Evaluation of a simple method for measuring blood cholesterol levels using non-laboratory observers.

J R Pearson1, L J Dusenbury, R Bakes-Martin, M Loverde, J Johnson, R L Byyny.   

Abstract

PURPOSE: Blood cholesterol is a good estimator of risk for coronary heart disease. Six criteria for a screening cholesterol measurement are: (1) high predictive value of elevated cholesterol compared with a "gold standard" cholesterol measurement; (2) high correlation with the national reference system for cholesterol; (3) acceptable results obtainable by volunteer observers; (4) results available quickly; (5) reproducible results; and (6) low cost. We hypothesized that cholesterol measurements performed on the Reflotron by volunteer nurses would meet these criteria. SUBJECTS AND METHODS: We recruited 345 subjects (160 men, 185 women, aged 20 to 84) and measured their blood cholesterol level by a method calibrated under the National Heart, Lung, and Blood Institute Lipid Standardization Program (Lab-chol) and on the Reflotron (R-chol method).
RESULTS: Comparison of the results showed good agreement (slope = 0.937; intercept = 0.27 mmol/liter). Comparison of the risk assignments made by the R-chol method against those made by the Lab-chol method, using the 1987 Adult Treatment Panel classification criteria, showed that the R-chol method has a sensitivity of 0.79 and a specificity of 0.98. Assuming a national prevalence of hypercholesterolemia of 25 percent, the positive predictive value is 0.92 and the negative predictive value is 0.93. The cost per screen was $1.25.
CONCLUSION: These data show that volunteer observers can obtain screening cholesterol results on the Reflotron that are accurate, reliable, quickly available at a low cost, and have high predictive value for hypercholesterolemia.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3414733     DOI: 10.1016/0002-9343(88)90589-x

Source DB:  PubMed          Journal:  Am J Med        ISSN: 0002-9343            Impact factor:   4.965


  4 in total

1.  Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.

Authors:  S Havas; R Bishop; L Koumjian; J Reisman; S Wozenski
Journal:  Am J Public Health       Date:  1992-03       Impact factor: 9.308

2.  Precision and accuracy of a portable blood analyzer system during cholesterol screening.

Authors:  P Greenland; N L Bowley; C A French; B Meiklejohn; S Gagliano; C E Sparks
Journal:  Am J Public Health       Date:  1990-02       Impact factor: 9.308

3.  Cardiovascular risk screening program in Australian community pharmacies.

Authors:  Gregory M Peterson; Kimbra D Fitzmaurice; Helen Kruup; Shane L Jackson; Rohan L Rasiah
Journal:  Pharm World Sci       Date:  2010-03-10

4.  Comparison of the use of a dry chemistry analyser in primary care in Norway and the United Kingdom.

Authors:  G Thue; S Sandberg; D G Bullock
Journal:  Br J Gen Pract       Date:  1993-01       Impact factor: 5.386

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.