Important Differences Between Manufacturers When Transitioning From a Contemporary Cardiac Troponin Assay to a High-Sensitivity Cardiac Troponin Assay.

Peter A Kavsak1, Kevin Um1, Shawn Mondoux1, Guillaume Paré1, Craig Ainsworth1, Andrew Worster1.   

Abstract

Entities:  

Year:  2021        PMID: 34169266      PMCID: PMC8209364          DOI: 10.1016/j.cjco.2021.01.017

Source DB:  PubMed          Journal:  CJC Open        ISSN: 2589-790X


× No keyword cloud information.
To the Editor: Despite high-sensitivity cardiac troponin (hsTn) assays having superior analytical and clinical performance compared with contemporary cTn assays, there are still analytical issues that may necessitate that clinical laboratories institute additional testing and inform clinicians of errors when reporting hsTn results.1, 2, 3 We have recently identified that the Ortho hsTnI assay misclassifies patients with myocardial injury (∼10% false positives for injury at a hospital and cancer centre), compared with the Abbott hsTnI assay. It is unclear what the impact would be for hospital sites transitioning from a contemporary cTn assay to Ortho’s hsTnI assay and whether the increase in positivity would be evident at another hospital setting. Given this, we assessed the impact of transitioning from a contemporary cTnI assay (Siemens EXL cTnI normal < 0.06 ug/L) to the Ortho hsTnI (female normal: <10 ng/L; male normal: <14 ng/L) using the published 99th-percentile cutoffs on the percentage (%) of positive results at a community hospital with an emergency department (ED) (West Lincoln Memorial Hospital [∼60 beds]). Over the first 18 weeks after commencing Ortho hsTnI testing at this hospital, the % positive was higher (26%; 95% confidence interval [CI]: 25%–29%; n = 1914 total results; 87.0% results from ED in 2020) vs the Siemens cTnI assay in the corresponding timeframe in 2019 (13%; 95% CI: 11%-15%; n = 1628 total results, with 85.6% of results from ED). After the laboratory program instituted duplicate testing for the Ortho hsTnI assay (to mitigate analytical outliers) with any positive results reflexed for Abbott hsTnI, over 4 weeks, the Ortho hsTnI assay still yielded higher % positive results (29%; 95% CI: 24%-34%; n = 452 total results) compared with the Siemens cTnI (21%; 95% CI: 17%-26%; n = 417; total results in 2019; P = 0.02). The Abbott hsTnI assay yielded positivity estimates over these 4 weeks (23%; 95% CI: 19%-28%) similar to those from the Siemens cTnI assay (P = 0.55). The number of patients that were positive by Ortho hsTnI and negative by Abbott hsTnI over these 4 weeks was 15, or 6.4% (95% CI: 3.9%-10.4%) of the population with hsTnI measured. None of these 15 patients had a diagnosis of acute coronary syndrome with the discordant findings between Ortho hsTnI and Abbott hsTnI (Supplemental Table S1). Removal of the Ortho false positives and results with macrocomplexes (Fig. 1) yielded a higher correlation between Ortho hsTnI and Abbott hsTnI in this subgroup (rho = 0.94; 95% CI: 0.91-0.96; n = 91), with closer agreement to what has been observed in patients with symptoms suggestive of acute coronary syndrome.
Figure 1

Comparison between Ortho high-sensitivity cardiac troponin (hsTn)I (y-axis) and Abbott hsTnI (x-axis) for patients at a community hospital who are positive with the Ortho hsTnI assay (red and grey filled circles are samples from females and males, respectively, who are positive for Ortho [upper reference limit (URL = 99th percentile); female URL < 10 ng/L; male URL < 14 ng/L]) and negative for Abbott hsTnI (ie, concentrations < sex-specific URLs).

Comparison between Ortho high-sensitivity cardiac troponin (hsTn)I (y-axis) and Abbott hsTnI (x-axis) for patients at a community hospital who are positive with the Ortho hsTnI assay (red and grey filled circles are samples from females and males, respectively, who are positive for Ortho [upper reference limit (URL = 99th percentile); female URL < 10 ng/L; male URL < 14 ng/L]) and negative for Abbott hsTnI (ie, concentrations < sex-specific URLs). These data support suboptimal performance of Ortho hsTnI for the detection of myocardial injury in the community-hospital setting, with confirmation by another hsTnI assay helpful to prevent a misdiagnosis.
  4 in total

1.  Misclassification of Myocardial Injury by a High-Sensitivity Cardiac Troponin I Assay.

Authors:  Peter A Kavsak; Shawn Mondoux; Andrew Worster; Janet Martin; Vikas Tandon; Craig Ainsworth; P J Devereaux
Journal:  Can J Cardiol       Date:  2021-01-13       Impact factor: 5.223

2.  Implications of adjustment of high-sensitivity cardiac troponin T assay.

Authors:  Peter A Kavsak; Stephen A Hill; Matthew J McQueen; P J Devereaux
Journal:  Clin Chem       Date:  2013-03       Impact factor: 8.327

3.  Non-reproducible cardiac troponin results occurring with a particular reagent lot.

Authors:  Julien Favresse; Julie Cadrobbi; Christine Eucher; Kim Laffineur; Catherine Rosseels; Denis Pieters; Marc Elsen; Jeremie Gras
Journal:  Clin Chem Lab Med       Date:  2020-07-20       Impact factor: 3.694

4.  High-Sensitivity Cardiac Troponin I vs a Clinical Chemistry Score for Predicting All-Cause Mortality in an Emergency Department Population.

Authors:  Peter A Kavsak; Joshua O Cerasuolo; Dennis T Ko; Jinhui Ma; Jonathan Sherbino; Shawn E Mondoux; Richard Perez; Hsien Seow; Andrew Worster
Journal:  CJC Open       Date:  2020-03-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.