Literature DB >> 3416442

Simultaneous liquid-chromatographic determination of prednisone and prednisolone in plasma.

M H Cheng1, W Y Huang, A I Lipsey.   

Abstract

This high-performance liquid-chromatographic (HPLC) method for simultaneous determination of prednisone and its metabolite, prednisolone, in plasma is a modification of the method of Frey et al. (Clin Chem 1979;25:1944-7). Heparinized plasma (1.0 mL) with 0.1 mL of internal standard solution (11-deoxy-17-hydroxycorticosterone, 2 mg/L) is extracted with 7.0 mL of dichloromethane, then washed sequentially with 0.1 mol/L HCl, 0.1 mol/L NaOH, and deionized water, 2.0 mL each. The extract is evaporated and the residue reconstituted with 75 microL of mobile phase, methanol/H2O (40/60 by vol). Thirty microliters of this is injected onto a reversed-phase C6 column, which is eluted at 1.4 mL/min. Analytical recoveries of prednisone and prednisolone were 94-98% and 102-106%, respectively. Day-to-day precision (CV) was 3.8% for prednisone, 6.1% for prednisolone. We encountered no interference from the 21 other steroids and 25 drugs tested. This method is simple, accurate, and precise.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3416442

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  HPLC determination and clinical significance of serum prednisone in patients with nephrotic syndrome.

Authors:  Chun-Mei Chen; Yun-Cheng Xia; Xu-Guang Zhang; Can-Hui Peng; Fu-You Liu; You-Ming Peng; Lin Sun
Journal:  Int J Clin Exp Med       Date:  2014-12-15
  1 in total