Exposure to radiofrequency electromagnetic fields: Comparison of exposimeters with a novel body-worn distributed meter.

Anke Huss1, Stefan Dongus2, Reza Aminzadeh3, Arno Thielens3, Matthias van den Bossche3, Patrick Van Torre4, René de Seze5, Elisabeth Cardis6, Marloes Eeftens2, Wout Joseph3, Roel Vermeulen7, Martin Röösli2.   

Abstract

BACKGROUND: Exposure to radiofrequency electromagnetic fields (RF-EMF) is often measured with personal exposimeters, but the accuracy of measurements can be hampered as carrying the devices on-body may result in body shielding. Further, the compact design may compromise the frequency selectivity of the sensor. The aim of this study was to compare measurements obtained using a multi-band body-worn distributed-exposimeter (BWDM) with two commercially available personal exposimeters (ExpoM-RF and EmeSpy 200) under real-life conditions.
METHODS: The BWDM measured power density in 10 frequency bands (800, 900, 1800, 2100, 2600 MHz, DECT 1900 MHz, WiFi 2.4 GHz; with separate uplink/downlink bands for 900, 1800 and 2100 MHz); using 20 separate antennas integrated in a vest and placed on diametrically opposite locations on the body, to minimize body-shielding. RF-EMF exposure data were collected from several microenvironments (e.g. shopping areas, train stations, outdoor rural/ urban residential environments, etc.) by walking around pre-defined areas/routes in Belgium, Spain, France, the Netherlands and Switzerland. Measurements were taken every 1-4 s with the BWDM in parallel with an ExpoM-RF and an EmeSpy 200 exposimeter. We calculated medians and interquartile ranges (IQRs) and compared difference, ratios and correlations of geometric mean RF-EMF exposure levels per microenvironment as measured with the exposimeters and the BWDM.
RESULTS: Across 267 microenvironments, medians and IQR of total BWDM measured RF-EMF exposure was 0.13 (0.05-0.33) mW/m2. Difference: IQR of exposimeters minus BWDM exposure levels was -0.011 (-0.049 to 0.0095) mW/m2 for the ExpoM-RF and -0.056 (-0.14 to -0.017) for the EmeSpy 200; ratios (exposimeter/BWDM) of total exposure had an IQR of 0.79 (0.55-1.1) for the ExpoM-RF and 0.29 (0.22-0.38) for the EmeSpy 200. Spearman correlations were 0.93 for the ExpoM-RF vs the BWDM and 0.96 for the EmeSpy 200 vs the BWDM. DISCUSSION AND
CONCLUSIONS: Results indicate that exposimeters worn on-body provide somewhat lower total RF-EMF exposure as compared to measurements conducted with the BWDM, in line with effects from body shielding. Ranking of exposure levels of microenvironments showed high correspondence between the different device types. Our results are informative for the interpretation of existing epidemiological research results.
Copyright © 2021 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Body-Worn Distributed Meter; Comparison exposimeters; EME SPY; ExpoM-RF; Measurements; Microenvironments; RF-EMF

Year:  2021        PMID: 34153890     DOI: 10.1016/j.envint.2021.106711

Source DB:  PubMed          Journal:  Environ Int        ISSN: 0160-4120            Impact factor:   9.621


  1 in total

Review 1.  Three Quarters of a Century of Research on RF Exposure Assessment and Dosimetry-What Have We Learned?

Authors:  Kenneth R Foster; Marvin C Ziskin; Quirino Balzano
Journal:  Int J Environ Res Public Health       Date:  2022-02-12       Impact factor: 3.390

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.