Heidi Thomas1, Felix Fries1, Max Gmelch1, Toni Bärschneider1, Martin Kroll1, Thaleia Vavaleskou2, Sebastian Reineke1. 1. Technische Universität Dresden, Dresden Integrated Center for Applied Physics and Photonic Materials (IAPP), Nöthnitzer Str. 61, 01187 Dresden, Germany. 2. Johann Wolfgang Goethe-Universität Frankfurt am Main, Institut für Anorganische und Analytische Chemie, Max-von-Laue-Straße 7, 60438 Frankfurt am Main, Germany.
Abstract
Currently, organic phosphorescent particles are heavily used in sensing and imaging. Up to now, most of these particles contain poisonous and/or expensive metal complexes. Environmentally friendly systems are therefore highly desired. A purely amorphous system consisting of poly(methyl methacrylate) particles with incorporated N,N,N',N'-tetrakis(4-carboxyphenyl)benzidine emitter molecules is presented in this work. Single particles with sizes between 400 and 840 nm show-depending on the environment-bright fluorescence and phosphorescence. The latter is observed when oxygen is not in the proximity of the emitting dye molecules. These particles can scavenge singlet oxygen, which is produced during the photoexcitation process, by incorporating it into the polymer matrix. This renders their use to be unharmful for the surrounding matter with possible application in marking schemes for living bodies.
Currently, organic phosphorescent particles are heavily used in sensing and imaging. Up to now, most of these particles contain poisonous and/or expensive metal complexes. Environmentally friendly systems are therefore highly desired. A purely amorphous system consisting of poly(methyl methacrylate) particles with incorporated N,N,N',N'-tetrakis(4-carboxyphenyl)benzidine emitter molecules is presented in this work. Single particles with sizes between 400 and 840 nm show-depending on the environment-bright fluorescence and phosphorescence. The latter is observed when oxygen is not in the proximity of the emitting dye molecules. These particles can scavenge singlet oxygen, which is produced during the photoexcitation process, by incorporating it into the polymer matrix. This renders their use to be unharmful for the surrounding matter with possible application in marking schemes for living bodies.
In the last years,
small organic phosphorescent particles have
been successfully used in a variety of devices for sensing and imaging.[1] They can be incorporated in chemical sensors[2−5] or light-emitting diodes,[6−8] used in O2 sensing,[9,10] for in vivo imaging,[11] or in phosphorescent
immunoassays.[12,13] Most of the organic particles
discussed in the literature contain metal complexes such as Pt(II),[14−16] Ir(III),[17−19] Eu(III),[20] or others.[21,22] Since the heavy metal effect induces an efficient spin–orbit
coupling, they have long been used to promote organic phosphorescence.[23,24] The aim for environmentally friendly alternatives, low cost, and
easy processability led to a huge effort in the preparation of small
purely organic particles showing room temperature phosphorescence
(RTP). Several strategies can be pursued, including H-Aggregates,[25−28] encapsulation,[29,30] polymerization,[31,32] and host-guest doping.[33,34] The big advantage of
RTP concerning in vivo imaging is the enhanced signal-to-noise ratio
compared to fluorescence since in the afterglow emission, no luminescent
background signal is present.[25] However,
many metal-free organic room temperature materials are nano- or microcrystals.
In contrast to this, purely amorphous systems are easy to process
and hence attract a lot of scientific interest.Here, we present
poly(methyl methacrylate) (PMMA) particles with
incorporated N,N,N′,N′-tetrakis
(4-carboxyphenyl)benzidine (TCATPB) emitter molecules. It has already
been shown that this material combination shows bright phosphorescence
in thin films and three-dimensional (3D) printed objects.[35] Downscaling such a fundamentally simple and
flexible system to particles smaller than 1 μm adds to the wide
application areas mentioned above. These particles do not contain
any harmful heavy elements and have the advantage over the already
mentioned techniques that they do not require high temperatures, expensive
educts, or complex workup. Depending on the oxygen content of the
environment, the fluorescence and phosphorescence of single particles
can be observed in a luminescence microscopy setup.
