MRI Highly Accelerated Wave-CAIPI T1-SPACE versus Standard T1-SPACE to detect brain gadolinium-enhancing lesions at 3T.

Augusto Lio M Goncalves Filho1,2, M Gabriela Figueiro Longo1, John Conklin1, Stephen F Cauley1,2, Daniel Polak3, Wei Liu4, Daniel N Splitthoff3, Wei-Ching Lo5, John E Kirsch1, Kawin Setsompop1,2, Pamela W Schaefer1, Susie Y Huang1,2, Otto Rapalino1.   

Abstract

BACKGROUND AND
PURPOSE: High-resolution three-dimensional (3D) post-contrast imaging of the brain is essential for comprehensive evaluation of inflammatory, neoplastic, and neurovascular diseases of the brain. 3D T1-weighted spin-echo-based sequences offer increased sensitivity for the detection of enhancing lesions but are relatively prolonged examinations. We evaluated whether a highly accelerated Wave-controlled aliasing in parallel imaging (Wave-CAIPI) post-contrast 3D T1-sampling perfection with application-optimized contrasts using different flip angle evolutions (T1-SPACE) sequence (Wave-T1-SPACE) was noninferior to the standard high-resolution 3D T1-SPACE sequence for visualizing enhancing lesions with comparable diagnostic quality.
METHODS: One hundred and three consecutive patients were prospectively evaluated with a standard post-contrast 3D T1-SPACE sequence (acquisition time [TA] = 4 min 19 s) and an optimized Wave-CAIPI 3D T1-SPACE sequence (TA = 1 min 40 s) that was nearly three times faster than the standard sequence. Two blinded neuroradiologists performed a head-to-head comparison to evaluate the visualization of enhancing pathology, perception of artifacts, and overall diagnostic quality. A 15% margin was used to test whether post-contrast Wave-T1-SPACE was noninferior to standard T1-SPACE.
RESULTS: Wave-T1-SPACE was noninferior to standard T1-SPACE for delineating parenchymal and meningeal enhancing pathology (p < 0.01). Wave-T1-SPACE showed marginally higher background noise compared to the standard sequence and was noninferior in the overall diagnostic quality (p = 0.03).
CONCLUSIONS: Our findings show that Wave-T1-SPACE was noninferior to standard T1-SPACE for visualization of enhancing pathology and overall diagnostic quality with a three-fold reduction in acquisition time compared to the standard sequence. Wave-T1-SPACE may be used to accelerate 3D post-contrast T1-weighted spin-echo imaging without loss of clinically important information.
© 2021 American Society of Neuroimaging.

Entities:  

Keywords:  MRI; Wave-CAIPI; acceleration; brain; parallel imaging

Year:  2021        PMID: 34081374     DOI: 10.1111/jon.12893

Source DB:  PubMed          Journal:  J Neuroimaging        ISSN: 1051-2284            Impact factor:   2.486


  3 in total

1.  Usefulness of Wave-CAIPI for Postcontrast 3D T1-SPACE in the Evaluation of Brain Metastases.

Authors:  H J Baek; Y J Heo; D Kim; S Y Yun; J W Baek; H W Jeong; H J Choo; J Y Lee; S-I Oh
Journal:  AJNR Am J Neuroradiol       Date:  2022-05-26       Impact factor: 4.966

2.  Time optimization of gadobutrol-enhanced brain MRI for metastases and primary tumors using a dynamic contrast-enhanced imaging.

Authors:  Jeemin Seo; Changmok Lim; Kye Young Lee; Young-Cho Koh; Won-Jin Moon
Journal:  BMC Med Imaging       Date:  2022-10-17       Impact factor: 2.795

3.  Scan-specific artifact reduction in k-space (SPARK) neural networks synergize with physics-based reconstruction to accelerate MRI.

Authors:  Yamin Arefeen; Onur Beker; Jaejin Cho; Heng Yu; Elfar Adalsteinsson; Berkin Bilgic
Journal:  Magn Reson Med       Date:  2021-10-02       Impact factor: 4.668

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.