| Literature DB >> 3410948 |
T Rappaport1, K P Sawyer, I Nachamkin.
Abstract
To develop laboratory methods for the rapid identification of gram-positive cocci from blood cultures, several commercial immunological and biochemical tests for identifying staphylococci and streptococci from two different blood culture systems (Thiol/TSB, Difco Laboratories; BACTEC, Johnston Laboratories, Inc.) were evaluated and compared with conventional identification methods. A total of 44 cultures contained Staphylococcus aureus as determined by conventional methods. Commercial immunological methods (six tested) ranged in overall sensitivity from 38.6 (Staphyloslide; BBL Microbiology Systems) to 77.3% (Staphaurex; Wellcome Diagnostics). All methods tested had 100% specificity. A total of 30 isolates of group D streptococci were tested with immunological and biochemical identification systems. The overall sensitivity ranged from 14.2% (Phadebact Group D; Pharmacia Diagnostics) to 100% with Streptex (Wellcome; immunological) and Identicult A-E (Scott Laboratories, Inc.; biochemical). The results of this study suggest that some reagents can be used to provide rapid identification of gram-positive cocci from blood cultures 24 h earlier than standard methods.Entities:
Mesh:
Substances:
Year: 1988 PMID: 3410948 PMCID: PMC266604 DOI: 10.1128/jcm.26.7.1335-1338.1988
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948