Nowadays, heavy metal pollution has attracted wide attention. Many electrochemical methods have been developed to detect heavy metal ions. The electrode surface usually needs to be modified, and the process is complicated. Herein, we demonstrate the fabrication of electrodes by direct laser sintering on commercial polymer films. The prepared porous carbon electrodes can be used directly without any modification. The electrodes were fixed in a 3D-printed flow reactor, which led to very little analyte required during the detection process. The velocities of the analyte under stirring and flowing conditions were simulated numerically. The results prove that flow detection is more conducive to improving detection sensitivity. The limit of detection is about 0.0330 mg/L for Pb2+. Moreover, the electrode has been proved to have good repeatability and stability.
Nowadays, heavy metal pollution has attracted wide attention. Many electrochemical methods have been developed to detect heavy metal ions. The electrode surface usually needs to be modified, and the process is complicated. Herein, we demonstrate the fabrication of electrodes by direct laser sintering on commercial polymer films. The prepared porous carbon electrodes can be used directly without any modification. The electrodes were fixed in a 3D-printed flow reactor, which led to very little analyte required during the detection process. The velocities of the analyte under stirring and flowing conditions were simulated numerically. The results prove that flow detection is more conducive to improving detection sensitivity. The limit of detection is about 0.0330 mg/L for Pb2+. Moreover, the electrode has been proved to have good repeatability and stability.
With the rapid development
of economy and technology, the pollution
of heavy metals such as lead ions has caused widespread concern.[1] Pb2+ is a kind of toxic heavy metal
ion that is harmful to human health and can damage the nervous and
digestive systems. Lead poisoning can cause intellectual and developmental
problems.[2,3] For human health, the detection and analysis
of heavy metal ions are critical. At present, the detection methods
of lead ions mainly include absorption photometry, atomic absorption
spectroscopy, atomic fluorescence spectrometry, inductively coupled
plasma atomic emission spectrometry, X-ray fluorescence spectrometry,
and mass spectrometry.[4−7] Although these methods are very sensitive, they have some disadvantages,
such as expensive equipment, complex preprocessing, long-time consumption,
and poor real-time detection.Compared with other detection
methods, electrochemical detection
has the advantages of fast analysis speed, high sensitivity, simple
operation, low detection cost, and no requirement of expensive large-scale
instruments. It is widely used in the detection of heavy metal ions.[8−11] The most commonly used electrochemical detection technique is voltammetry,
including cyclic voltammetry (CV), linear scanning voltammetry (LSV),
anode stripping voltammetry (ASV), square wave voltammetry (SWV),
and differential pulse voltammetry (DPV).[12−14] In ion trace
analysis, the most common detection method is stripping voltammetry,
which combines the constant potential analysis with the voltammetric
analysis.[15] The analysis method has two
steps. The first step is to make the analyte perform constant potential
electrolysis in the stirred state, which will make the analyte be
reduced and deposited on the cathode. This process is called the enrichment
process. The second step is to apply a reverse scanning voltage to
the electrode after a period of quiescence so that the analyte deposited
on the cathode is oxidized and redissolved in the solution. This process
is called the stripping process.[16,17] Compared with
the direct voltammetric analysis of the original solution, the biggest
advantage of this method is that the analyte can be pre-enriched on
the electrode, so the dissolution current is not affected by the charging
current and the impurity residual current.Metal oxides, expensive
metals, and carbon are often used as electrodes
for detecting heavy metal ions.[18,19] But metal oxide has
low conductivity. Noble metals are scarce, and their high price limits
their widespread use. However, as one of the most common elements
in nature, carbon has good biocompatibility and is a kind of a low-consumption
material,[20,21] so it is often used to make electrodes.
Traditional carbon materials are activated carbon, graphite, carbon
nanotubes, and graphene. To improve material performance, researchers
have introduced the concept of porosity into the construction of carbon
materials. Porous carbon has many advantages, such as a large specific
surface area, good conductivity, and many active sites. It is a very
suitable electrode material.[22]Herein,
porous carbon electrodes were prepared by CO2 infrared
laser sintering on commercial polymer films. The process
is simple to operate and is a new manufacturing technology, which
significantly improves production efficiency and reduces costs. Tour et al. used this technique to make polymer-written electronic
and energy storage devices.[23] The application
of this technology in microsupercapacitors has been widely reported,[24−26] but the application in detection is scarce. In this work, we use
this technology to prepare electrochemical sensors for the detection
of lead ions.In the electrochemical detection process, lead
ions were first
deposited on the electrode surface and then dissolved by DPV.[27−29] The critical step to detect heavy metal ions is the ion deposition
process. High deposition efficiency will improve the sensitivity of
detection. The traditional method to improve the ion deposition efficiency
is usually by stirring the solution. However, it has some drawbacks,
such as the generation of a non-uniform and time-dependent advection,
which influences the detection signal, and it usually requires a large
number of analytes. Compared with stirring in batch reactors, flow
detection is easy to describe and control and requires fewer analytes.
