Camille Zimmer1, Alexandra Cassivi1, Celia C Baía2, Elizabeth Tilley3,4, Robert Bain5, Richard Johnston6, Caetano C Dorea1. 1. Department of Civil Engineering, University of Victoria, Victoria, BC, Canada. 2. Environment and Regional Development Graduate Program (PGDRA), Universidade Federal de Rondônia (UNIR), Porto Velho, Rondônia, Brazil. 3. Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland. 4. The Polytechnic, University of Malawi, Chichiri, Malawi. 5. Division of Data, Analysis, Planning and Monitoring, UNICEF, New York, NY, USA. 6. Department of Environment, Climate Change and Health (ECH), WHO, Geneva, Switzerland.
Abstract
To monitor safely managed drinking water services, an increasing number of countries have integrated water quality testing for Escherichia coli into nationally-representative household surveys such as the Multiple Indicator Cluster Surveys (MICS). However, plastic waste generated during such water quality testing programs, mostly through the use of pre-sterilized disposable materials, is non-negligible. The objective of this study was to evaluate several re-use protocols for disposable filter funnels used by the MICS water quality test kits. Decontamination and re-use protocols were assessed in centralized laboratory and decentralized field settings and neither yielded positive results. Re-use of 100 mL sterile funnels decontaminated with an alcohol wipe resulted in a higher incidence of false positive results (i.e., positive contamination when processing sterile water), both in the laboratory and field; therefore, a higher proportion of positives tests can be expected if these components are re-used. Further improvements to the decontamination technique and training are needed before material re-use can be reliably adopted. Autoclaving the funnels for re-use is feasible, provided that there is capacity to re-package and distribute funnels in a sterile manner.
To monitor safely managed drinking water services, an increasing number of countries have integrated water quality testing for Escherichia coli into nationally-representative household surveys such as the Multiple Indicator Cluster Surveys (MICS). However, plastic waste generated during such water quality testing programs, mostly through the use of pre-sterilized disposable materials, is non-negligible. The objective of this study was to evaluate several re-use protocols for disposable filter funnels used by the MICS water quality test kits. Decontamination and re-use protocols were assessed in centralized laboratory and decentralized field settings and neither yielded positive results. Re-use of 100 mL sterile funnels decontaminated with an alcohol wipe resulted in a higher incidence of false positive results (i.e., positive contamination when processing sterile water), both in the laboratory and field; therefore, a higher proportion of positives tests can be expected if these components are re-used. Further improvements to the decontamination technique and training are needed before material re-use can be reliably adopted. Autoclaving the funnels for re-use is feasible, provided that there is capacity to re-package and distribute funnels in a sterile manner.
In 2015, the United Nations’ Sustainable Development Goals (SDGs) were set out to
guide world development agendas to the year 2030. Under Goal 6, the mandate was
established to “ensure available and sustainable management of water and sanitation
for all.”[1] Specific to drinking water, Goal 6.1 was established to “achieve universal
and equitable access to safe and affordable drinking water for all.”[1] Obtaining reliable information on the safety of drinking water supplies is a
major undertaking on the national and global scale.[2] The World Health Organization (WHO) and the United Nations Children’s Fund
(UNICEF) jointly gather demographic, social and public health data in collaboration
with national governments under the Joint Monitoring Programme (JMP), including
information on water supply, sanitation, and hygiene, with the support and use of
UNICEF’s Multiple Indicator Cluster Surveys (MICS) and its water quality module,
alongside other nationwide household surveys.The MICS water quality testing questionnaire includes verbal questions regarding
access and availability of drinking water, and 2 in situ water quality tests carried
out by a trained enumerator: 1 from a glass of water in the household (i.e.,
point-of-use, PoU), and 1 from the source of the glass of water (i.e.,
point-of-collection, PoC).[3] The water quality tests are processed in or near the household under study
and normally uses the membrane filtration technique to enumerate E.
coli, an indicator of the presence of faecal contamination, in 100 mL
of water sample.[4] Such portable water quality test methods require somewhat large quantities of
pre-sterilized and pre-packaged disposable materials (depending on the survey size),
cost approximately USD $2.40 per test and generate an estimated 10 m3 of
plastics waste for surveys published in 2019 (calculations available in Supplemental Table S2). Aside from the high cost and generation of
plastics waste, single-use materials can present logistical difficulties in terms of
the transport and distribution before field work, as well as the return transport
and disposal. We hypothesized that some single-use materials were sufficiently
robust to be cleaned and re-used.Current guidance on reducing laboratory plastic waste generally assumes access to a
centralized laboratory, recycling program infrastructure, and/or high-cost items
such as an autoclave,[5,6]
thus reducing the potential for plastic waste reduction to be realized in
low-resource and portable applications such as the MICS water quality tests.
Therefore, the objective of this study was to evaluate different re-use protocols
for disposable filter funnels used by the MICS water quality test.
