Literature DB >> 34191071

Feasibility and Significance of Dose Adaptation via Linear Couch Translations to Correct for Rotational Shifts During Frameless Brain Radiosurgery with the Gamma Knife Icon™.

Joey P Cheung1, Olivier Morin1, Steve E Braunstein1, Penny K Sneed1, Philip V Theodosopoulos2, Michael W McDermott2, Lijun Ma3.   

Abstract

OBJECTIVE: The present study aimed to examine the technical feasibility and effectiveness of adapting the radiation dose distributions with three-dimensional (3D) linear couch translations in contrast to full six-dimensional couch maneuvers to correct for rotational shifts during frameless radiosurgical treatment with the Gamma Knife Icon™ (Elekta AB; Stockholm, Sweden).
METHODS: The original magnetic resonance images used for radiosurgery treatment planning (15 targets) were digitally processed to simulate rotational shifts of ±1, ±2, ±3, ±5, and ±10 degrees in the transverse plane and imported back into Leksell GammaPlan® (Elekta AB), creating "uncorrected" treatment plans. In addition, geometrically optimized 3D translation shifts were consequently applied to each isocenter in all "uncorrected" treatment plans to account for systematically introduced rotational shifts and to produce "corrected" treatment plans. The differences in the dose distribution between the original treatment plans and the "uncorrected" and "corrected" treatment plans were calculated and compared at each rotational shift position.
RESULTS: The "uncorrected" treatment plans resulted in a significant deterioration in target coverage (by 8-72%) and selectivity (by 2-42%), with some targets being missed completely with rotations of ±3 or more degrees. In contrast, in all "corrected" treatment plans, the average decreases in target coverage and selectivity were only 1% (maximum values 4-5%).
CONCLUSION: Applications of 3D linear couch translations successfully overcome gross uncertainties in dose distributions caused by up to ±10 degrees of rotational shifts in a target. As a result, rapid dose adaptation with 3D couch translations is unique and effective for frameless radiosurgery with the Gamma Knife Icon™.

Keywords:  Cone beam CT; Dose adaptation; Frameless radiosurgery; Gamma Knife Icon™; Gamma Knife radiosurgery; Rotational shift; Stereotactic radiosurgery; Treatment accuracy

Year:  2021        PMID: 34191071     DOI: 10.1007/978-3-030-69217-9_16

Source DB:  PubMed          Journal:  Acta Neurochir Suppl        ISSN: 0065-1419


  1 in total

1.  A technique for achieving submillimeter accuracy of volume-staged stereotactic radiosurgery.

Authors:  Lijun Ma; Shannon Fogh; Nalin Gupta; Andrew Hwang; Dilini Pinnaduwage; Jean Nakamura; Igor Barani; Michael McDermott; Penny K Sneed; David A Larson; Arjun Sahgal
Journal:  J Radiosurg SBRT       Date:  2012
  1 in total