticles for additional functionality or attributes.70,137 MOFs are made of metal clusters coordinated with organic linkers to generate a large Langmuir surface area and small-to-medium-sized pores.54 MOFs are defined as porous structures constructed from the coordinative bonding between metal ions and organic ligands or bridging ligands.138 The linkers or bridging ligands consist of carboxylates, or anions, such as phosphonate, sulfonate, and heterocyclic compounds while the inorganic units are the metal ions or clusters called secondary building units (SBUs).138 The coordination number, geometry of the metal ions and the nature of the functional groups determine the geometry of MOFs. Based on this we have octahedron with six points of extension, trigonal prism with five points, square paddle‐wheel (four points), and triangle with three points. Some commonly used metals for synthesis of MOFs include La, Zn, Cr, Cu, In Co, Fe and Ag while some common organic linkers or ligands include 1,4‐benzenedicarboxylate or terephthalate moiety (H2bdc), Benzene‐1,3,5‐tricarboxylate moiety (H3btc), 4,4′‐biphenyldicarboxylate (H2bpdc), 1,4-bis(imidazol-1-ylmethyl)benzene (Bix), 1,3,5-benzenetriphosphoric acid, 1,5-naphthalenedisulfonic acid, 4,4-bipyridine, 2,5‐dihydroxybenzene‐1,4‐dicarboxylic acid (H4dhbdc), 2,6-naphthalenedicarboxylic acid (H2ndc), adamantane tetracarboxylic acid (H4atc), 4,4′,4″-benzene-1,3,5-tryyl-benzoic acid (H3btb).138 Figure 1 shows the typical skeletal structure of MOF and some examples of ligand structures.
There are different sensing platforms, viz. luminescence, surface plasmon resonance, electrochemical, impedance, fluorescence imaging (magnetic resonance imaging MRI), interferometry and solvatochromism.18,49,69,97,98,142 Recently, MOFs have been explored as Biological and biochemical sensors.11,36 Hao and Yan36 developed a lanthanide-functionalized MOF as a fluorescent probe for hippuric acid in urine which was considered as the biological indicators of toluene exposure. The fabricated sensor, according to the authors has several attractive features, including high sensitivity, excellent selectivity, fast response time (~ 1 min), broad linear range (0.05–8.0 mg/mL), and good reversibility and regeneration.36 The sensor was successfully applied to determination of hippuric acid in human urines with recoveries in the range of 93.5–102.9%. The high porosity, tuneable chemical composition, large surface area, high crystallinity, and potential for post synthetic modification for molecular recognition have made MOFs promising candidates for biosensing application.80 Besides, the inherent luminescence of many MOFs have made it useful in sensing platforms.97,98 Some MOFs and their biosensing applications are summarized in Table 1.
Table 1
Some MOFs and their biosensing applications/detection limits
S. no.
MOF
Type
Biosensing applications
Detection limits (nM/ppm/ppb)
References
Nucleic acid sensing
1
Cu(H2dtoa)
Transition metal MOFs
i. Detection of HIV DNA
ii. Detection of Thrombin
DNA—3 nM
Thrombin—1.3 nM
144
2
[Cu3(Cmdcp)2(dps)·H2O)4(SO4)]n
Transition metal MOFs
i. Detection of HIV-1 dsDNA sequence
ii. Sudan Virus RNA sequence
HIV DNA-196 pm
SUDV RNA-73 pm
130
3
Zn(L).(HDMA)2(DMF)(H2O)6
Transition metal MOFs
Nucleic Acid Detection
0.05 nM
111
4
Cd(L).(HDMA)2(DMF)(H2O)3
Transition metal MOFs
Nucleic Acid Detection
0.05 nM
111
Enzymes and proteins sensing
5
HKUST-1
Transition metal MOFs
Detection of thiamine with intrinsic peroxidase-like activity
–
105
6
La-atp
Transition metal MOFs
Label-free assay of polyphenol oxidase
0.00012U mL−1
60
7
UiO-66-NH2
REM MOFs
Detection of Hg2+ ions using FAM-labeled ssDNA
17.6 nM
121
8
ATP-Ce-Tris
REM MOFs
Artificial peroxidase like activity and detection of H2O2
0.6 nM
139
9
Ru-PEI@ZIF-8 complex
Heterometallic MOFs
Assay for telomerase activity
11 cells
126
10
Eu@Sc-MOFs
Heterometallic MOFs
Detection of PS biomarker PGA in serum and urine
4.16ppb
63
11
ZIF-8@BHb composites
Composite MOF structures
Peroxidase-like activity and detection of H2O2 and phenol
1.0 µM for each analyte
135
Miscellaneous
12
Ir-Cd-Eggsshell membrane-GOx layer
Heterometallic MOFs
Biosensing of glucose
0.01 mM
39
13
Ir-Zne MOFs
Heterometallic MOFs
Biosensing of glucose
0.05mM
17
14
UiO-66-NH2 encapsulated metal ions (Cd2+ or Pb2+) and aptamer
Composite
MOF
Structures
Multiplex antibiotic