Methods
Overview
MICS water quality tests utilize the membrane filtration technique in order to
enumerate presumptive E. coli; a 100 mL water sample is passed
through a filter of 0.45 µm pore size via vacuum pressure, thus isolating
bacteria on the filter. The filter is removed and placed in a single-use petri
dish on top of a gel media containing nutrients that are selective for
E. coli (Compact Dry™ EC plates—Nissui Pharmaceutical Co.,
Ltd., Japan) and incubated for 24 hours at approximately 37°C. After incubation,
each blue colony, measured as a colony forming unit (CFU), is enumerated by
visual inspection and is assumed to have originated from a single bacterium.[4] Further details of the water quality sampling and testing are available elsewhere.[4] Two portable field kit designs were used in this study (Figure 1): a standard
field kit (SFK; Millipore, USA, see Supplemental Table S1) currently in use by MICS surveys,[4] and a new low-cost field kit (LCFK; glass base from VWR, USA, filter
support designed by UNICEF, and not commercially available) that has been
trialed during non-MICS surveys in Afghanistan[2] and water quality field research in Malawi.[7,8] An upcoming publication
will describe the LCFK in detail, including CAD files to enable others to mass
produce this single component.
Figure 1.
(a) Photo of SFK. (b) Photo of new LCFK. Labels: 1. 100 mL syringe and
PVC tube for vacuum application (a 150 mL syringe was used for this work
however the standard MICS catalog uses a 100 mL syringe); 2. Filtration
base (current or low-cost); 3. Funnel; 4. Filter support (current or
low-cost).
(a) Photo of SFK. (b) Photo of new LCFK. Labels: 1. 100 mL syringe and
PVC tube for vacuum application (a 150 mL syringe was used for this work
however the standard MICS catalog uses a 100 mL syringe); 2. Filtration
base (current or low-cost); 3. Funnel; 4. Filter support (current or
low-cost).Decontamination and re-use protocols in this study focussed on filter funnels
(item 3, Figure 1)
which are disposable and come pre-sterilized in packages of 30. The objective of
this study was addressed by assessing re-use protocols in centralized laboratory
and decentralized field settings.
Decontamination Techniques for Funnel Re-Use
To assess potential funnel re-use, 2 general decontamination approaches were
evaluated: alcohol wipes or autoclaving, selected on the basis described below.
Other potential decontamination approaches could include boiling or washing with
chlorinated water; however, these methods were not investigated in this study.
Laboratory and/or field evaluations were performed depending on the techniques
as summarized in Table
1. Based on the methods used, only 1 trial at a time could be
processed (i.e., duplicate or triplicate measurements were not possible).
However, we undertook multiple trials (minimum 30) for each testing environment
(single alcohol wipe, double alcohol wipe, autoclave).
Table 1.
Summary of decontamination techniques evaluated in this work.
Decontamination technique
Details
Testing environment (number of trials)
Single alcohol wipe
Wiping with a single 70% isopropyl alcohol wipe
Laboratory (39) and field (71)
Double alcohol wipe
Consecutive wiping with two 70% isopropyl alcohol wipes; air
dried in between wipes
Laboratory (30)
Autoclave
Autoclaving at 121°C and 17 PSI for 15 min
Laboratory (34)
Summary of decontamination techniques evaluated in this work.Wipes containing 70% isopropyl alcohol were investigated (150 mm by 170 mm size,
PDI Healthcare Inc., United Kingdom) because these products are standard for
MICS modules to decontaminate reusable components such as the filter support and
forceps during membrane filtration.[4] Other alcohol-based agents include ethyl alcohol (ethanol) and N-propanol
(1-propanol or propan-1-ol), however these are not used by MICS so were not
evaluated. The PDI alcohol wipes are no longer available from the manufacturer;
consequently, current MICS surveys use Medipal Sterile Alcohol wipes (200 mm by
125 mm size, CLH Healthcare, United Kingdom), which similarly contain 70%
isopropyl alcohol.When using the alcohol wipes, care was taken to wipe all water contact surfaces
of the funnel at least twice with each wipe, to not touch the inside of the
funnel with anything other than the wipe, and, after wiping, the cleaned funnel
was placed upside-down on the alcohol wipe. Any further contact time with 70%
alcohol wipes has been shown to have a negligible effect on bacterial reduction
on plastic surfaces.[9]For the double alcohol wipe technique, a second wipe was used following the use
of a first wipe that was discarded after use. Notably, the double alcohol wipe
tests were done as a laboratory verification of possible alternative funnel
re-use solutions following the relatively high occurrence of false positive
results observed during fieldwork (see “Results” section). To ensure that there
was no residual alcohol left over after decontaminating the funnels with the
alcohol wipe(s), care was taken that the funnel had fully dried by waiting
approximately 1 minute before use. In the lab, to minimize the likelihood of the
funnel becoming contaminated during cleaning, gloved hands were sprayed with a
70% alcohol solution (Commercial Alcohols, Canada) prior to each trial. During
the fieldwork, gloves were not used during testing, but hands were cleaned
immediately prior to each test with a 70% alcohol-based hand sanitizer gel
(Purell®, Canada).The autoclave method for decontamination was also investigated because field
water quality testing materials could be repurposed for laboratory use in the
host country once MICS activities are concluded. For such scenarios, the funnel
could be decontaminated and re-used for other bacteriological assays in a
centralized laboratory by, for example, a local water or health authority.
