We investigated 18 pyrogenic reactions (PRs) that occurred between July 1 and 13, 1987, in 16 patients receiving long-term hemodialysis at one dialysis center in Illinois. We defined a case of PR as the onset of chills or fever (oral temperature, greater than or equal to 37.8 degrees C) in a patient who was afebrile and had no signs or symptoms of infection before a dialysis treatment. Pyrogenic reactions to dialysis sessions during the epidemic period (July 1 through 14) were associated with the use of a reused dialyzer (risk ratio, 6.2; 95% confidence interval, 0.8 to 45). The water used to rinse the dialyzers after use and to dilute the germicide for disinfecting the dialyzers contained high concentrations of endotoxins (greater than 6 ng/mL) and bacteria (greater than 10(4) colony forming units/mL). After reuse of dialyzers was discontinued at the center, the PR rate fell to pre-epidemic levels. We concluded that PRs were associated with reuse of dialyzers, probably contaminated with endotoxins. Active surveillance for PRs among patients undergoing long-term hemodialysis should be routine at all dialysis centers; an increase in PRs should prompt timely review of water treatment and dialyzer reprocessing systems.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3418872

Source DB:  PubMed          Journal:  JAMA        ISSN: 0098-7484            Impact factor:   56.272


  4 in total

1.  Effects of incubation time and temperature on microbiologic sampling procedures for hemodialysis fluids.

Authors:  M J Arduino; L A Bland; S M Aguero; M S Favero
Journal:  J Clin Microbiol       Date:  1991-07       Impact factor: 5.948

Review 2.  Hemodialysis and water quality.

Authors:  Angela D Coulliette; Matthew J Arduino
Journal:  Semin Dial       Date:  2013 Jul-Aug       Impact factor: 3.455

3.  The reuse of hemodialyzers: an assessment of safety and potential savings.

Authors:  E Baris; M McGregor
Journal:  CMAJ       Date:  1993-01-15       Impact factor: 8.262

Review 4.  Endotoxemia: methods of detection and clinical correlates.

Authors:  J C Hurley
Journal:  Clin Microbiol Rev       Date:  1995-04       Impact factor: 26.132

  4 in total