Detection
OTC - 0.18 pM
KAN - 0.15 pM
16
15
Chitosan-immobilized Cu-MOFs and tyrosinase
Transition metal MOFs
Biosensing of bisphenol
15.33 nM
72
16
MOF-5
Transition metal MOFs
Sensing of BSA
–
52
PEI polyethyleneimine, ZIF zeolitic imidazole frameworks, UiO University of Oslo
Some MOFs and their biosensing applications/detection limitsi. Detection of HIV DNAii. Detection of ThrombinDNA—3 nMThrombin—1.3 nMi. Detection of HIV-1 dsDNA sequenceii. Sudan Virus RNA sequenceHIV DNA-196 pmSUDV RNA-73 pmCompositeMOFStructuresMultiplex antibioticDetectionOTC - 0.18 pMKAN - 0.15 pMPEI polyethyleneimine, ZIF zeolitic imidazole frameworks, UiO University of Oslo
MOFs as Biosensing Materials in Pandemic Outbreaks
Viruses are often the culprit in epidermic and pandemic outbreaks. They are infectious agents, mostly in nanoscale capable of causing various diseases.82 MOFs have been used as biosensors during epidermic and pandemic outbreaks.
Sensing of Human Immunodeficiency Virus
The retrovirus is a RNA virus whereby its DNA is integrated into its host chromosomal DNA.8,82 Detection at the early stage of infection may be difficult due to the rare proviral DNA expression in the infected host.99 The Human Immunodeficiency Virus (HIV), which belongs to the genus Lentivirus within the family of Retroviridae and subfamily Orthoretrovirinae99 is a human retrovirus. Based on the genetic characteristics and differences in the viral antigens, there are two types of HIV: HIV-1 and HIV-2.99,100 The HIV-1 type is believed to have evolved from non-human primate immunodeficiency viruses from the Central African chimpanzees (SIVcpz)30,32 while the HIV-2 type is linked to the West African sooty mangabeys (SIVsm) as the origin.100 HIV infection results in acquired immunodeficiency syndrome (AIDS), a disease that is associated with the depletion of the CD4+T cell of the host.8 According to the WHO, at the end of 2019, an estimated 38.0 million people are living with HIV and about 33 million deaths have resulted from AIDS-related sicknesses.118 Because early diagnosis and treatment of HIV can improve survival and reduce morbidity, the Centers for Disease Control and Prevention have recommended routine testing.4 Examples of such routine test are the western blot and enzyme-linked immunosorbent assay (ELISA) assay.82 Nevertheless, because of reaction of samples with one or more of the antigens, these methods suffer from some false positive and negative outcomes.82 Researchers have taken advantage of large specific surface area, high porosity, fluorescence quenching, high loading efficiency, easy functionalization, and tunable pore properties of MOFs to deploy them in biosensing applications including the biosensing of HIV.Yang et al.130 applied [Cu3(Cmdcp)2(dps)4.(H2O)4(SO4)]n for the detection of human immunodeficiency virus-1 double-stranded DNA (HIV-1 ds-DNA). The 3-dimentional structure of the MOF enhanced the distinction between the ds-DNA and ss-DNA molecules. The intrinsic quenching properties of the unsaturated Cu(II) metal ion coordination centre and the conjugated π-electron system of the aromatic groups on both linkers enabled electrostatic and hydrogen bonding via
π-stacking interactions of the probe DNAs with the MOF, leading to photoinduced electron transfer (PET) fluorescence quenching. There was also a strong interaction between the probe DNA and the target DNA sequence. The non-target DNA sequences were between 50 and 86% less fluorescence than the target sequence in the dsDNA assay due to the diminished effect of its concentration. The probe recorded a high selectivity and 196pM detection limit for the viral dsDNA.130 Notably, the interaction of the MOF [Cu3(Cmdcp)2(dps)4·(H2O)4(SO4)]n with the complimentary sequences of HIV ds-DNA: carboxyfluorescein FAM-labeled probe ss-DNA, 5′-FAM-TTCTTCTTTTTTCT-3′ (P-DNA-1) and SUDV RNA: 5-FAM-TTAAAAAGTTTGTCCTCATC-3 (P-DNA-2) showed that the fluorescence intensity of the complimentary sequences of both HIV ds-DNA and SUDV RNA decreased upon the addition of the MOF. The quenching efficiency (QE%) of both HIV ds-DNA and SUDV RNA sequences were 65 and 76% respectively, indicating that the MOF efficiently quenched the fluorescence of both P-DNA-1 and P-DNR-2 sequences. The fluorescence spectra of both HIV-1 ds-DNA and SUDV RNA complementary sequences are presented in Fig. 4.