Laboratory Evaluation of Funnel Decontamination Efficacy
A laboratory evaluation of funnel decontamination efficacy was carried out using
the SFK and consisted of initially processing 2 test waters (Figure 2), namely: (1)
water spiked with E. coli to approximately
105 CFU/100 mL (high spike, HS); or (2) a sterile, “blank” water.
Funnel decontamination efficacy testing consisted of filtering a 100 mL sample
of HS test water through the SFK to ensure exposure of the filtration apparatus
to E. coli. The level of baseline funnel contamination due to
such exposure was estimated by processing 100 mL of sterile blank water samples
for subsequent E. coli enumeration (number of trials, N = 34,
Figure 2a).
Similarly, the different decontamination techniques (Table 1) were assessed by initially
processing 100 mL of HS test water. However, prior to filtering 100 mL of
sterile blank water samples, specific decontamination techniques were applied to
each funnel (Figure
2b).
Figure 2.
(a) Procedure for establishing a baseline number of E.
coli remaining in the funnel after funnel re-use with no
decontamination. (b) Procedure for assessing decontamination efficacy
using an alcohol wipe or autoclave in the laboratory. High spike (HS)
water contained approximately 105 CFU/100 mL E.
coli.
(a) Procedure for establishing a baseline number of E.
coli remaining in the funnel after funnel re-use with no
decontamination. (b) Procedure for assessing decontamination efficacy
using an alcohol wipe or autoclave in the laboratory. High spike (HS)
water contained approximately 105 CFU/100 mL E.
coli.After each decontamination technique was applied to the funnel, 2 tests (Figure 3) were conducted
to verify whether any decontamination residual (i.e., leftover alcohol after
wiping) could potentially compromise subsequent E. coli
enumeration. Firstly, 100 mL of water containing approximately
1 × 102 CFU/100 mL (enumeration spike, ES), was enumerated in
triplicate using new, sterile funnels, in order to provide a baseline
enumeration of the ES. Secondly, the ES baseline enumeration was compared to the
enumeration of 100 mL of the same batch of ESwater, using a different filter
funnel subject to 1 of the decontamination techniques.
Figure 3.
Procedure for the assessment of a residual effect of the alcohol wipe and
autoclave decontamination methods. Enumeration spike (ES) water
contained approximately 1 × 102 CFU/100 mL E.
coli.
Procedure for the assessment of a residual effect of the alcohol wipe and
autoclave decontamination methods. Enumeration spike (ES) water
contained approximately 1 × 102 CFU/100 mL E.
coli.The above evaluations were carried out in the laboratory by experienced
personnel; in addition to these tests, a laboratory evaluation of the double
alcohol wipe decontamination method was undertaken with a group of newly trained
University of Victoria undergraduate students (N = 16). First, each student
plated a standard blank test, using sterile blank water and a new sterile funnel
(Figure 4a).
Directly following this, each student filtered a 100 mL sample of ESwater,
cleaned the funnel using 2 alcohol wipes, then processed a 100 mL sample of
sterile blank water (Figure
4b). The evaluation protocol carried out by the students (Figure 4b) closely
follows that which was carried out by experienced laboratory personnel (Figure 2b), with the
exception that ESwater was used (as opposed to a HS water) to more closely
mimic potential field concentrations.
Figure 4.
Procedure for the assessment of decontamination effectiveness with a
group of newly trained students. (a) Procedure for establishing a
baseline number of E. coli remaining in the funnel
after funnel re-use with no decontamination. (b) Procedure for assessing
decontamination efficacy using an alcohol wipe or autoclave in the
laboratory. Enumeration spike (ES) water contained approximately
1 × 102 CFU/100 mL E. coli.
Procedure for the assessment of decontamination effectiveness with a
group of newly trained students. (a) Procedure for establishing a
baseline number of E. coli remaining in the funnel
after funnel re-use with no decontamination. (b) Procedure for assessing
decontamination efficacy using an alcohol wipe or autoclave in the
laboratory. Enumeration spike (ES) water contained approximately
1 × 102 CFU/100 mL E. coli.The HS and ES test water used during the laboratory evaluation of decontamination
protocols consisted of different E. coli spike concentrations
in a matrix of autoclaved isotonic quarter-strength Ringers solution (Oxoid
Ltd., England). E. coli was sourced from a
commercially-available probiotic Mutaflor® (Pharma-Zentrale GmbH,
Germany) which was incubated in Tryptic Soy Broth (TSB, Sigma-Aldrich, Germany),
stirred at approximately 500 rotations per minute at 37°C overnight prior to
testing (further probiotic E. coli details and characterization
are available elsewhere[10]).
Mechanical Fatigue Testing
Fatigue testing of funnels was carried out in the laboratory to determine the
number of times funnels could be decontaminated and re-used before leakage was
observed. Once 100 mL tap water was filtered through the LCFK, the funnels were
removed, decontaminated (alcohol wipes or autoclave) and air-dried before
repeating the procedure (the number of cycles for each funnel was recorded).
This procedure was carried out on 20 funnels for each the single-wipe alcohol
decontamination (not conducted for the 2-wipe protocol) and autoclave
sterilization methods until leakage between the funnel and the vacuum filter was
observed, or until 25 cycles had been carried out; whichever came first.