Results and Discussion
Preparation
of the Particles
PMMA:TCATPB particles
were formed using the so-called emulsion polymerization,[36] a type of radical polymerization. Methyl methacrylate
(MMA), ethylene glycol dimethacrylate (EGDMA, linker), and the emitter
were emulsified in water containing sodium dodecyl sulfate (SDS, surfactant)
and potassium persulfate (K2S2O8,
initiator) by vigorous stirring. Upon heating the solution, the particles
were formed. The polymer particles were separated by centrifugation
and washed three times with water to remove unbound monomers and residues
such as of the initiator. The particles were redispersed and stored
in water to prevent clogging. TCATPB was dissolved in cyclohexanol
since it is not soluble in MMA and/or EGDMA. The advantage of cyclohexanol
is that it does not dissolve the formed particles. Particles with
two different sizes were prepared. For the formation of the smaller
ones, no linker was used. The reason is that even though the increase
of the SDS concentration led to smaller particle sizes, as was expected,
the particles lacked phosphorescence. Polymerization using only the
monomer instead gave smaller particles that showed RTP. This may be
due to the better incorporation of the emitter into the PMMA particle.
Physical and Photophysical Properties of the Particles
Figure shows scanning
electron microscopy (SEM) images as well as size-distribution diagrams
of the respective particles having mean sizes of around 840 nm (P840) and 400 nm (P400).
Figure 1
SEM images (a, c) as
well as size-distribution diagrams (b, d)
of the larger P840 (blue) and smaller particles P400 (orange).
SEM images (a, c) as
well as size-distribution diagrams (b, d)
of the larger P840 (blue) and smaller particles P400 (orange).To measure the luminescent properties
of the particles, they were
applied onto a quartz substrate at a high concentration. This is achieved
by embedding the particles into an oxygen barrier. The barrier (Exceval,
a water-soluble ethylene vinyl alcohol polymer) does not emit itself
and allows measurements in air. To detect the phosphorescence, the
remaining oxygen within the particles has to be removed by activation
with UV light. The activation over time by continuous excitation is
shown in Figure S9 by the example of P840. A detailed description of this process has been published
recently.[37] As a reference system, a thin
layer of TCATPB in PMMA on a quartz substrate was used. To ensure
the same conditions as for the particles, the TCATPB sample has also
been covered with a thin barrier layer. Figure shows a comparison of the luminescent characteristics
of P840, P400, and the reference system. The
fluorescence maximum of P840 is located at 416 nm and does
not differ from TCATPB (Figure a), whereas the maximum of P400 shows an 8 nm blue
shift, which we attribute to random fluctuations. The phosphorescence
ratio is weakened in the microparticle samples compared to the thin-film
sample. It is assumed that the phosphorescence of a considerable amount
of emitters in the microparticle samples is quenched with very high
efficiency, leading to some of the emitters showing fluorescence only.
Thus, the total phosphorescence ratio is reduced. Presumably, this
is caused by the geometry of the two systems. In the microparticles,
a significant amount of emitter molecules is embedded at the particles’
surface, where they are prone to triplet-exciton quenching mediated
through molecular oxygen present at this very surface. In contrast,
the reference sample decouples all emitters from direct contact with
oxygen.
Figure 2
Thin-film emission data of TCATPB (gray) in PMMA covered with Exceval
as well as P840 (blue) and P400 (orange) embedded
in Exceval in the ambient atmosphere at an excitation wavelength of
λexc = 365 nm. (a) Prompt emission showing both fluorescence
and phosphorescence. The inset shows delayed spectra showing the phosphorescence
only. (b) Corresponding phosphorescence decays with τPhos = 575 ms (TCATPB), 585 ms (P840), and 610 ms (P400).
Thin-film emission data of TCATPB (gray) in PMMA covered with Exceval
as well as P840 (blue) and P400 (orange) embedded
in Exceval in the ambient atmosphere at an excitation wavelength of
λexc = 365 nm. (a) Prompt emission showing both fluorescence
and phosphorescence. The inset shows delayed spectra showing the phosphorescence
only. (b) Corresponding phosphorescence decays with τPhos = 575 ms (TCATPB), 585 ms (P840), and 610 ms (P400).Except for the signal-to-noise
ratio, the phosphorescence of both
kinds of particles resembles the spectrum of the reference (the inset
of Figure a). All
three samples show two peaks at around 516 and 550 nm. The triplet
energy is therefore not impacted by the size of the system, which
is in accordance with previous observations, suggesting that the triplet
energy is mainly affected by modifications of the emitter molecules
themselves.[38] The phosphorescence lifetime,
on the other hand, potentially depends on the surrounding environment.