Ebola virus disease (EVD), formerly known as Ebola haemorrhagic fever is a viral hemorrhagic fever of humans and other primates that first appeared in 1976 in two simultaneous outbreaks.44 EVD is caused by Ebola viruses (EBOV), single-strand RNA viruses of the family Filoviridae.78 There are about five species of EBOV, namely Zaire, Bundibugyo, Sudan, Reston and Tai Forest. Although the fatality rate varies from specie to specie of EVD, it is in the range of 50–90%.44,104 The chronology of previous Ebola virus disease outbreaks and the actual fatality rate can be found in the WHO recent reports.119 EBOV infects its host cell by attaching to the receptors through the GP glycoprotein and getting endocytosed in host vesicles.44 The C-type lectins DCSIGN and DC-SIGNR is pivotal in the process as they bind to Ebola glycoproteins.44 The entry pathway of EBOV into host cell, the binding to cell-surface receptors, the slashing of the viral GP1 protein into N-terminal fragment Ebola within the endosome, and the digestion of cathepsin B into GP2 are illustrated in Fig. 7. The laboratory diagnosis of EBOV includes polymerase chain reaction, enzyme-linked immunosorbent assay (ELISA), antigen ELISA, immunohistochemistry, fluorescence assay, electron microscopy, indirect immunofluorescence assay (IFA), immuno-blot (western blot), biosensors SPR, QCM, optical, and DNA-based fluorescence nanobarcodes methodology.19,44,104
Sensing of Zika Virus and Dengue Virus RNA Sequence
Zika and Dengue are mosquito-borne flavivirus.116,117 Zika virus (ZIKU) was first identified in Uganda in 1947 in monkeys and later in humans in 1952 in Uganda and the United Republic of Tanzania.116 Beside Africa, ZIKU disease outbreaks have been recorded in Asia, the Pacific, and the Americas.116 Dengue virus (DENV) is prevalent in the tropical and sub-tropical climates worldwide, mostly in urban and semi-urban areas. There are four DENV serotypes (DENV-1, DENV-2, DENV-3 and DENV-4), implying that it is possible to be infected four times.117 Presently, the diagnoses of DENV and ZIKV are based on virus isolation cultures, Enzyme-linked immunosorbent assay (ELISA), genomic RNA detection using Polymerase Chain Reaction (PCR), and fluorescent biosensor.57 As earlier mentioned, the fluorescent biosensing technology is gaining attention in the detection of virus nucleic acids because of the advantages of high sensitivity and rapid response time.57,60,105,111Xie et al.124 successfully applied a three-dimensional Cu-based zwitterionic MOF, [Cu(Dcbcp)(bpe)]n (1, Dcbcp = N-(3,5-dicarboxylbenzyl)-(3-carboxyl) pyridinium, bpe = 1,2-bis(4-pyridyl)ethylene), for simultaneous detection of ZIKV and DENV RNA sequences. The detection limits for ZIKV were 192 pM and 121pM for single and synchronous fluorescence analysis respectively. For DENV, single and synchronous fluorescence analysis recorded 332 and 184pM detection limits respectively. The MOF so synthesized and characterized could form electrostatic, π stacking and/or hydrogen bonding interactions with two different fluorophore-labeled DNA probes which created two sensing systems used as fluorescent sensing platforms for the viruses’ detection.124 The single and synchronous fluorescence analysis recorded high selectivity and sensitivity. There was less interference with other mismatched RNA sequences. The authors reported that the sensing system showed potential application in the diagnosis of other virus associated infectious diseases which have similar clinical manifestations.124
Sensing of Thrombin
The MOF Cu(H2dtoa) had also been used to target thrombin.80,144 Thrombin is an endogenous protein or natural enzymes involved in the coagulation cascade, where it has a key role in the formation of fibrin clots by converting fibrinogen to fibrin.53 The same assay principle for HIV-1 U5 long terminal repeat DNA sequence was applied but with detection limit of 1.3 and 5–100 nM linear range. The sensor system was highly sensitive and selective as well and there was no effect on lysome bovine serum albumin and human IgG enzymes on signal.80,144 The fluorescence spectra of the FAM-labeled probe DNA 2–Cu(H2dtoa) in the presence of different concentrations of thrombin is presented in Fig. 5b.