Field Piloting of Funnel Re-Use
The funnel re-use protocol was piloted during the water quality testing component
of a study with a household survey component conducted in April 2019 in Southern
Malawi. This study obtained ethical approval from the National Committee on
Research in the Social Sciences and Humanities (NCRSH) in Malawi (P.10/18/326)
and the Human Research Ethics Board at the University of Victoria (18-1129). The
survey consisted of a questionnaire and water quality test and was conducted by
6 enumerators who had undergone 5 days of training on questionnaires and in situ
water quality testing, of which approximately 2 days were spent solely on water
quality training. The training followed a similar structure to MICS Water
Quality Testing Modules (MICS, 2017) with E. coli sampling and
enumeration of two 100 mL samples: PoU and PoC. In total, 375 randomly selected
households were surveyed in informal settlements and rural areas over a 3-week
period. In a typical day, each enumerator surveyed 5 households, with a maximum
of 20 households surveyed by each enumerator each week.To undertake sample enumeration, the MICS protocol[4] was followed, with 2 exceptions: first, the LCFK was used (Figure 1b) to process all
samples; second, the funnels were decontaminated and re-used instead of 1 new
funnel per test as per the MICS protocol. Before use, funnels were cleaned with
1 alcohol wipe (the same wipe as used to clean forceps and funnel support
contact area), with care taken to not touch the inside of the funnel during
cleaning. Following testing, the funnels were dried with a towel and stored in
the enumerators’ backpack with the rest of the kit components. Each enumerator
was provided with 2 new funnels at the outset of each week (1 for main use and 1
as a backup), with each funnel being used for a maximum of 24 water quality
tests.The effectiveness of funnel decontamination (with a single alcohol wipe) was
evaluated in the field by processing a 100 mL sample of a freshly-opened bottled
of water “field blank” (Figure
5). During the field program, each enumerator conducted 1 field blank
test per day, or approximately 4 per week per enumerator, for a total of 71
blank tests. Field tests were conducted following the third household survey of
each day. Bottled water used for field blanks was tested beforehand by the
supervisors to ensure no E. coli was detected.
Figure 5.
Procedure for assessing decontamination effectiveness in the field.
Procedure for assessing decontamination effectiveness in the field.
Statistical Methods
Relevant geometric mean colony counts and upper/lower 95% Confidence Intervals
(CIs) were computed for all blank (laboratory and field) test data, using a
value of 0.5 CFU/100 mL, half of the minimum detectable E. coli
(1 CFU/100 mL), for non detects. Colony counts greater than
1 × 102 CFU/100 mL were recorded as 1.01 × 102 CFU/100 mL
in accordance with MICS protocols.[4]The proportion of positive blank tests were calculated for previous MICS surveys
and for laboratory and fieldwork. For field data, the proportion of positive
blank tests was computed as a function of the enumerator who conducted the test,
the day of the week and the week number (first, second, or third weeks).
Statistical comparisons were drawn using a binomial regression. Model outputs
were used as the input into a 1-way analysis of variance (ANOVA). Significant
ANOVA results were further analyzed using multiple comparisons with the post hoc
Bonferroni correction, to reduce the likelihood of a Type I error.For each decontamination technique, an ANOVA test was used to test for
differences in ES counts when new funnels were used, as compared to
decontaminating and re-using the funnels. A separate ANOVA test was undertaken
for the 1 alcohol wipe and autoclaving decontamination methods; ES data were
analyzed based on batch (i.e., blocking in 1-way ANOVA), so that comparisons
were only drawn on counts within each batch of ES. A Kruskal-Wallis rank sum
test (the non-parametric version of an ANOVA) was undertaken for the 2 alcohol
wipe decontamination method as the dataset violated the normality assumption.
All datasets used for an ANOVA were tested to verify the normality assumption
using a Shapiro-Wilk’s test.Results from all statistical tests used in the analysis were considered
significant at the α ⩽ 0.05 significance level. In the case of statistically
significant (P ⩽ .05) results, the data were examined further
to determine if the result would have an impact on the categorization of results
into such risk categories. All statistical tests were performed with R
statistical software, in RStudio, version 3.6.3. Water quality data were, in
some cases, compared to the a priori waterborne risk categories defined by Lloyd
et al[11]: “very low risk” (<1 CFU/100 mL); “low risk” (1-10 CFU/100 mL);
“moderate risk” (11-1 × 102 CFU/100 mL); “high risk”
(1.01 × 102-1 × 103 CFU/100 mL).
Results
Funnel Decontamination Efficacy and Field Effectiveness
The proportion of blank tests that returned a positive (⩾1 × CFU/100 mL) result
during laboratory and fieldwork are displayed in Figure 6. The proportion of positive
blank tests in the laboratory and fieldwork using 1 alcohol wipe to
decontaminate the funnels was found to be significantly higher than the
proportion reported in MICS surveys in 2019[12-19] (P = .02
and P < .01, respectively; Table 2). When funnels were
decontaminated using 2 alcohol wipes (either with HS or ESwater preceding the
blank) or the autoclave in the laboratory, the proportion of positive blank
tests did not differ significantly from that reported in MICS surveys from
2019[12-19]
(P = .07, .80, and .63, respectively, Table 2). The proportion f positive
blank tests in the laboratory did not differ when using 1 or 2 alcohol wipes
(either with HS or ESwater preceding the blank; P = .74 and
.28, respectively; Table
2). For more information on false positive rates reported in MICS
surveys conducted in 2019 (N = 4667), see Supplemental Table S2.
