Bilateral Bispectral Index Monitoring Performance in the Detection of Seizures in Nonanesthetized Epileptic Patients: An Observational Study.

Neus Fàbregas1, Jose F Valencia2, Isabel Belda1, Ana Tercero3, Adriana Hervias1, Santiago Villafuerte2, Pedro L Gambús1, Mar Carreño3, Ricard Valero1,4.   

Abstract

BACKGROUND: The aim of this observational study was to determine whether bilateral bispectral index (BIS) monitoring can detect seizures in epileptic patients.
METHODS: Four-channel frontal BIS monitoring and standard 40-channel electroencephalography monitoring were conducted in epileptic patients undergoing evaluation for epilepsy surgery. The BIS numerical value, signal quality index, electromyography, suppression ratio, and color density spectral array were continuously recorded. In patients with electroencephalography-confirmed seizures, the mean value and trend (slope of linear regression) of bilateral BIS monitor parameters were analyzed from 1 minute before to 1 minute after seizure onset.
RESULTS: Of 48 patients included in the study, 21 (43.8%) had at least 1 seizure. BIS numerical value was not able to detect focal or focal to bilateral tonic-clonic seizures. Considering all seizures, the only significant differences between recordings 1 minute before and 1 minute after seizure onset were a decrease in the signal quality index slope from 1 hemisphere (0.039±0.297 vs. -0.085±0.321, respectively; P =0.029) and in the mean signal quality index recorded from both hemispheres (left hemisphere: 65.775±30.599 vs. 61.032±26.285; P =0.016 and right hemisphere: 63.244±31.985 vs. 59.837±27.360; 0.029); these differences were not maintained after Hochberg adjustment for multiple comparisons. In seizures occurring during sleep, there was a change in the electromyography slope of 1 hemisphere before and after seizure onset (-0.141±0.176 vs. 0.162±0.140, respectively; P =0.038). There were variable responses in BIS parameters in the 3 patients who developed focal nonconvulsive seizure clusters.
CONCLUSION: Bilateral BIS monitoring was not able to detect the occurrence of seizures in epileptic patients.
Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2021        PMID: 34091549     DOI: 10.1097/ANA.0000000000000780

Source DB:  PubMed          Journal:  J Neurosurg Anesthesiol        ISSN: 0898-4921            Impact factor:   3.969


  1 in total

1.  Processed Electroencephalogram-Based Monitoring to Guide Sedation in Critically Ill Adult Patients: Recommendations from an International Expert Panel-Based Consensus.

Authors:  Frank A Rasulo; Philip Hopkins; Francisco A Lobo; Pierre Pandin; Basil Matta; Carla Carozzi; Stefano Romagnoli; Anthony Absalom; Rafael Badenes; Thomas Bleck; Anselmo Caricato; Jan Claassen; André Denault; Cristina Honorato; Saba Motta; Geert Meyfroidt; Finn Michael Radtke; Zaccaria Ricci; Chiara Robba; Fabio S Taccone; Paul Vespa; Ida Nardiello; Massimo Lamperti
Journal:  Neurocrit Care       Date:  2022-07-27       Impact factor: 3.532

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.