Detection Assays Relevant to SARS-COV-2 and COVID-19 Diagnosis
The detection assays relevant to COVID-19 diagnostic testing and screening are categorized based on detection targets: (i) nucleic acid tests: which detects the presence of viral ribonucleic acids (RNA) using amplification steps based on reverse transcriptase polymerase chain reactions (RT-PCR); (ii) antigen tests: which detect the presence of a viral antigen as part of a surface protein; and (iii) antibody tests: which detect the presence of the antibodies such as immunoglobulin M (IgM) and IgG generated against the severe acute respiratory coronavirus-2 (SARS-COV-2).1,62In any case, enzyme-linked immunosorbent assays (ELISA), lateral flow assays (LFA) and chemoluminescence assays (CLIA) are the three most used assays. Besides, virus neutralisation tests and whole genome sequencing may be used to specifically detect neutralising antibodies and determine the sequence of the SARS-CoV-2 virus respectively. LFA is used for testing pregnancy and ovulation at home. It usage is simple, equipment-free hence cost-effective, and rapid thus very essential in biosensing and nanotechnology.47,76,81,85 In addition to the understanding of the assays, the diagnostic specimens are important factors to consider. Data comparing the accuracy of test methods suggest that test sensitivity may vary by type of specimen.1,55 Specimen for SARS-CoV-2 diagnostic tests can be taken from the upper (nasopharyngeal/oropharyngeal swabs, nasal aspirate, nasal wash or saliva) or lower (sputum or tracheal aspirate or bronchoalveolar lavage) respiratory tracts.84 According to Yang et al.,133 the most accurate sample for the diagnosis of SARS-CoV-2 is sputum, followed by nasal swabs and throat swabs.
Current Biosensing Techniques Focused on COVID-19 Diagnosis
The technology behind testing is biosensing.22,81 A biosensor is an analytical device which detect biological and biochemical component by employing biomimetic elements or biologically derived materials (nucleic acids, antibodies, antigen, enzymes, organelles, tissues, microorganisms etc.) that interact with, bind with, or recognize the analyte under study.7 The detector or transducer makes use of optical, electrochemical, electrochemiluminescence and piezoelectric principles to transform one signal into another for easy measurement and quantification.7 The biosensing techniques for SARS-COV-2 basically employ nucleic acids/DNA and antibody/antigen biomimetic components or biologically derived materials.94 The methods available in literature for COVID-19 testing are presented in Table 2 and the current diagnostic methods and potential portable biosensors for COVID-19 demonstrated diagrammatically in Fig. 9.
Table 2
Current biosensing techniques focused on COVID-19 diagnosis.