Thus, a 3D-printed reactor was prepared to realize the flow detection
of lead ions.[30−34] When the electrochemical analysis is performed, the analyte flows
over the electrode surface, and then the corresponding electrochemical
signal can be quickly displayed. Therefore, lead ions can be effectively
detected in real time.
Results and Discussion
Characterization of the Laser-Induced Porous
Carbon (LIPC)
When the laser irradiates the polyimide (PI)
film, PI begins to depolymerize and carbonize. Because laser sintering
takes place in the atmosphere, O2 in the air can quickly
react with carbon at high temperatures to produce carbon monoxide
and carbon dioxide gases. The thermolysis of the PI film also produces
these gases to form pores. Therefore, the carbonization layer on the
PI film has a porous structure.[35] The porous
carbon was produced by laser irradiation at a power of 27 W. Characterization
was performed by scanning electron microscopy (SEM), Raman spectrometry,
X-ray diffraction (XRD), and Brunauer–Emmett–Teller
(BET) test.The SEM image clearly shows that the laser-induced
structure has a porous morphology, as shown in Figure A. The cross-sectional SEM image (Figure B) compares the LIPC
structure with the PI substrate. It can be clearly seen from the figure
that the structure of PI changed significantly after decomposition
and carbonization. The inset in Figure A shows the confocal Raman spectra of the porous carbon.
The yellow line is the Raman spectrum of porous carbon on the laser-engraved
track, and the blue line is the Raman spectrum of porous carbon between
the two tracks. There are three strong peaks, namely, the D peak at
1352 cm–1, the G peak at 1585 cm–1, and the 2D peak at 2701 cm–1. The D peak is induced
by defects or bent sp2-carbon bonds. The G peak and 2D
peak are associated with second-order zone-boundary phonons.[23] In addition, the D/G peak intensity ratio represents
the graphitization degree of carbon, and the smaller the D/G peak
intensity ratio is, the better the graphitization degree of carbon
is. It can be seen from the figure that the graphitization degree
of carbon between two laser-engraved tracks is higher than that on
the laser-engraved track. This is because the sintering temperature
at the laser-engraved track is very high. Excessive temperature will
destroy the carbon structure, making the graphitization degree of
the porous carbon in this region lower than that of the porous carbon
between two laser-engraved tracks.
Figure 1
Characterization of LIPC. (A) SEM image
of LIPC. The scale bar
is 500 μm. The inset is the Raman spectra of carbon on the region
of the laser scratch track and between the two tracks. (B) Cross-sectional
SEM image of LIPC on the PI film. The scale bar is 100 μm. (C)
XRD of powdered LIPC scraped from the PI film. (D) N2 adsorption–desorption
isotherm of powdered LIPC scraped from the PI film. The inset is the
PSD of the sample calculated from the desorption isotherm.
Characterization of LIPC. (A) SEM image
of LIPC. The scale bar
is 500 μm. The inset is the Raman spectra of carbon on the region
of the laser scratch track and between the two tracks. (B) Cross-sectional
SEM image of LIPC on the PI film. The scale bar is 100 μm. (C)
XRD of powdered LIPC scraped from the PI film. (D) N2 adsorption–desorption
isotherm of powdered LIPC scraped from the PI film. The inset is the
PSD of the sample calculated from the desorption isotherm.The XRD pattern shows two peaks at 2θ = 25.88°
(002)
and 42.88° (100), respectively (Figure C). The 002 peak intensity is relatively
low. It indicates that the porous substance is carbon with a low degree
of graphitization. Figure D shows the N2 adsorption–desorption isotherm
of powdered LIPC. The inset is the pore size distribution diagram
(PSD) of the sample calculated from the desorption isotherm. BET multipoint
method analysis results show that the specific surface area of LIPC
is 191 m2/g, the total pore volume is 0.34 mL/g, and the
average hole diameter is 7.1 nm.
