Prosthesis Position after TAVI with Balloon-Expandable SAPIEN 3 in Bicuspid Aortic Valves.

Philipp Breitbart1, Jan Minners1, Martin Czerny2, Manuel Hein1, Franz-Josef Neumann1, Philipp Ruile1.   

Abstract

BACKGROUND: Prior data suggest a correlation between the position of transcatheter heart valves (THV) and the occurrence of complications after transcatheter aortic valve implantation (TAVI) in patients with tricuspid aortic valves (TAV). However, data including a detailed analysis of prosthesis positioning in bicuspid aortic valves (BAV) are limited. Therefore, the purpose of this study was to investigate THV position after TAVI in BAV.
METHODS: We evaluated the THV position in 50 BAV and 50 TAV patients (all received the balloon-expandable Sapien 3 prosthesis) using fusion imaging of pre- and post-procedural computed tomography angiography. According to the manufacturers' recommendations, a low implantation position was defined as >30% of the prosthesis below the annulus.
RESULTS: THV position was appropriate in the majority of the patients within both groups (90.0% for BAV vs. 96.0% for TAV, p = 0.240). In BAV, we observed a more pronounced THV waist (7.4 ± 4.5% vs. 5.8 ± 3.0%, p = 0.043) and a lower average THV expansion (91.9 ± 12.2% vs. 95.5 ± 2.7% of nominal expansion, p = 0.044).
CONCLUSIONS: Accurate positioning in relation to the aortic annulus of the TAVI Sapien 3 prosthesis is possible in patients with BAV with results comparable to TAV. However, there is a more pronounced prosthesis waist and a lower average THV expansion in BAV.

Entities:  

Keywords:  TAVI; bicuspid aortic valves; computed tomography angiography; fusion imaging; prosthesis positioning; transcatheter aortic valve implantation

Year:  2021        PMID: 34207892     DOI: 10.3390/jcm10122561

Source DB:  PubMed          Journal:  J Clin Med        ISSN: 2077-0383            Impact factor:   4.241


  1 in total

1.  Patients With Bicuspid Aortic Stenosis Undergoing Transcatheter Aortic Valve Replacement: A Systematic Review and Meta-Analysis.

Authors:  Yi Zhang; Tian-Yuan Xiong; Yi-Ming Li; Yi-Jun Yao; Jing-Jing He; Hao-Ran Yang; Zhong-Kai Zhu; Fei Chen; Yuanweixiang Ou; Xi Wang; Qi Liu; Xi Li; Yi-Jian Li; Yan-Biao Liao; Fang-Yang Huang; Zhen-Gang Zhao; Qiao Li; Xin Wei; Yong Peng; Sen He; Jia-Fu Wei; Wen-Xia Zhou; Ming-Xia Zheng; Yun Bao; Xuan Zhou; Hong Tang; Wei Meng; Yuan Feng; Mao Chen