Prospective View of MOFs as Biosensing Materials for COVID-19
At present, the reverse transcription polymerase chain reaction (RT-PCR) remains the recommended laboratory diagnostic method for COVID-19.31,106 However, there are reports of a number of RT-PCR false-negative results on confirmed infection cases.91,125 It is a fact that a single negative PCR result does not rule out COVID-19 infection and the culture currently is two negative PCR results.31,106 In a recent study of 167 COVID-19 infection patients by Xie et al.,125 five patients with positive chest computed tomography (CT) received false-negative COVID-19 infection after RT-PCR test. Repeated swab tests on the five patients confirmed the patients as positive to COVID-19. More so, the current RT-PCR-based detection techniques require high manpower and long processing time.125 Other diagnostic techniques like the CT scan and culture methods are not suitable for fast-response detection and real-time analysis.134 Hence, it is imperative that scientists begin to think in other directions. Despite the roll-out of COVID-19 vaccines, the virus might not disappear with a brink of eyes. As earlier mentioned, MOFs possess outstanding properties that endeared them for the development of biosensors. Besides, there are proven cases where MOFs have been used in detection of viruses (HIV, Ebola, Zika, Dengue, etc.) and infectious diseases.58,92,121,130 The high specificity and selectivity of MOFs towards the detection of other viruses and diseases as presented in Section 4 should also be the motivation for the development of MOFs biosensors for COVID-19 detection. This could be a pointer to the huge potentials of MOFs in the fight against SARS outbreaks.
Advantages of MOFs as a Potential Biosensor for SARS-COV-2 Nucleic Acids, Antigens and Antibody Assay
Nucleic Acids Detection
MOF possesses a unique fluorescence quenching/recovery with continued decline of fluorescence intensity compared to the traditional fluorescence nanoquenchers such as gold nanotubes, graphenes, graphene oxides etc.42,59,77,89 This makes it promising in detection of SARS-CoV-2 viral RNA (nucleic acid). The fluorophore-labeled probes could be adsorbed on MOF through various interactions including electrostatic interactions, hydrogen bonding, and π–π stacking with negatively charged aromatic nucleic acid sequences similar to those presented in Sections 4.1, 4.2 and 4.3 for HIV-1 ssDNA, SUDV and HIV ss-DNA. The fluorescence of dyes on probes could be quenched by metal ions, such as Cu2+, Fe3+, Zn2+, Dy3+ or coplanar structure via the process of fluorescence resonance energy transfer (FRET).112,113 The specific hybridization of probe DNA (P-DNA) with target virus-related nucleic acids sequences could form stable rigid double or triple-stranded DNA structures and would be released from the surface of MOF due to their low affinity toward nanomaterials, leading to efficient recovery of fluorescence.26,41,112,113 The possible nucleic acid detection mechanism and quenching/recovery processes are depicted in Fig. 10. Again, the abundant functional groups and positively charged metal ions in MOF may provide various interactions, such as electrostatic interactions, hydrogen bonding, and π–π stacking for adsorption of fluorophore-labeled probes as presented in Fig. 11.
With MOF-based detection platform, many viruses such as Avian leukosis virus (ALV),66 Japanese encephalitis virus (JEV),129 HIV,25 hepatitis A virus (HAV) 73 have been effectively detected by antigen detection methods which are primarily based on the ELISA and molecular imprinting technology. Molecular imprinting technology is antigen detection technology that possesses remarkable advantages of brief and inexpensive preparation and prospective selectivity.113 The imprinted particles can capture the target virus quickly and show excellent selectivity for viruses because the imprinted sites are generated during imprinting.73 Although earlier approaches to viruses’ detection using molecular imprinting faced a challenge of large size viruses of 20–900 nm, this can be overcome by finding an imprinted carrier with sufficient surface area for providing additional imprinting sites. MOF is a typical material with high surface area, which is basically related to its porous structure. For example, MIL-100 (Cr) and MIL-101 (Cr) have a large pore of 2.5–2.9 and 2.9–3.4 nm, respectively. The larger surface area of the MOF can provide more imprinting sites for virus detection, which is beneficial to expand the linear range and improve sensitivity. With the large surface area advantage, other viral antigens detection methods, such as electrochemiluminescent immunosensing,74 immunochromatographic assay,68 sandwich-immunoassay LSPR chip46 may be adopted. Electrochemical immunoassay has many advantages, including good reproducibility, high sensitivity, low cost, fast and accurate analysis, which has attracted widespread attention in virus detection.113
Antibody Detection
MOFs-based fluorescence biosensor for antigen detection has a suitable adsorption capacity for molecular probes which is more conducive to fluorescence recovery compared to similar class of materials, such as single-walled carbon nanotubes (SWCNTs) and grapheme oxides (GO).114 Its application in SARS-COV-2 antibody detection will offer a high selectivity which is mainly based on the specific recognition of antibodies and antigens. Wei et al.114 developed a Cu-based novel biosensor MOF (H2dtoaCu) for the detection of influenza virus H5N1 antibody and the ss-DNA linked with H5N1 antigens and fluorescent dye 5′ 6-carboxyfluorescein were adsorbed by the MOF, resulting in the effective quenching of fluorescence making it easy to recover. This was because aromatic compounds (such as FAM) could be less strongly bound to MOF through non-covalent van der Waals interactions, making the fluorescence easier to be recovered compared to GO and SWCNT which is quite difficult. The MOF was proven to be advantageous in terms of sensitivity, rapidity, cracking and stable period’s detection, simplicity in operation and cost-effectiveness.114 This could be adopted in SARS-COV-2 or similar strands of virus. The composition of the MOFs such as metal ions, functional groups; and properties such as geometry, size, porosity and stability are important factors to consider when developing MOF based biosensors for diagnoses of SARS-COV-2 and similar viruses. Stock and Biswas102 and Wang et al.113 have presented detailed information on these factors.