Optimization
of the Laser-Induced Porous Carbon
Electrode (LIPCE)
The power and path of laser irradiation
can affect the carbonization of the PI film, which further affects
the sensitivity of lead ion detection. To obtain the best electrochemical
detection performance, the laser irradiating power and path were optimized.LIPCs with powers of 15 W (LIPC15), 21 W (LIPC21), 27 W (LIPC27),
and 30 W (LIPC30) were fabricated. SEM images of LIPC samples (Figure A–D) show
the effect of laser power on the laser-induced carbon structure. The
specific surface areas of LIPC15, LIPC21, LIPC27, and LIPC30 were
tested by a specific surface area analyzer, and their specific surface
areas are 261.476, 230.525, 215.033, and 314.867 m2/g,
respectively. The result shows that the specific surface area of LIPC30
is much larger than those of LIPC15, LIPC21, and LIPC27. It indicates
that when the laser power is 30 W, more pores are formed on the surface
of the PI film. LIPC15, LIPC21, LIPC27, and LIPC30 were also tested
by the confocal Raman spectrometer (Figure F). It shows that the degree of graphitization
of LIPC21, LIPC27, and LIPC30 is significantly higher than that of
LIPC15.
Figure 2
Optimization of the LIPCE formed at different laser powers. SEM
images of LIPCs formed at different laser powers: (A) 15 W, (B) 21
W, (C) 27 W, and (D) 30 W. (E) Digital photo of the LIPCE. (F) Confocal
Raman spectrometry of LIPCs with different powers. (G) CV curves of
different power LIPCEs in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. (H) DPV curves of different power LIPCEs
in 0.4 mg/L Pb2+ solution (buffer: 0.1 M NaAc-HAc containing
0.1 M NaCl, pH = 4).
Optimization of the LIPCE formed at different laser powers. SEM
images of LIPCs formed at different laser powers: (A) 15 W, (B) 21
W, (C) 27 W, and (D) 30 W. (E) Digital photo of the LIPCE. (F) Confocal
Raman spectrometry of LIPCs with different powers. (G) CV curves of
different power LIPCEs in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. (H) DPV curves of different power LIPCEs
in 0.4 mg/L Pb2+ solution (buffer: 0.1 M NaAc-HAc containing
0.1 M NaCl, pH = 4).Then, the PI films sintered
with 15, 21, 27, and 30 W laser powers
were made into electrodes to detect lead ions. Figure E shows a digital photo of the LIPCE. These
electrodes were tested by CV (Figure G). It shows that the performance of LIPCE21, LIPCE27,
and LIPCE30 is better than that of LIPCE15. Then, the electrodes were
used to detect 0.4 mg/L Pb2+ by stripping voltammetry. Figure H shows the corresponding
DPV curves. Compared with LIPCE15, we can see that LIPCE21, LIPCE27,
and LIPCE30 are more sensitive to lead ion detection. It shows that
the highly graphitized LIPC makes the concentration of lead ions deposited
on the electrode surface higher and the rate of dissolution faster.
The difference of their specific surface area has no effect on the
electrochemical test results. There is almost no difference between
LIPCE21, LIPCE27, and LIPCE30. Here, we choose LIPCE30 for the follow-up
experiments.Then, the porous carbon electrodes prepared by
six different laser
sintering paths were studied. The first one (1) was perpendicular
to the long side of the electrode, the second one (2) was at an angle
of 45° to the long side of the electrode, the third one (3) was
parallel to the long side of the electrode, the fourth one (4) was
sintered by path (1) twice, the fifth one (5) was sintered by path
(3) twice, and the sixth one (6) was sintered by path (1) and then
sintered by path (3). SEM images are shown in Figure A. It can be seen from the figure that the
porous structure of the LIPC surface sintered through paths (4) and
(5) is more obvious. However, when the PI film was sintered in accordance
with the designed electrode pattern, the surface of the LIPCE prepared
through path (2) would be damaged. This is because the length of the
sintering path is uneven and irregular, leading to uneven heating
on the PI surface. This caused some porous carbon layers to warp off
from the PI surface. Incomplete surfaces could negatively affect the
results of electrochemical detection. Therefore, it should not be
made into an electrode. The performance of other five LIPCE30 electrodes
prepared by different laser irradiation paths was compared by CV (Figure B). It can be seen
from the figure that the redox peak intensity of the fourth electrode
is higher and the peak potential difference (ΔEP) is smaller than others. It shows that the laser irradiation
path (4) makes the charge transfer on the electrode surface easier
and faster. Then, the sensitivity of the five electrodes to lead ion
detection was compared by DPV (Figure C). The LIPCE30 with the laser irradiation path (4)
(LIPCE30 (4)) shows the highest DPV response. It indicates that the
laser irradiation path (4) makes the lead ion concentration deposited
on the electrode surface higher and the dissolution rate faster. The
results show that LIPCE30 (4) had excellent electrochemical performance,
and we used it in the subsequent experiments.