Figure 6.
Percent of blank tests (using sterile or mineral water for the laboratory
and field tests, respectively) returning a count of >1 CFU/100 mL;
Error bars represent the Bernoulli variance; “N,” denotes the number of
blank samples processed; MICS proportion calculated using a weighted
average by number of blank tests conducted in each survey (see Supplemental Table S2).
Table 2.
Summary of P-values for Binomial t-test
analysis of proportion of positive blank test, using all funnel
decontamination methods in laboratory and fieldwork.
MICS
One alcohol wipe; laboratory
Two alcohol wipes; laboratory
Two alcohol wipes with ES water;
laboratory
Autoclave; laboratory
One alcohol wipe; field
MICS
N/A
0.02[a]
0.07
0.80
0.63
<0.01[a]
One alcohol wipe; laboratory
N/A
0.74
0.28
0.07
<0.01[a]
Two alcohol wipes; laboratory
N/A
0.33
0.10
0.02[a]
Two alcohol wipes with ES water; laboratory
N/A
N/A[b]
0.02[a]
Autoclave; laboratory
N/A
<0.01[a]
One alcohol wipe; fieldwork
N/A
Significant result.
Not possible to compute; both rates of positive blanks 0%.
Percent of blank tests (using sterile or mineral water for the laboratory
and field tests, respectively) returning a count of >1 CFU/100 mL;
Error bars represent the Bernoulli variance; “N,” denotes the number of
blank samples processed; MICS proportion calculated using a weighted
average by number of blank tests conducted in each survey (see Supplemental Table S2).Summary of P-values for Binomial t-test
analysis of proportion of positive blank test, using all funnel
decontamination methods in laboratory and fieldwork.Significant result.Not possible to compute; both rates of positive blanks 0%.The summary of counts resulting from blank tests when the funnels were re-used
and decontaminated with the alcohol wipe(s) in the laboratory and field are
shown in Figure 7. The
geometric means for the laboratory blank test data were all <1 CFU/100 mL
(upper and lower 95% CI both <1 CFU/100 mL) using 1 or 2 alcohol wipes, or
the autoclave, to decontaminate the funnel. In the laboratory, when no
decontamination took place between processing the HS water and processing and
plating the blank water (Figure 2a), a baseline contamination average of 42 CFU/100 mL (95%
CI 37-47 CFU/100 mL) were enumerated in the blank water.
Figure 7.
(a) Distribution of counts resulting from blank tests in the laboratory
using 1 alcohol wipe to decontaminate funnel (N = 39). (b) Distribution
of counts resulting from blank tests in the laboratory using 2 alcohol
wipes to decontaminate funnel (N = 30). (c) Distribution of counts
resulting from blank tests in the fieldwork using 1 wipe to
decontaminate funnel (N = 71).
(a) Distribution of counts resulting from blank tests in the laboratory
using 1 alcohol wipe to decontaminate funnel (N = 39). (b) Distribution
of counts resulting from blank tests in the laboratory using 2 alcohol
wipes to decontaminate funnel (N = 30). (c) Distribution of counts
resulting from blank tests in the fieldwork using 1 wipe to
decontaminate funnel (N = 71).All field blank tests produced a geometric mean of <1 CFU/100 mL (95% CI
<1-2 CFU/100 mL), and exclusively positive field blank tests produced a
geometric mean of 3 CFU/100 mL (95% CI <1-7 CFU/100 mL). There was no
significant difference between the frequency of positive field blank test
results with respect to the day of the week it occurred, week number, or the
enumerator (excluding supervisors) who conducted the blank test
(P = .13, .09, .80, respectively), indicating that a
positive blank result was equally likely to happen to all enumerators during
fieldwork, regardless of the day and week. None of the field blank tests
conducted by the experienced counters, either during training or fieldwork,
produced positive results.In laboratory testing for the presence of a decontamination residual, there was
no statistical difference between the enumeration of ESwater when new funnels
were used, compared to decontaminating and reusing the funnels using any of the
decontamination methods examined. An ANOVA test comparing differences in ES
counts when new funnels were used, as compared to decontaminating and re-using
the funnels, yielded P-values of .99 and .07 for 1 alcohol wipe
and autoclaving decontamination methods, respectively. For the 2 alcohol wipe
decontamination method, a Kruskal-Wallis Wilcoxon rank sum test for differences
in ES counts when new funnels were used, as compared to decontaminating and
re-using the funnels, yielded a P-value of .30.Counts from positive field blank tests were compared to their respective
preceding water quality samples (Figure 8), and the correlation between
the 2 was not found to be significant (P = .50). Colony counts
greater than 1 × 102 CFU/100 mL were recorded as
1.01 × 102 CFU/100 mL; counts higher than this did not have an
impact on the correlation analysis.
Figure 8.
