Literature DB >> 34120656

Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis.

Shabnam Dildar1, Sheharbano Imran2, Farah Naz3.   

Abstract

BACKGROUND AND
OBJECTIVE: High Performance Liquid Chromatography (HPLC) technique is considered as a gold standard for HbA1c analysis however all laboratories cannot adopt it due to certain limitations. Our aim was to compare Particle Enhanced Immunoturbidimetry (PEIT) method with High Performance Liquid Chromatography (HPLC) for HbA1c analysis.
METHOD: All blood samples were analyzed by HPLC assay on a Bio-Rad D-10 analyzer and PEIT on an Erba XL-200 analyzer. Precision studies were undertaken and Coefficient of Variation (%CV) calculated. Systemic Error (SE), Random Error (RE) and Total Error (TEcalc) were obtained. The Total Allowable Error (TEa) set by the National Glycohemoglobin Standardization Program (NGSP) for HbA1c is 6%.The acceptable evaluation method is where TEcalc is less than TEa.
RESULTS: The Precision studies were satisfactory with Coefficient of Variation (%CV) being less than 4% for both techniques. Mean HbA1c levels were slightly higher from HPLC than PEIT 9.07 ± 2.23% and 8.93 ± 2.10% respectively, although the difference was minimal. RE was 1.41%, TEcalc was 1.55%, which was less than TEa set by the NGSP. Both methods strongly correlated with the correlation coefficient (r) 0.9716, p < 0.0001.
CONCLUSION: Our study showed HbA1c analysis by PEIT technique is precise, accurate, rapid and convenient and can be employed as an alternative to HPLC technique in countries where cost is a major problem for diagnosis and treatment.

Entities:  

Keywords:  HbA1c; High performance liquid chromatography (HPLC); Particle enhanced immunoturbidimetric test (PEIT)

Year:  2021        PMID: 34120656     DOI: 10.1186/s40842-021-00123-w

Source DB:  PubMed          Journal:  Clin Diabetes Endocrinol        ISSN: 2055-8260


  2 in total

1.  A method evaluation decision chart (MEDx chart) for judging method performance.

Authors:  J O Westgard
Journal:  Clin Lab Sci       Date:  1995 Sep-Oct

2.  Diagnosis and classification of diabetes mellitus: highlights from the American Diabetes Association.

Authors:  W F Rayburn
Journal:  J Reprod Med       Date:  1997-09       Impact factor: 0.142

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.