Here, the lifetime slightly increases from 575 ms (TCATPB) to 585
ms (P840) and 610 ms (P400), with decreasing
system size. A reason might be the missing linker in the case of P400 and therefore a slightly larger amount of methacrylate
groups consequently exhibiting more hydroxy functions (by tautomerisation
of methyl ester groups) per emitter molecule. As has already been
reported,[34,39,40] hydrogen bonds
formed between the emitter and the host lead to an enhancement of
the phosphorescence. Nonetheless, the difference in the lifetime is
not drastic, which is as expected, as each particle still contains
plenty of emitter molecules (2.7 × 106 molecules per
particle for P840 and 2.9 × 105 molecules
per particle for P400). Compared to other recently published
organic phosphors,[41−43] this particle system compares well with a variety
of organic molecules showing RTP, where a typical lifetime value range
from a few hundred milliseconds up to single-digit seconds (Table ).
Table 1
Summary of the Emission (Fluorescence
and Phosphorescence) of the Different Targets: PMMA:TCATPB, P840, and P400 along with the Measured Fluorescence
and Phosphorescence Lifetimes, Quantum Yield, and Radiative and Nonradiative
Ratesa
sample
λfluo [nm]
λphos [nm]
τfluo [ns]
τphos [ms]
kr_P [s–1]c
knr_P [s–1]c
Φfluo+phos [%]
TCATPB
416
516/550
2.0
575/740b
1.35
0.39
36.2
P840
416
516/548
1.8
585/760b
1.32
0.39
32.9
P400
408
516/550
1.8
610/810b
1.23
0.40
33.2
Thin film covered
with and particles
embedded in Exceval. All measurements performed under air at room
temperature.
Measured at
77K.
See the Supporting Information for details.
Thin film covered
with and particles
embedded in Exceval. All measurements performed under air at room
temperature.Measured at
77K.See the Supporting Information for details.
Phosphorescence Imaging of the Particles
Using a fluorescence
microscope, the particles were investigated with respect to their
individual luminescence properties. Especially, the switching between
purely fluorescent emission in air and additional phosphorescence
in nitrogen needs to be confirmed to prove that phosphorescence is
visible on the single-particle level. Highly diluted samples were
prepared on microscopy glasses, which were placed on the microscope
stage. A closed sample box on top of the table allows measuring under
nitrogen conditions. Several different regions on the sample have
been found, containing some agglomerated particles, but also isolated
individual ones. Figure shows that single particles are nicely identifiable.
Figure 3
Images of the particles
as visible in the microscope. It can be
seen that both agglomerated particles and single particles are present
in the film.
Images of the particles
as visible in the microscope. It can be
seen that both agglomerated particles and single particles are present
in the film.Interestingly, the optical properties
between single particles
and agglomerates do not differ strongly, which is a hint that agglomeration
effects do not play a major role in our samples. The recorded image
is spectrally divided into two channels: the first (Channel 1) is
the wavelength regime of the fluorescence (below 450 nm) and the second
(Channel 2) contains both the residual fluorescence and the phosphorescence
(above 550 nm). The time behavior of a particle’s emission
can be determined using a series of pictures, where the cycle time,
which is needed for one picture, determines the time resolution. The
particle of interest is chosen in both the left and the right channel
and the intensity of this spot is evaluated for the whole series of
pictures. To isolate the phosphorescent contribution in Channel 2,
the residual fluorescence needs to be quantified. Under ambient conditions
(e.g., at the initial time t = 0), Channel 2 only
contains fluorescent emission. Now, the signal of Channel 1 can be
scaled to this intensity and subsequently subtracted for the whole
series of pictures. The residual signal is the remaining phosphorescence
(for more details, see Figure S7). If no
phosphorescence was present, this procedure would result in a constant
zero intensity in Channel 2. For the particles, a clear switch-on
process of the phosphorescence is visible, as can be seen in Figure , where the nitrogen
supply was switched on right after the start of the detection. Hence,
the phosphorescence signal increases steadily. Additionally, the fluorescence
decreases, which is due to singlet–triplet annihilation and
a saturation effect.[44] The latter is based
on the lifetime of the phosphorescence, which has been determined
to be around 600 ms, whereas the fluorescent lifetime is in the nanosecond
range. Therefore, on the timescale of excitation and fluorescence,
molecules that have undergone ISC to the triplet state are no longer
available for further excitation, resulting in a lowered fluorescence
signal. The timescale of the phosphorescence increase is determined
by the nitrogen flow, the gas tightness of the sample box, and the
permeability of the matrix.
Figure 4
Right after having started the measurement,
the nitrogen supply
was opened to create an inert atmosphere (a) P840 and (b)
P400. Consequently, the phosphorescence increases, while
the fluorescence decreases. Each measurement point is averaged over
10 images with an exposure time of ∼0.1 s, showing the behavior
of one exemplary particle each.
