| Literature DB >> 34108549 |
J Nguyen1, K Kesper2, G Kräling3, C Birk4, P Mross5, N Hofeditz4, J Höchst6, P Lampe6, A Penning6, B Leutenecker-Twelsiek7, C Schindler4, H Buchenauer8, D Geisel4, C Sommer4, R Henning9, P Wallot9, T Wiesmann9, B Beutel10, G Schneider8, E Castro-Camus11, M Koch4.
Abstract
The worldwide shortage of medical-grade ventilators is a well-known issue, that has become one of the central topics during the COVID-19 pandemic. Given that these machines are expensive and have long lead times, one approach is to vacate them for patients in critical conditions while patients with mild to moderate symptoms are treated with stripped-down ventilators. We propose a mass-producible solution that can create such ventilators with minimum effort. The central part is a module that can be attached to CPAP machines and repurpose them as low-pressure ventilators. Here, we describe the concept and first measurements which underline the potential of our solution. Our approach may serve as a starting point for open-access ventilator technologies.Entities:
Year: 2021 PMID: 34108549 PMCID: PMC8190155 DOI: 10.1038/s41598-021-91673-7
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Plots of the airway pressure, flow, ventilation volume as a function of time for the PIP values (a) 20 cmH2O and (b) 15 cmH2O. Each curve corresponds to a ventilator with a different CPAP machine model (referred to as u1, u2, u3, u4 and u5). All measurements are performed on an artificial lung. Respiratory rate: 20 bpm, I:E ratio: 1:2.
Figure 2Plots of the airway pressure, flow, ventilation volume as a function of time. (a) Ventilation data plots acquired for the I:E ratios 1:1 and 1:2 while the respiratory rate and the PIP value are set to 20 bpm and 20 cmH2O, respectively. (b) Ventilation data plots acquired for the respiratory rates 16 bpm, 20 bpm and 26 bpm while the I:E ratio is 1:2 and the PIP value is 20 cmH2O. All measurements are conducted on an artificial lung and the pressure is generated with the CPAP machine u1 (see Table 3).
List of CPAP machine models.
| Manufacturer | Model | Reference |
|---|---|---|
| Löwenstein Medical GmbH | Prisma SOFT | u1 |
| Löwenstein Medical GmbH | Prisma 25 S | u2 |
| Respironics Inc | REMstar C-Flex Int’l | u3 |
| DeVilbiss healthcare | 9055SE | u4 |
| Löwenstein Medical GmbH | Phönix 3 | u5 |
Figure 3Illustration of patient-triggered and patient-cycled ventilation. The plots show the measured airway pressure (blue) and flow (red) as a function of time in the event of spontaneous breathing during ventilation (from ~ 16 s). The patient-initiated start of the inhalation and exhalation phase is marked as ‘o’ and ‘x’, respectively.
Figure 4Plots of the airway pressure, flow, ventilation volume as a function of time for the CPAP machines (a) u1 and (b) u4 (see Table 3). The curves represent the ventilation performance after turning on the ventilator (0 h) and after running for 72 h without a break (72 h). All measurements are performed on an artificial lung. PIP: 20 cmH2O, Respiratory rate: 20 bpm, I:E ratio: 1:2.
Specifications for a minimally acceptable ventilator to treat COVID-19 patients with mild or moderate symptoms.
| Ventilation variable | Description | Values |
|---|---|---|
| Triggering mechanism | A method to initiate the inspiration phase | Time-triggering, flow-triggering |
| Cycling mechanism | A method to end the inspiration phase | Time-cycling, flow-cycling |
| Breath control variable | A parameter that describes the mechanism to assist the patient’s breathing | Pressure-control (PC) |
| Breath sequence | A pattern of mandatory and/or spontaneous breaths | Intermittent mandatory ventilation (IMV) |
| Targeting Scheme | A method used by the ventilator to achieve a specific ventilation pattern | Set-point (s) |
| Ventilation Mode | A term to describe the set of ventilation operations based on the selected breath control variable, breath sequence and targeting scheme | PC-IMVs,s |
| I:E ratio | The ratio of the inspiration (I) and the expiration (E) phase during ventilation | 1:1, 1:2 |
| Respiratory rate | The number of breaths per minute (bpm). Here, it is the number of mandatory breaths per minute. The number of spontaneous breaths per minute cannot be lower than the number of mandatory breaths per minute | 8–30 bpm |
| Positive end-expiratory pressure (PEEP) | The airway pressure at the end of the exhalation phase | At least 5 cmH2O |
| Peak inspiratory pressure | The airway pressure set at the CPAP machine | 15–20 cmH2O |
| Maximum airway pressure | The upper-pressure limit in the patient’s airways | 30 cmH2O |
Design requirements for CARL to ensure the patient’s safety during ventilation.
| Function | Description |
|---|---|
| Contamination protection | Protection of patient and medical equipment from pathogens |
| Constant monitoring (settings) | Visual display of airway pressure, tidal volume, and current ventilation settings |
| Constant monitoring (hardware and software) | Visual and acoustic signals for hardware and software failures during operation |
Figure 5Illustrations of the module CARL. (a) 3D model of the module CARL. (b) Assembled model of CARL.
Figure 6Schematic of a basic ventilator with a CPAP machine and the module CARL.
Figure 7Photo of the user interface of CARL. (a) Entire interface with a display of a flow curve. Ventilation parameters are highlighted in orange. (b) Display of the current ventilation parameters and ventilation quantities.
Figure 8Comparison of measured pressure (blue) and flow (red) data acquired from an external calibrated system ( “pnt”) and CARL (“carl”). All measurements are conducted on an artificial lung and the pressure is generated with the CPAP machine u1 (see Table 3). PIP: 20 cmH2O Respiratory rate: 20 bpm, I:E ratio: 1:2.