Application of SF-Cube 2.0 Technology in Platelet Count in Patients with EDTA-Dependent Pseudothrombocytopenia.

Zhi He, Guofang Shu, Huixia Lu.   

Abstract

BACKGROUND: To explore the application of SF-Cube 2.0 technology in platelet count in patients with EDTA-dependent pseudothrombocytopenia (EDTA-PTCP).
METHODS: Twenty-two out-patients and in-patients with EDTA-PTCP in our hospital were selected from February 1, 2020, to August 31, 2020. With the permission of each patient, blood samples of EDTA and sodium citrate anticoagulant tubes were collected. EDTA anticoagulant samples were tested by using the "complete blood cell count and white blood cell differential count"(CD) mode, and "complete blood cell count, white blood cell differential count, and reticulocyte count" (CDR) mode of the Mindray BC-6800Plus automated hematology analyzer, and the platelet counts were compared with the sample of sodium citrate anticoagulant from this patient.
RESULTS: As SF-Cube 2.0 technology (implemented by using the CDR mode of the Mindray BC-6800Plus) was used, the platelet count of EDTA-PTCP sample was consistent with that of sodium citrate anticoagulant, which was significantly higher than that of CD detection mode.
CONCLUSIONS: SF-Cube 2.0 technology can effectively correct the platelet counts in people with known or suspected EDTA-PTCP.

Entities:  

Mesh:

Substances:

Year:  2021        PMID: 34107626     DOI: 10.7754/Clin.Lab.2020.201021

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138