Challenges Associated with Incorporating MOFs in Viral Diagnosis and Steps to Address Them in Future Researches
The major challenges associated with MOF-based viral detection and diagnostic strategies are the detection limit, instability of some MOFs in aqueous medium and complex physiological environments,64,109 biological toxicity of some metal ions in the MOFs and complex in vivo environment.14 The detection limit of MOFs ranged from pM to nM and is not low enough to detect a low level of viruses in the clinical samples, hence may affect sensitivity. To improve on this, future researches should consider the following (1) designing structures of MOFs with reduced particle sizes to amplify fluorescence signal as well as enhance high quenching efficiency and good selectivity; (2) integrating viral RNA in developing novel biosensors using nanoparticles, such as MOF-derived porous carbon41; (3) generating automated ultrasensitive MOFs-based biosensors with high-throughput screening in a large-scale, combining electrochemistry technologies21 and high class devices like lab-on-a-chip96; (4) combining MOF with metal oxides, metal particles, or carbon materials for electrochemical detection of viruses will be a promising direction since oxide is known to enhance binding ability, biological activity, and sensing performance.101To address the challenge of MOFs instability in aqueous medium and complex physiological environment, water-stable MOFs can be constructed with zwitterionic carboxylate ligands131 or zwitterionic thiolate56 and some metal ions including Zn2+110 and Zr4+13 Moreover, water-stable MOFs can be obtained by doping metal ions.145 For instance, small amounts of second metals such as Mn2+, Zn2+, Mg2+ and Ca2+ can be incorporated into metal clusters to enhance the stability of MIL-100 (Fe) moiety.3In the case of biological toxicity of some metal ions in the MOFs and in vivo environment, the detection performance and safety of MOF-based biosensing need to be further verified in vivo and some biocompatible metal ions (such as Fe2+/Fe3+, Ni3+, Zr4+, Mn2+, Mg2+, Ca2+, Zn2+), or biomaterials (such as liposomal bio-conjugates). Green synthetic routes using Sonochemistry should also be considered.27,75 MOFs functional modification with target ligands (viz: aptamers)12,38 enhance its functionalities, with detection performances much better than the biosensor platforms based on graphene oxide.26
Conclusions
The science of biosensors as diagnostic tools is highly promising. Embracing these tools can ensure life-saving decisions on the handling, understanding, and treatment of pandemic strains. As the world continues to be ravaged with COVID-19 pandemic and with the uncertainties surrounding the current COVID-19 diagnosis approaches, this review tends to draw the attention of researchers towards thinking in the direction of metal organic frameworks (MOFs) as a biosensor in COVID-19 pandemic outbreak. The application of MOFs as biosensors in pandemic outbreaks such as Human Immunodeficiency virus, Ebola virus, Zika virus, Dengue virus, and as biosensor for sensing Thrombin have been highlighted. The attractive features of MOFs that endeared them for biosensing applications include the large surface area, tailorable structure, high porosity, tunable size, and versatile functionality. Because COVID-19 pandemic may not disappear anytime soon despite the rolling out of vaccine, early testing and isolation would remain the safest approach and taking proactive steps to utilize MOFs biosensor may be a significant milestone. As a promising potential platform for detecting SARS-CoV-2, MOF has the following advantages (1) it can be used as an adsorption and fluorescence quenching platform; (2) MOF with unique structure can be designed using molecular imprinting technology; (3) MOFs can serve as a simple and effective fluorescence anisotropy amplification platform for SARS-COV-2 detection.