| Literature DB >> 34217352 |
Shaobai Wang1, Xiaolong Zeng2,3, Liang Huangfu1, Zhenyan Xie4, Limin Ma2, Wenhan Huang5, Yu Zhang6.
Abstract
BACKGROUND: The Opti_Knee system, a marker-based motion capture system, tracks and analyzes the 6 degrees of freedom (6DOF) motion of the knee joint. However, the validation of the accuracy of this gait system had not been previously reported. The objective of this study was to validate and the system. Two healthy subjects were recruited for the study.Entities:
Keywords: Biplanar fluoroscopy; Kinematics; Knee; Marker-based motion analysis
Year: 2021 PMID: 34217352 PMCID: PMC8254326 DOI: 10.1186/s13018-021-02576-2
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Fig. 1Experimental setup for gait analysis.
Fig. 2Scene of the setup of tibiofemoral coordinate systems. a This graph showing that the two marker sets were fixed to the thigh and shin. A digitizer was used to calibrate patient-specific bone landmark points (i.e., great trochanter (GT), medial epicondyle (ME), lateral epicondyle (LE), medial tibia plateau (MP), lateral tibia plateau (LP), medial malleolus (MM), and lateral malleolus (LM)) with the participants in a neutral standing position. b This graph exhibited the tibiofemeral coordinate systems. By the bony landmarkers, a tibiofemoral coordinate system was estabilshed [6]
Fig. 3The procedure of the experiment
Accuracy of the knee kinematics determined by the Opti_Knee gait system and biplanar fluoroscopy system in flexion–extension cycles
| Variables | RMSE | Mean differences | SD of differences | 95% limits of agreement | R | R square |
|---|---|---|---|---|---|---|
| Add/Abd (°) | 1.9 | − 0.1 | 2.2 | (− 4.4, 4.2) | 0.995 | 0.990 |
| Int/Ext rotation (°) | 3.8 | − 3.8 | 2.0 | (− 7.7,0.1) | 0.998 | 0.996 |
| Flexion/extension (°) | 1.7 | − 0.9 | 1.7 | (− 4.2, 2.4) | 0.999 | 0.998 |
| Ant/Pos translation (mm) | 3.3 | 0.9 | 3.7 | (− 6.2, 8.4) | 0.858 | 0.736 |
| Dis/Pro translation (mm) | 3.7 | 0.8 | 3.9 | (− 7.6, 8.4) | 0.994 | 0.988 |
| Med/Lat translation (mm) | 2.7 | 0 | 3.1 | (− 6.1, 6.1) | 0.911 | 0.830 |
Accuracy of knee kinematics determined by the Opti_Knee gait system and biplanar fluoroscopy system in gait cycles
| Variables | RMSE | Mean differences | SD of differences | 95% limits of agreement | R | R square |
|---|---|---|---|---|---|---|
| Add/Abd (°) | 2.3 | − 0.7 | 2.2 | (− 5.2, 3.8) | 0.979 | 0.958 |
| Int/Ext rotation (°) | 3.2 | 0.6 | 3.1 | (− 5.5, 6.7) | 0.964 | 0.929 |
| Flexion/extension (°) | 1.4 | − 0.1 | 1.4 | (− 2.8, 2.6) | 0.999 | 0.998 |
| Ant/Pos translation (mm) | 4.2 | − 0.1 | 4.2 | (− 8.3, 8.1) | 0.938 | 0.880 |
| Dis/Pro translation (mm) | 3.3 | − 1.6 | 2.8 | (− 7.1, 3.9) | 0.883 | 0.780 |
| Med/Lat translation (mm) | 3.2 | − 2.3 | 2.3 | (− 6.8, 2.2) | 0.912 | 0.832 |
Notes: RMSE root mean square error, SD standard deviation, Add/Abd abduction/adduction, Int/Ext internal/external, Flex/Ext flexion/extension, Ant/Pos anterior/posterior, Dis/Pos distal/proximal, Med/Lat medial/lateral