| Literature DB >> 34150799 |
Shan Jiang1,2, Zhongbiao Jiang1,2, Li-Hua Luo1,3, Kun Yu1,2, Yeyu Cai1, Xingzhi Xie1, Wei Zhao1, Weijun Situ1, Jun Liu1, Zhu Chen1,2.
Abstract
Background: To investigate the value of automatic positioning technology in improving the protection of radiographers in the relocatable CT room of a Fang Cang hospital during the outbreak of coronavirus disease 2019 (COVID-19).Entities:
Keywords: COVID-19; Fang Cang hospital; automatic positioning technology; health protection; relocatable CT
Year: 2021 PMID: 34150799 PMCID: PMC8212982 DOI: 10.3389/fmed.2021.659520
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Figure 1The structure of the relocatable CT system. (A) The outward appearance of the relocatable CT system. (B) The interior structure of the relocatable CT system. The figures were provided through the courtesy of Shanghai United Imaging Healthcare Co., Ltd.
Figure 2Camera-based intelligent auxiliary positioning system. (A) The working principle of the camera-based intelligent auxiliary positioning system. (B) The relocatable CT system with AI positioning technology in Wuchang Fang Cang Temporary Hospital at Hongshan Stadium in Wuchang district, Wuhan. The figures were provided through the courtesy of Shanghai United Imaging Healthcare Co., Ltd.
Figure 3Operation process of relocatable CT aided by automatic positioning technology. (A) The inner structure of the relocatable CT system. (B) The patients enter the scanning room from a separate door by themselves. (C) The examination bed moves to the ideal position automatically with the help of the camera-based intelligent auxiliary positioning system. (D) The patient's real-time image is transmitted to a computer screen in the controlling room. (E) The automatic positioning system determines the scan range automatically. (F) The technologist starts scanning.
Demographic characteristics of 269 patients treated in a Fang Cang hospital.
| Female | 166 (61.7%) |
| Male | 103 (38.3%) |
| All population | 48.35 ± 12.38 |
| Female group | 48.15 ± 12.84 |
| Male group | 48.66 ± 11.65 |
| ≤ 20 | 5 (2%) |
| 21–30 | 14 (5.2%) |
| 31–40 | 56 (20.8%) |
| 41–50 | 71 (26.4%) |
| 51–60 | 77 (29%) |
| 61–70 | 44 (16.5%) |
| >70 | 2 (0.1%) |
Imaging analysis of 700 scans conducted by re-locatable CT.
| Accurate position group | 642 (91.7%) |
| Artificial adjustment group | 58 (8.3%) |
| Accurate range group | 660 (94.3%) |
| Adjustment group | 40 (5.7%) |
| High quality | 700 (100%) |
| Poor quality | 0 |