Survey of dental radiographic equipment and radiation doses in Finland.

R Havukainen1.   

Abstract

The radiation dose exposure, and the faults in about 1,700 dental units inspected at dental surgeries by the Finnish Centre for Radiation and Nuclear Safety in 1981-1985, were analysed. The mean value of skin doses in the bite-wing projection was about 6.2 mGy, the range 0.5 to 151 mGy. The mean energy imparted per bite-wing examination was estimated as 0.68 mJ and that per panoramic examination as 1.2 mJ. That gives a total imparted energy of about 600 J per year for conventional dental examinations and about 420 J per year for panoramic examinations. This gives a total of 0.13 mJ from conventional and 0.089 mJ from panoramic examinations per inhabitant per year. The collective effective dose equivalent was calculated as about 9 manSv for conventional dental examinations and about 6 manSv for panoramic examinations. Twenty per cent of units had some fault which was capable of decreasing radiation safety. Forty per cent of units were served reparation orders or other remarks were made in inspection documents. Large doses were usually accounted for by incorrect film processing and malfunction of the exposure timer.

Entities:  

Mesh:

Year:  1988        PMID: 3408611

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

1.  Compliance of Iranian dentists with safety standards of oral radiology.

Authors:  S Shahab; A Kavosi; H Nazarinia; S Mehralizadeh; M Mohammadpour; M Emami
Journal:  Dentomaxillofac Radiol       Date:  2012-02       Impact factor: 2.419

2.  Oral Radiology Safety Standards Adopted by the General Dentists Practicing in National Capital Region (NCR).

Authors:  Monica Chaudhry; K Jayaprakash; K K Shivalingesh; Vartika Agarwal; Bhuvandeep Gupta; Richa Anand; Abhinav Sharma; Sumedha Kushwaha
Journal:  J Clin Diagn Res       Date:  2016-01-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.