Right after having started the measurement,
the nitrogen supply
was opened to create an inert atmosphere (a) P840 and (b)
P400. Consequently, the phosphorescence increases, while
the fluorescence decreases. Each measurement point is averaged over
10 images with an exposure time of ∼0.1 s, showing the behavior
of one exemplary particle each.This process is reversible, which means as soon as the nitrogen
supply is stopped and oxygen penetrates back into the sample, the
signal in the phosphorescent channel drops and only starts to rise
again as soon as nitrogen is applied again (cf. Figure ). The variations in the heights of the phosphorescent
plateaus are due to the consequent scaling using the fluorescent signal.
During the measurement, the emitter molecules and possibly the embedding
polymers are subject to photodegradation (cf. Figure S4), influencing the fluorescence-to-phosphorescence
ratio and thus changing the absolute height of the plateaus. Possible
reasons are the formation of new molecular species with additional
luminescence contributions to the overall signal and the increase
of exciton quenching channels as a result of chemical decomposition
of the embedding material.
Figure 5
Appearance of phosphorescent emission of (a)
P840 and
(b) P400 is reversible, as can be seen in the switching
behavior. The green area shows the time slot where nitrogen was switched
on. Degradation leads to the fact that the initial signal level is
not restored. Each measurement point is averaged over 10 images with
an exposure time of ∼0.1 s.
Appearance of phosphorescent emission of (a)
P840 and
(b) P400 is reversible, as can be seen in the switching
behavior. The green area shows the time slot where nitrogen was switched
on. Degradation leads to the fact that the initial signal level is
not restored. Each measurement point is averaged over 10 images with
an exposure time of ∼0.1 s.The advantage of this particle system is that compared to other
particles used in, e.g., tumor recognition, the singlet oxygen that
is produced during irradiation does not harm the environment but is
scavenged by PMMA.[45] So, in contrast to
particles that help to destroy malignant cells, our particles might
be used in bodies without harm after coverage with an oxygen barrier.
Besides, they do not contain any toxic or costly metal compounds.
To prevent the surrounding from getting damaged by UV light, RTP materials
that absorb in the visible range, like PhenDpa[38] or BF2(HPhN) (ref (5)), could be used.
Conclusions
To
summarize, we produced PMMA particles containing an organic
emitter showing bright room temperature phosphorescence using a simple
preparation route based on methyl methacrylate and TCATPB. Depending
on the usage of a linker, the particles have a diameter between 840
and 400 nm. Upon illumination with 365 nm UV light, they can be detected
and visualized using a single-molecule spectroscopy setup. The fluorescent
particles possess long-lived room temperature phosphorescence, which
is activated in an inert nitrogen atmosphere or even in water in a
sealed environment (see Figure ) as soon as the oxygen is consumed. Exposure to the ambient
atmosphere lets the phosphorescence disappear.
Figure 6
Photographs of a dispersion
of P400 in water at (a)
ambient light and (b, c) illumination with a 365 nm LED. (b) The prompt
emission and (c) the delayed emission.
Photographs of a dispersion
of P400 in water at (a)
ambient light and (b, c) illumination with a 365 nm LED. (b) The prompt
emission and (c) the delayed emission.Overall, the switching between the two luminescence states is reversible.
Since the PMMA matrix scavenges the singlet oxygen that is produced
by illumination, these particles might be used in living bodies for
in vivo imaging.
Experimental Section
Materials
TCATPB
was synthesized from aniline and methyl-4-bromobenzoate
by double Buchwald–Hartwig palladium coupling followed by oxidative
coupling and saponification. A detailed procedure has been published
recently.[35]Methyl methacrylate (stabilized
with hydroquinone monomethylether) was purchased from Sigma-Aldrich
Chemical Co. and ethylene glycol dimethacrylate as well as sodium
dodecyl sulfate from TCI chemicals. These chemicals were used without
further purification.
Particle Preparation
P840: 7.5 mg of K2S2O8 in 2.5 mL of
1 μM SDS added
to 2.5 μL of EGDMA, 247.5 μL of MAA, and 200 μL
of 10 mg/mL TCATPB in cyclohexanol, and the reaction was stirred at
75 °C for 4 h.P400: 2.5 mg of K2S2O8 in 2.5 mL of 1 μM SDS added to 250
μL of MAA and 200 μL of 10 mg/mL TCATPB in cyclohexanol,
and the reaction was stirred at 75 °C for 4 h.