Correlation between positive field blank test counts and respective
preceding samples. A value of 1.01 × 101 CFU/100 mL was used
for counts >1.01 × 101 CFU/100 mL.
Correlation between positive field blank test counts and respective
preceding samples. A value of 1.01 × 101 CFU/100 mL was used
for counts >1.01 × 101 CFU/100 mL.No funnels leaked or showed signs of wear at any time during testing, regardless
of the decontamination type (either an alcohol wipe or autoclave), up to the
maximum of 25 uses (after which trails ceased). During field implementation of
funnel re-use with the new LCFK, no leakage was reported by the enumerators or
by the experienced counters. The funnels, which are intended for single use, are
robust enough for a minimum of 25 uses with decontamination by either alcohol
wipe or autoclave, confirming our hypothesis.
Discussion
General
Considering the laboratory evaluation, we found no additional benefit (in terms
of proportion of false positive blank results) to wiping the funnels down a
second time. Autoclaving funnels for re-use was only undertaken in the
laboratory; such a method could present logistical difficulties for fieldwork,
such as the need to carry used funnels for the remainder of the day or week
until they are returned to a central location for decontamination (as opposed to
disposing them), and the need for funnels to be re-packaged in sterile wrapping
for re-use. However, if the funnels are being re-used in a centralized
laboratory or if there is capacity to re-package and redistribute funnels in a
sterile manner, autoclaving funnels is a viable alternative to disposing of the
funnels after a single use. Low-cost alternatives to autoclaves can be used; for
example, the use of pressure cookers to sterilize reagents in low-resource areas.[20]There were noticeable differences between laboratory- and field-based
assessments, in terms of both the frequency and magnitude of positive blank test
results (Figures 6 and
7; Table 2), although
protocols in the laboratory and field were, in principle, the same and all
surfaces (including gloves or hands as appropriate) were cleaned with 70%
alcohol (liquid solution or hand sanitizer gel in the laboratory and field,
respectively) prior to testing. The laboratory and field contexts were
characterized by several main differences: where tests were performed (i.e., a
benchtop as opposed to a plastic board on the ground); the person conducting the
tests (i.e., experienced laboratory personnel vs newly trained enumerators); and
the composition of blank water and water preceding each blank (discussed below,
“Correlation of Positive Field Blank Tests with Preceding and Succeeding Water
Quality Data”).It should be noted that positive blank test results do not necessarily result
from re-using funnels. Positive blank tests can result from a number of factors
including, but not limited to, contaminated blank water, contaminated forceps or
filtration head contact area, accidental contact of membrane filter with
non-sterile surfaces or improper opening/closing of petri dishes. The funnel
decontamination and re-use in this study was the only methodological difference
to the standard MICS technique and there are 2 plausible ways in which funnels
could become contaminated during re-use: cross-contamination from the previous
sample, with residual E. coli not sufficiently decontaminated
by the alcohol wipe, or by hand contact with the inside of the funnel inside
during cleaning. Although hand cleanliness has never been evaluated in the
context of field-based membrane filtration techniques, during the MICS survey
the hand does not reach inside of the funnel.[4] It should be noted that some MICS surveys have reported
higher-than-average (1%) positive blank tests; notably, surveys conducted in
Cote d’Ivoire and The Gambia reported percentages of positive blank tests of
8.2% and 6.2%, respectively,[2] and such an investigation was not indicated by these surveys to discern
the contamination source. The comparison of false positive rates in this study
was made against the pooled results from MICS surveys (see Supplemental Table S2) and we did not examine heterogeneity in
false positive rates between surveys or survey teams.Occasionally, in the field, when funnels were decontaminated and re-used with an
alcohol wipe, there were pale yellow or colorless colonies present on the blank
test results, even if E. coli colonies were absent. Therefore,
it is unlikely the alcohol wipes achieved full “sterility”—full reduction of all
bacteria to numbers below detection minimum—but rather, in cases where
<1 CFU/100 mL was reported, the alcohol wipe reduced the E.
coli to numbers below the detection minimum (1 CFU/100 mL). This
may also be the case during MICS surveys, however as per the MICS methods,[4] only blue colonies (presumptive for E. coli) are counted
and recorded by enumerators, so any such occurrences have not been reported by
the surveys.
Log10 Reduction Achieved by Alcohol Wipe in the Laboratory
In the laboratory work, a “baseline” was established to estimate the number of
E. coli remaining on the funnels after processing the HS
water (Figure 2a). The
results for the established baseline (42 CFU/100 mL; 95% CI 37-47 CFU/100 mL)
indicated that, using either 1 or 2 alcohol wipes, a reduction of 1.6
log10 CFU/100 mL was achieved. It should be noted that this
reduction value is limited by the relatively low numbers of E.
