Ten electrophoretic methods compared with a selected method for quantifying lactate dehydrogenase isoenzymes in serum.

G C Moses1, M L Ross, A R Henderson.   

Abstract

Using the Selected Method of McKenzie and Henderson (Selected Methods Clin Chem 1983;10:59-67) as a reference method, we compared the performance of 10 commercially available methods for determination of lactate dehydrogenase (LD, EC 1.1.1.27) isoenzymes. Results were expressed as percentage of total LD activity, as determined with two different types of densitometers shown to have an average difference less than 1.4% for each isoenzyme. All methods gave generally comparable results, as judged by Bland-Altman plots and correlation analyses. However, in general, estimates by the commercial methods for LD-1, LD-2, and LD-3 were lower, and for LD-4 and LD-5 were higher than with the Selected Method. The overall CV was less than 20% for all methods and isoenzymes, except for LD-4 and LD-5 by the Beckman Paragon, Helena LD-VIS, Gel LDH, Gel PC, and Iso Dot, Gelman LDH Isozyme, and Sebia Hydragel assays, for which it was greater than 20%. Overall, accuracy was best with the Helena Iso Dot and LD-VIS assays, followed by the Corning LD Flur assay; accuracy was poorest with the Gelman LDH Isozyme, Sebia Hydragel, and Beckman Paragon assays.

Mesh:

Substances:

Year:  1988        PMID: 3416440

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  1 in total

1.  High sensitive C-reactive protein as a systemic inflammatory marker and LDH-3 isoenzyme in chronic obstructive pulmonary disease.

Authors:  Anup N Nillawar; J S Bardapurkar; S J Bardapurkar
Journal:  Lung India       Date:  2012-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.