Tobias Erlöv1, Rafi Sheikh2, Ulf Dahlstrand2, John Albinsson2, Malin Malmsjö2, Magnus Cinthio1.
Abstract
In translation from preclinical to clinical studies using photoacoustic imaging, motion artifacts represent a major issue. In this study the feasibility of an in-house algorithm, referred to as intensity phase tracking (IPT), for regional motion correction of in vivo human photoacoustic (PA) images was demonstrated. The algorithm converts intensity to phase-information and performs 2D phase-tracking on interleaved ultrasound images. The radial artery in eight healthy volunteers was imaged using an ultra-high frequency photoacoustic system. PA images were motion corrected and evaluated based on PA image similarities. Both controlled measurements using a computerized stepping motor and free-hand measurements were evaluated. The results of the controlled measurements show that the tracking corresponded to 97 ± 6% of the actual movement. Overall, the mean square error between PA images decreased by 52 ± 15% and by 43 ± 19% when correcting for controlled- and free-hand induced motions, respectively. The results show that the proposed algorithm could be used for motion correction in photoacoustic imaging in humans.Entities:
Year: 2021 PMID: 34221662 PMCID: PMC8221956 DOI: 10.1364/BOE.421644
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732