coli present on the funnel following the filtration of HS water. It
is not possible to make such a comparison for the field data because no such
baseline was established to estimate the contamination remaining on the funnels
if no decontamination were to take place. The alcohol wipes used in this study
are intended for use in a healthcare setting[9,21] and a study examining
bacterial reduction by a similar disposable wipe on hospital computer keyboards
indicated that bacterial reduction of up to 99.99% (4 log10) is possible.[21]
Correlation of Positive Field Blank Tests with Preceding and Succeeding Water
Quality Data
Although it is important to minimize positive blank tests whenever possible, it
is not clear how best to address them when they do occur, either for specific
field teams or for an overall survey such as those performed for MICS. In this
study, we tried to assess whether there was any evidence to suggest that the
positive blank tests observed in the field were correlated with the preceding
blank tests (Figure 8),
for which we find no evidence. In light of such lack of evidence, it seems
plausible that the field positive blank results did not originate from the
preceding water, but perhaps from setting-related environmental contamination or
accidental contact of the funnel during the course of cleaning. Similar such
correlation between positive blank tests and the preceding water quality test is
not available from MICS survey findings reports published in 2019[12-19] (summarized in Supplemental Table S2).Such a correlation analysis was not conducted for laboratory results, however the
HS water preceding blank tests in the laboratory consistently had counts on the
order of 105 CFU/100 mL, several orders of magnitude higher than that
seen in the field. It therefore seems plausible that in the laboratory, the
positive blank test results may have originated from a residual contamination of
the preceding HS water, which was spiked to a concentration not seen in the
field.
Effect of Funnel Re-use on Categorical Risk Assessments
Typically, the E. coli enumeration data gathered in MICS surveys
are used to make risk assessments based on a priori waterborne risk categories
defined by Lloyd et al.[11] For water having a count of 1 or 2 CFU/100 mL lower than the cut-off for
the next highest risk category (eg, 9 CFU/100 mL or 1 × 102
CFU/100 mL), or for water in the very low risk category (<1 CFU/100mL), a
contaminated funnel could raise the risk assessment. Based on this, 2
hypothetical projections were conducted by subtracting from the non-blank counts
obtained during the fieldwork (i.e., PoC and PoU samples), to ascertain the
possible blank-adjusted risk categories of the water quality data.First, based on the result that a positive blank result was equally likely to
happen to all enumerators during all points in the field, we subtracted
1 CFU/100 mL, the geometric mean of all field blank test data, from the count of
all non-blank water quality data. The adjusted assumed risk category assessments[11] are presented in Table 3; under this analysis, 38 of 563 (6.7%) water quality samples
underwent a shift in risk category, with 33 of 126 (26%) low risk
(1-10 CFU/100 mL) shifted to very low risk (<1 CFU/100 mL). Water quality
data which are projected to have given a result of <1 CFU/100 mL (very low
risk category) are the most impacted by occurrences of positive blank results.
Under this analysis, all water quality data collected with a count of
1 CFU/100 mL (low risk) is assumed to have returned an adjusted count of
<1 CFU/100 mL (very low risk). Although some of our collected water quality
data may have returned their respective counts without any supposition of
contamination (i.e., not all water quality data collected with 1 CFU/100 mL were
contaminated), this projection demonstrates the impact of contamination on
samples in the very low or low risk categories.
Table 3.
Projected shift in risk assessment data of non-blank samples, according
to an estimated shift of −1 CFU/100 mL to every non-blank sample. Risk
categories defined according to Lloyd et al.[11]
Assumed blank-adjusted risk
category
Risk category according to water
quality data collected
Very low
Low
Moderate
High
Total
Very low
184
33
0
0
217
Low
0
93
5
0
98
Moderate
0
0
139
0
139
High
0
0
0
109
109
Total
184
126
144
109
563
Projected shift in risk assessment data of non-blank samples, according
to an estimated shift of −1 CFU/100 mL to every non-blank sample. Risk
categories defined according to Lloyd et al.[11]The second hypothetical projection was based on the result that 23% of the field
blank tests returned a positive result. In this projection we subtracted
3 CFU/100 mL, the geometric mean of the positive field blank tests, from a
randomly selected 23% of field non-blank water quality data. The resulting
blank-adjusted risk category assessments are presented in Table 4; under this analysis, 23 of
563 (4.1%) water quality samples underwent a shift in risk category, with 20 of
126 (16%) low risk (1-10 CFU/100 mL) shifted to very low risk
(<1 CFU/100 mL). Under this second analysis, low risk samples are not as
highly impacted as the first analysis, under which all samples having
1 CFU/100 mL were shifted into very low risk.
Table 4.
Projected shift in risk assessment data of non-blank samples, according
to an estimated shift of −3 CFU/100 mL to 23% randomly selected
non-blank samples. Risk categories defined according to Lloyd et al.[11]
Assumed blank-adjusted risk
category
Risk category according to water
quality data collected
Very low
Low
Moderate
High
Total
Very low
184
20
0
0
204
Low
0
106
3
0
109
Moderate
0
0
141
0
141
High
0
0
0
109
109
Total
184
126
144
109
563
Projected shift in risk assessment data of non-blank samples, according
to an estimated shift of −3 CFU/100 mL to 23% randomly selected
non-blank samples. Risk categories defined according to Lloyd et al.[11]A paired Wilcoxon rank-sum test was used to assess differences between the risk categories[11] of collected water quality data and each projection, revealing
statistically significant differences between the collected water quality data
and both the first and second projections depicted in Tables 3 and 4 (both comparisons
P < .01).
Future Implementation of Funnel Re-Use
There was no indication that any residual alcohol remaining on funnels following
decontamination by alcohol wipe or autoclave, and the funnels did not leak
during any testing either the laboratory or field up to a maximum 25 uses.
