| Literature DB >> 34246285 |
Takuya Ishikawa1, Hiroyuki Sakai2, Tokutaro Itaya1, Suwanai Hirotsugu1, Jumpei Shikuma1, Takashi Miwa1, Ryo Suzuki1, Masato Odawara1.
Abstract
BACKGROUND: Anti-streptavidin antibodies are causal determinants of analytical interference during Thyroid function tests, and numerous reports have detailed such interference, with anti-streptavidin antibodies attracting attention. CASEEntities:
Keywords: Anti-streptavidin antibody; Immunoassay interference; Pre-wash; Thyroid function tests
Year: 2021 PMID: 34246285 PMCID: PMC8272342 DOI: 10.1186/s13044-021-00108-y
Source DB: PubMed Journal: Thyroid Res ISSN: 1756-6614
Detection and characterization of interference using polyethylene glycol, heterophilic blocking tube, and streptavidin-coated magnetic particle SA-MP treatment
| TSH, μIU/mL (RR: 0.50–5.00) | FT4, ng/dL (RR: 0.90–1.70) | FT3, pg/mL (RR: 2.3–4.0) | |
| Patient | |||
| Before PEG Tx | 4.67 | 2.66 | 5.88 |
| After PEG Tx (actual measured value) | 4.77 | 0.87 | 2.11 |
| After PEG Tx (calculated value) | 9.54 | ||
| TSH | FT4 | FT3 | |
| Recovery (%) | 204 (RR: 30–80) | 33 (RR: 60–90) | 36 (RR: 60–90) |
| TSH, μIU/mL (RR: 0.50–5.00) | FT4, ng/dL (RR: 0.90–1.70) | FT3, pg/mL (RR: 2.3–4.0) | |
| Patient | |||
| Before HBT Tx or SA-MP Tx | 4.36 | 2.15 | 4.27 |
| After HBT Tx | 6.49 | 1.68 | 3.86 |
| After SA-MP Tx | 9.81 | 1.04 | 2.7 |
| Control | |||
| Before HBT Tx and SA-MP Tx | 3.65 | 1.63 | 3.62 |
| After HBT Tx | 3.69 | 1.64 | 3.64 |
| After SA-MP Tx | 3.64 | 1.66 | 3.61 |
PEG Tx resulted in a high TSH level but low FT4 and FT3 recovery. HBT Tx or SA-MP Tx resulted in a higher TSH but lower FT4 and FT3 values. However, HBT Tx or SA-MP Tx showed almost the same results in the control sample
Thyroid function test parameters as measured by different immunoassays (Elecsys and Architect)
| Elecsys | Architect | ||
|---|---|---|---|
| Cobas8000 e602 (with pre-wash) | Cobas e411(without pre-wash) | ||
| Patient | |||
| TSH, μIU/mL | 4.36 (RR: 0.50–5.00) | 4.77 (RR: 0.50–5.00) | 8.57 (RR: 0.35–4.94) |
| FT4, ng/dL | 2.15 (RR: 0.90–1.70) | 1.31 (RR: 0.90–1.70) | 0.86 (RR: 0.70–1.48) |
| FT3, pg/mL | 4.27 (RR: 2.3–4.0) | 3.66 (RR: 2.3–4.0) | 2.44 (RR: 1.71–3.71) |
Cobas8000 e602, which is performed with pre-washing, resulted in almost the same level of TSH and with higher FT4 and FT3 values compared to Cobas e411, which is not performed with pre-wash. Architect provided the highest TSH level and lowest FT4 and FT3 values among the three instruments