Michael Sadowski, Clark Houghtling, Sopheak Srun, Tim Carlson, Jason Hedrick, Andrew Porteous, Ken Gordon.
Abstract
The ethylene oxide (EO) product test of sterility (ToS) can be conducted to comply with ANSI/AAMI/ISO 11135:2014 for the generation of data to demonstrate the appropriateness of the biological indicator (BI) that is used to develop and qualify the EO sterilization process. Clause D.8.6 of 11135 provides an option to perform a sublethal EO process, followed by conducting a product ToS, performing sterility testing of BIs from the process challenge device, and comparing the test results. Certain limitations for the EO product ToS should be considered when conducting studies that feature the use of this test, in order to support compliance with this requirement. Limitations for any sterility test include sample size, testing frequency, detection sensitivity, and/or the potential for false-positive/false-negative results, each of which must be recognized and well understood in order to support compliance with the standard. In addition, the experimental design of any study featuring the use of a sterility test should be carefully developed to ensure the generation of scientifically sound results and conclusions to support the study objective. © Copyright AAMI 2021. Copying, networking, and distribution prohibited.Entities: