Valentina Kutyifa1, Mary W Brown1, Christopher A Beck2, Scott McNitt1, Crystal Miller3, Karlene Cox3, Wojciech Zareba1, Spencer Z Rosero1, Marye J Gleva4, Jeanne E Poole5.
Abstract
BACKGROUND: Outcomes of patients with nonischemic cardiomyopathy and low ejection fraction implanted with an implantable cardioverter-defibrillator (ICD) or cardiac resynchronization therapy with a defibrillator (CRT-D), especially in contemporary, real-life cohorts, are not fully understood.Entities:
Keywords: Cardiac resynchronization therapy; Implantable cardioverter-defibrillator; Left ventricular dysfunction; Sex differences; Ventricular tachyarrhythmias
Year: 2020 PMID: 34113895 PMCID: PMC8183973 DOI: 10.1016/j.hroo.2020.11.002
Source DB: PubMed Journal: Heart Rhythm O2 ISSN: 2666-5018
Figure 1Screening, implantation, and enrollment in the BIO-LIBRA study.
Recommended home monitoring parameters
| Parameter | Recommended device settings |
|---|---|
| BIOTRONIK (BIOTRONIK SE & Co. KG, Berlin, Germany) Home Monitoring | ON |
| Capture control RA, RV, and LV leads | ON or ATM |
| IEGMs for therapy episodes | ON |
IEGMs = intracardiac electrograms; LV = left ventricular; RA = right atrial; RV = right ventricular.
Recommended detection and therapy settings
| Parameter | Detection | Therapy |
|---|---|---|
| In patients with no VT history | ||
| VF | 207 bpm, 30/40 intervals | ATP One-Shot, 1 burst of 8 pulses at 88% cycle length, then full output shocks |
| VT2 | 188 bpm, 30 intervals | ATP ≥1 burst of 8 pulses at 88% cycle length, 10 ms scan decrement, all shocks ON |
| VT1 | Monitor zone, 150 bpm | Monitor zone |
| In patients where VT cycle length is known | ||
| VF | 207 bpm, 24/30 intervals | ATP One-Shot, 1 burst of 8 pulses at 88% cycle length, then full output shocks |
| VT2 | 188 bpm (or 10-20 bpm < VT rate), 30 intervals | ATP ≥1 burst of 8 pulses at 88% cycle length, 10 ms scan decrement, all shocks ON |
| VT1 | Therapy at 10-20 bpm < VT rate or monitoring zone at 150 bpm | Monitor zone or Therapy as for VT2 (favoring more ATP) |
ATP = antitachycardia pacing; bpm = beats per minute; VT = ventricular tachycardia.
Recommended supraventricular tachyarrhythmia discrimination settings
| Parameter | Recommended settings |
|---|---|
| Single-chamber devices | |
| MorphMatch | ON |
| Onset | ON, 20% |
| Stability | ON, 40 ms |
| Sustained VT timer | OFF |
| DX, dual-chamber, and CRT devices | |
| SMART Detection | ON (default settings or adapted to known VT) |
CRT = cardiac resynchronization therapy; VT = ventricular tachycardia.
Figure 2Follow-up schedule in the BIO-LIBRA study.
Figure 3Organizational structure in BIO-LIBRA.