Mohit Singhala1, Ryan Bell1, Brielle Cenci1, Youseph Yazdi1, Mark A Marzinke2, Derek M Fine2, Chirag R Parikh2.
Abstract
Entities:
Keywords: acute kidney injury; dialysis connector; disaster planning; kidney replacement therapy; three-dimensional printing
Year: 2021 PMID: 34127951 PMCID: PMC8189730 DOI: 10.1016/j.ekir.2021.05.026
Source DB: PubMed Journal: Kidney Int Rep ISSN: 2468-0249
Figure 1Components required for in-house dialysate production. (a) The complete system attached to a Fresenius 2008T hemodialysis machine. (b) The Fresenius Optiflux Dialyzer. (c) close-up of the connector attached to the dialyzer effluent port (left) and total parenteral nutrition (TPN) bag inlet (right). (d) The Baxter Exactamix TPN bag. (a) and (b) show the test connector with slots used for leak tests and visual observation of the mating surfaces (please see Supplementary Document, S3).
Stability of electrolytes in the dialysate at room temperature and 4°C
| Analyte concentrations | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Calcium, mg/dl | Chloride, mmol/l | Bicarbonate, mmol/l | Glucose, mg/dl | Potassium, mmol/l | Lactate, mmol/l | Magnesium, mg/dl | Sodium, mmol/l | |||||||||
| Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | Room Temp | 4°C | |
| Time point | ||||||||||||||||
| BL | 5.2 | 5.1 | 104 | 104 | 29.9 | 29.8 | 108 | 107 | 2.1 | 2.2 | 0 | 0 | 1.29 | 1.25 | 140 | 140 |
| ~6 h | 5.3 | 5.2 | 103 | 103 | 31.9 | 32.1 | 106 | 107 | 2.2 | 2.2 | 0 | 0 | 1.28 | 1.3 | 140 | 140 |
| ~9 h | 5.2 | 5.1 | 104 | 104 | 31.6 | 32.7 | 109 | 109 | 2.1 | 2.1 | 0 | 0 | 1.28 | 1.23 | 140 | 140 |
| ~12 h | 5.2 | 5.1 | 104 | 104 | 32.4 | 31.3 | 104 | 107 | 2.2 | 2.1 | 0 | 0 | 1.25 | 1.26 | 141 | 140 |
| ~24 h | 5 | 5 | 104 | 103 | 31.3 | 32.9 | 106 | 106 | 2.1 | 2.1 | 0 | 0 | 1.32 | 1.33 | 139 | 140 |
| Percent difference (BL–24 h), % | -3.8 | -2 | 0 | -1 | 4.7 | 10.4 | -1.9 | -0.9 | 0 | -4.5 | N/A | N/A | 2.3 | 6.4 | -0.7 | 0 |
BL, baseline; Temp, temperature; N/A, not applicable.
The hemodialysis machine dialysate settings were: sodium 140 mEq/l , potassium 2.0 mEq/l, calcium 2.5 mEq/l, and bicarbonate 30 mEq/l.