Sample Preparation
For the bilayer reference system,
164 μL of a 17 mM solution of TCATPB in anisole was added to
a 1 mL solution of PMMA (80 mg/mL) also dissolved in anisole. Exceval
(50 mg/mL) was dissolved in pure water at 120 °C in a closed
vial. For spin coating, a speed of 2000 rpm and volumes of 150 μL
(host and emitter) and 500 μL (pure barrier) were used to form
a uniform film. The layers were coated on top of each other. Cleaned
quartz glass substrates of a size of 25 mm × 25 mm were used.
For P840 and P400 samples, 50 μL of the
respective aqueous particle solution was added to 500 μL of
a 50 mg/mL aqueous Exceval solution. The complete mixture was drop
cast onto the substrate and dried at ambient conditions. For fluorescence/phosphorescence
spectroscopy, P840 and P400 suspensions were
diluted until optical clearness. Afterward, 500 μL of these
solutions were drop cast onto quartz substrates and dried at ambient
conditions.
Spectral Measurements
Direct and
delayed emission measurements
were performed using CAS 140CTS from Instrument Systems. To trigger
the detection and the 300 nm (Thorlabs, M300L4) LED, a TGP3122 pulse
generator (AIM-TTI Instruments) was used. The delayed spectra were
recorded in the time span from a few to a few hundred milliseconds
after turning off the LED. All measurements were performed in darkness
under ambient conditions.
Lifetime Measurements
The phosphorescent
lifetime was
determined using a silicon photodetector PDA100A by Thorlabs. The
decay was recorded and fitted using a two-exponential fit, yielding
an intensity-averaged lifetime. The fluorescent lifetime was determined
using time-correlated single-photon counting (TCSPC). The measurements
were performed with an excitation laser PicoQuant LDHDC375, a PicoQuant
PMA Hybrid-40 detector, a PicoQuant TimeHarp platine, and a monochromator
SpectraPro HRS-300 from Princeton Instruments. To extract the lifetimes
from the nanosecond decays, the reconvolution software FluoFit from
PicoQuant was used.
Scanning Electron Microscopy (SEM)
SEM characterization
was performed with an Amray 1920 ECO SEM (SEMTech Solutions, Inc.,
Billerica, MA), operating at 10 keV, equipped with a scintillation
detector (Everhart–Thornley). The particle suspensions were
drop cast onto a sample holder equipped with a conductive sticker
and left in air to dry for around one day. Afterward, the samples
were sputtered with gold, using an Edwards S1508 sputter coater with
an Ar pressure of 1.5 × 10–1 mbar for 60 s.
Fluorescence/Phosphorescence Microscopy
Fluorescence
microscopy experiments were performed using a Nikon Ti2-A inverted
microscope. The samples are placed on the microscope’s table,
which was applied with a custom-made box to allow measurements under
inert conditions. The excitation light was from a 373 nm collimated
free-beam laser diode (LDH-D-C-375, PicoQuant), passing a clean-up
filter (370/36 BrightLine HC, Semrock) and a lambda fourth plate (355
nm, Edmund Optics). The beam was expanded using a 10× UV beam
expander (BE10-UVB, Thorlabs, Inc.) and then focused on the back-focal
plane of the objective to enable far-field microscopy. It entered
the microscope through the backside port and was mirrored to the sample
stage via a dichroic mirror (zt 375 RDC, Chroma). Emitted light from
the sample was collected by the objective and passed the dichroic
mirror to be led to a side port of the microscope. Here, it was spectrally
separated into two parts using color filters (FESH0450 and FELH0500,
Thorlabs) and a dichroic mirror (zt 514 RDC, Chroma) mounted on an
Optosplit II (Acal BFi Germany GmbH). The two resulting images represented
the wavelength regimes. The image detection was done using a back-illuminated
CCD camera (iXon Ultra 897, Andor). Time-resolved measurements were
realized by taking a series of images and subsequent post-procession
of the data with a self-written evaluation script.
Authors: Pablo Llano Suárez; Marta García-Cortés; María Teresa Fernández-Argüelles; Jorge Ruiz Encinar; Marta Valledor; Francisco Javier Ferrero; Juan Carlos Campo; José Manuel Costa-Fernández Journal: Anal Chim Acta Date: 2018-08-09 Impact factor: 6.558
Authors: E G Vlakh; E V Grachova; D D Zhukovsky; A V Hubina; A S Mikhailova; J R Shakirova; V V Sharoyko; S P Tunik; T B Tennikova Journal: Sci Rep Date: 2017-02-03 Impact factor: 4.379