Therefore, the main barrier to implementing the practice of re-using funnels in
MICS surveys may be the observed incidences of positive blank test results.If further implemented, the proportion of positive blank test results would have
to be reduced via intensive training and/or refresher training during fieldwork.
Close supervision would need to be implemented to evaluate practices that may
lead to contamination of the funnel, membrane filter, or sample. Such a
reduction of false positive blank tests may be possible as none of the field
blank tests conducted by skilled counters either during training or fieldwork
produced positive results. In addition, we find promising the result of 0% false
positive results by undergraduate trainees in the laboratory when a double wipe
was used and with the previous sample having approximately
1 × 102 CFU/100 mL concentration (Figure 4). However, the sample size for
this finding was small (N = 16) and this testing would need to be repeated.
Alternate chemicals for surface decontamination, such as 70% ethyl alcohol (ethanol)[22] and chlorine-based, phenol-based and quaternary ammonium-based wipes[21] have been found to be comparable to 70% isopropyl alcohol (used in this
study) in the medical context. Based on these findings, it would appear to be
unlikely that alternative commercially available disinfectant wipes would
greatly improve results in this study.We note that other field-based membrane filtration techniques such as the DelAgua
kit utilize re-useable funnels made of stainless steel.[23] In the field, these funnels are decontaminated by sealing the top and
bottom of the funnel and burning methanol within the sealed space. When methanol
is burned in the low-oxygen space, formaldehyde gas is created, which acts as a
disinfectant, although is not as simple as using alcohol wipe(s). The DelAgua
kit has been recommended for portable water quality testing[24] and is commonly used to undertake field-based water quality
monitoring.[20,25-28]The generation of plastics waste by water quality testing conducted to monitor
progress toward SDG 6.1 is somewhat at odds with SDG goal 12.6, established to
“substantially reduce waste generation through prevention, reduction, recycling,
and reuse.”[29] Plastics waste generated by scientific research is non-negligible; the
University of Exeter bioscience department estimated their plastic waste
consumption to be approximately 247 tonnes in 201430—generated in
1 year by 1 department laboratory in 1 university. There have been calls to
reduce consumption of single-use plastic items in laboratory work,[30] for example by re-using items such as pipettes[5] or recycling nitrile gloves[6] if possible.To generate the MICS surveys published in 2019[12-19] (English language only),
31 354 household water quality questionnaires were conducted, for a total of
65 506 water quality tests, including blanks (see Supplemental Table S2). If funnel decontamination and re-use had
been implemented for all MICS surveys published in 2019, an estimated
10 m3 or 525 kg of plastic waste could have been saved and a cost
savings of approximately USD $1000 would have been realized, simply for funnels,
excluding all other disposable supplies required for MICS surveys such as petri
dishes (see Supplemental Table S2). Although the results of this study were
not positive, reducing the cost, logistical barriers and plastics waste via
decontamination, and re-use of water quality test kit components is a valuable
idea and is worth pursuing further.
Conclusion
The objective of this study was to evaluate different re-use protocols for the
funnels used in water quality testing. The proportion of positive blank tests during
both laboratory and fieldwork when alcohol wipes were used to decontaminate funnels
for re-use was significantly higher than those achieved in the MICS programs
published in 2019 (7% and 23% for laboratory and field respectively, compared to 1%
for MICS). Therefore, although it is worthwhile to study the idea further, we
recommend that funnels not be decontaminated via alcohol wipe for re-use in MICS
surveys unless blank test results can be improved. Given the reductions in cost and
waste, as well as the logistical difficulties presented in the transport,
distribution, and disposal of the funnels as they are currently used, we recommended
that further work be conducted to either find a way to re-use the current funnels,
or switch to using a funnel designed for re-use. Such work would support a
meaningful reduction in plastic waste by MICS surveys and would simultaneously
support SDGs 6.1 (drinking water for all) and 12.5 (reduce waste generation).
Autoclaving the funnels for re-use is feasible, provided that there is capacity to
re-package and redistribute funnels in a sterile manner.Click here for additional data file.Supplemental material, sj-pdf-1-ehi-10.1177_11786302211014400 for Assessment of
Decontamination and Reuse of Disposable Filter Funnels Used in Microbiological
Water Quality Tests by Camille Zimmer, Alexandra Cassivi, Celia C Baía,
Elizabeth Tilley, Robert Bain, Richard Johnston and Caetano C Dorea in
Environmental Health Insights
Authors: A H Parker; R Youlten; M Dillon; T Nussbaumer; R C Carter; S F Tyrrel; J Webster Journal: J Water Health Date: 2010-03-09 Impact factor: 1.744
Authors: Hussein Mohamed; Thomas Clasen; Robert Mussa Njee; Hamisi M Malebo; Stephen Mbuligwe; Joe Brown Journal: Trop Med Int Health Date: 2015-11-19 Impact factor: 2.622
Authors: Andrea E Berendt; LeeAnn Turnbull; Donald Spady; Robert Rennie; Sarah E Forgie Journal: Am J Infect Control Date: 2011-02-09 Impact factor: 2.918