Spontaneous Repositioning of a Dislodged Atrial Pacemaker Lead.

Yoshiyasu Aizawa1, Ryo Konno1, Akio Kawamura1.   

Abstract

Entities:  

Keywords:  complication; dislodgement; pacemaker; passive fixation lead

Mesh:

Year:  2021        PMID: 34176842      PMCID: PMC8810246          DOI: 10.2169/internalmedicine.7675-21

Source DB:  PubMed          Journal:  Intern Med        ISSN: 0918-2918            Impact factor:   1.271


× No keyword cloud information.
A 95-year-old woman was implanted with a dual chamber pacemaker for a symptomatic 2:1 atrio-ventricular block with a heart rate of 40 bpm. Passive fixation leads were inserted into the right ventricular apex and right atrial appendage (Picture A). The next day after implantation, the atrial lead became dislodged and migrated into the superior vena cava (Picture B, C, arrowheads). Taking the patient's age and the risk of infection associated with a re-operation into consideration, we choose to conservatively observe the patient's course instead of performing a reoperation. The pacemaker was programmed from the DDD to VVI. A pacemaker check, which was performed 1 month later at an outpatient clinic, demonstrated the normalization of the atrial lead parameter. A chest X-ray revealed that the atrial lead had repositioned itself into the right atrial appendage (Picture D, E). The pacemaker was then reprogrammed to DDD and thereafter no further abnormality was observed. The incidence of lead dislodgement has been reported to range from 1% to 2.69% (1). Although a dislodged pacemaker lead can be repositioned using a catheter (2), we herein presented a rare case in which the malpositioning of an atrial lead resolved spontaneously.
Picture.
The authors state that they have no Conflict of Interest (COI).
  2 in total

1.  Successful percutaneous repositioning of a dislodged pacemaker lead.

Authors:  Naoki Yoshida; Takumi Yamada; H Thomas McElderry
Journal:  Europace       Date:  2013-10-03       Impact factor: 5.214

Review 2.  Meta-analysis of the incidence of lead dislodgement with conventional and leadless pacemaker systems.

Authors:  Yan Wang; Wenbo Hou; Chao Zhou; Yuxia Yin; Shoutao Lu; Guang Liu; Cuihai Duan; Mingkun Cao; Maoquan Li; Egon Steen Toft; Hai-Jun Zhang
Journal:  Pacing Clin Electrophysiol       Date:  2018-08-27       Impact factor: 1.976

  2 in total
  1 in total

1.  A Trick for a Dislodged Atrial Pacemaker Lead.

Authors:  Morio Shoda
Journal:  Intern Med       Date:  2021-06-26       Impact factor: 1.271

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.