Gora Lô1,2, Anna Julienne Selbé Ndiaye2, Amina Sow2, Amina Fall1, Aissatou Sow2, Oumar Barry1, Ndèye Marième Diop1, Halimatou Diop Ndiaye2,3, Ndèye Coumba Touré Kâne2,4, Souleymane Mboup2.
Abstract
INTRODUCTION: quantification of hepatitis B virus DNA, a key element in the management of chronic hepatitis B, allows a more direct and reliable measurement of viral replication and monitoring of the virological response to therapy. Polymerase chain reaction (PCR) platforms performing this quantification and adaptable to intermediate laboratories have been developed. Thus, this study was conducted to evaluate the on-site performance of the AMPLIX® hepatitis B virus (HBV) real-time PCR technique in comparison with the COBAS AmpliPrep™ technique.Entities:
Keywords: Amplix®; Roche COBAS AmpliPrep/COBAS Taqman; deoxyribonucleic acid; hepatitis B virus
Mesh:
Substances:
Year: 2021 PMID: 34178213 PMCID: PMC8197065 DOI: 10.11604/pamj.2021.38.294.24819
Source DB: PubMed Journal: Pan Afr Med J
characteristics of two Hepatitis B virus (HBV) DNA quantification assays
| Assays | ||
|---|---|---|
| Variables | CAP/CTM v2.0 | Amplix |
| DNA extraction principle | Magnetic particle | Magnetic particle |
| Sample volume (μL) | 650 μl | 200 μl |
| Elution volume (μL) | 65 μl | 50 μl |
| Sample capacity per batch | 4-72 | 1-48 |
| Extraction runtime (min) | 2h 15 mn | 1h 30 mn |
| Amplification runtime (min) | 3h 30 mn | 1h 40 mn |
| Target HBV genome region | Precore and Core | X |
| Claimed low limit of detection (IU/mL) | 20 | 26 |
| Claimed dynamic range (IU/mL) | 2.00x101 to 1.70x108 | 2.6 0x101 to 1.3x108 |
comparison and difference of quantification results of the HBV DNA between Amplix and CAP/CTM technical
| Amplix | ||||||
|---|---|---|---|---|---|---|
| 4 | 4 | 0 | 0 | 8 | ||
| 3 | 11 | 3 | 1 | 18 | ||
| 0 | 0 | 0 | 3 | 3 | ||
| 0 | 0 | 1 | 12 | 13 | ||
| 7 | 15 | 4 | 16 | 42 | ||
difference of quantification results of the HBV DNA between Amplix and CAP/CTM technical
| Log10 IU /ml viral load difference | ||||
|---|---|---|---|---|
| Δ ≤ 0.5 | 0.5< Δ≤1 | Δ >1 | Total | |
| < 3,3 | 9 (64.3%) | 4 (28.6%) | 1(7.1%) | 14 |
| [3,3 - 4,30] | 1 (33.1 %) | 1 (33.1%) | 1 (33.3%) | 3 |
| [4,30 - 5,30] | 2 (28.6%) | 4 (57.1%) | 1(14.3) | 7 |
| > 5,30 | 3 (30%) | 4 (40%) | 3 (30%) | 10 |
Figure 1the Passing-Bablok equation
Figure 2the mean difference (Amplix - CAP/CTM) in Bland-Altman plot
performance criteria of Amplix Assay
| Mean (M) | 4.86 | |
| Repeatability (4 log IU/ml) | Standard deviation (ET) | 0.09 |
| CV (%) | 2.12 | |
| Mean (M) | 7.76 | |
| Repeatability (7 log IU/ml) | Standard deviation (ET) | 0.12 |
| CV (%) | 1.60 | |
| Mean (M) | 3.91 | |
| Reproducibility (4 log IU/ml) | Standard deviation (ET) | 0.26 |
| CV (%) | 6.65 | |
| Bias | -2.10 | |
| Accuracy | -0.08 | |
| Uncertainty | 0.92 | |
| Mean (M) | 7.22 | |
| Reproducibility (7 log IU/ml) | Standard deviation (ET) | 0.23 |
| CV (%) | 3.15 | |
| Bias | 3.20 | |
| Accuracy | 0.22 | |
| Uncertainty | 0.98 |
CV: coefficient of variation
Figure 3diagram of levey-jennings of concentration level 4 log10 IU/ml
Figure 4diagram of levey-jennings of concentration level 7 log10 IU/ml