Figure 3
Optimization of LIPCE30
formed at different laser sintering paths.
(A) SEM of porous carbon generated by six different laser sintering
paths. (The laser sintering path of (1) was perpendicular to the long
side of the electrode. The laser sintering path of (2) was at an angle
of 45° to the long side of the electrode. The laser sintering
path of (3) was parallel to the long side of the electrode. (4) The
laser sintered the same path as (1) twice. (5) The laser sintered
the same path as (3) twice. (6) The laser sintered the same path as
(1) and then sintered the same path as (3).) (B) CV curves of LIPCE30
with different laser-irradiated paths in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. (C) DPV curves of the LIPCE30
with different laser-irradiated paths in 0.4 mg/L Pb2+ solution
(buffer: 0.1 M NaAc-HAc containing 0.1 M NaCl, pH = 4).
Optimization of LIPCE30
formed at different laser sintering paths.
(A) SEM of porous carbon generated by six different laser sintering
paths. (The laser sintering path of (1) was perpendicular to the long
side of the electrode. The laser sintering path of (2) was at an angle
of 45° to the long side of the electrode. The laser sintering
path of (3) was parallel to the long side of the electrode. (4) The
laser sintered the same path as (1) twice. (5) The laser sintered
the same path as (3) twice. (6) The laser sintered the same path as
(1) and then sintered the same path as (3).) (B) CV curves of LIPCE30
with different laser-irradiated paths in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. (C) DPV curves of the LIPCE30
with different laser-irradiated paths in 0.4 mg/L Pb2+ solution
(buffer: 0.1 M NaAc-HAc containing 0.1 M NaCl, pH = 4).
Electrochemical Characterization of the LIPCE
The electrochemical characterization of LIPCE30 (4) was performed
by CV. As shown in Figure A, the CVs of 5 mM [Fe(CN)6]3–/4– were recorded using LIPCE30 (4) at different scan rates (20–200
mV/s). The CVs have two peaks: the oxidation peak and the reduction
peak. ΔEP is greater than 0.058
V, and iPa/iPc is greater than 1. With the increase in the scan rate, the oxidation
peak position shifts to the positive direction, and the reduction
peak position shifts to the negative direction. Moreover, the peak
current increases with the increase in the scan rate and is proportional
to the square root of the scan rate (Figure B). It indicates that this is a quasi-reversible
electrode reaction process. In the electrode reaction, electrically
active substances first reached the electrode surface through diffusion
and then participated in the reaction by adsorption on the electrode
surface. The relationship between the peak current and the square
root of the scan rate is linear, which indicates that the electrode
process was mainly controlled by diffusion. The electrochemical surface
area (ECSA) of LIPCE30 (4) was estimated by the Randles–Sevčik
equation:[36−41]where IP is the
peak current, A is the effective surface area, D is the diffusion coefficient of K3Fe(CN)6 (7.6 × 10–6 cm2/s), n is the number of electrons transferred (n = 1), v is the scan rate, and C is the concentration of K3Fe(CN)6. For LIPCE30
(4), the ECSA is about 0.3275 cm2.
Figure 4
Electrochemical performance
of LIPCE30 (4). (A) CV curves of LIPCE30
(4) in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. Scan rates from 20 to 200 mV/s. (B) Plots of peak current versus
the square root of the scan rate corresponding to (A).
Electrochemical performance
of LIPCE30 (4). (A) CV curves of LIPCE30
(4) in 0.1 M KCl and 5 mM [Fe(CN)6]3–/4–. Scan rates from 20 to 200 mV/s. (B) Plots of peak current versus
the square root of the scan rate corresponding to (A).
Optimization of Analytical Conditions
Determining the best experimental conditions was the key to obtain
the highest detection sensitivity.[21] The
analytical conditions were optimized before evaluating the electrochemical
detection performance of LIPCE30 (4).The enrichment process
has a great influence on the detection sensitivity of lead ions, and
usually, the solution must be stirred to reduce the thickness of the
diffusion layer. The flow of the solution allows the lead ions to
effectively approach the electrode surface and more easily adsorb
on the porous carbon electrode. Here, we used a flow reactor to make
the liquid flow. The velocity distribution near the electrode surface
was simulated numerically when the electrode was in batch and flow
reactors (Figure ).
It can be seen that agitation leads to uneven dynamics, and mass transfer
is mainly limited to the liquid phase far away from the solid phase
(Figure B), which
is not beneficial for the adsorption of the ions. However, in the
flow reactor, the flow of solution can not only produce more stable
advection in a controlled and predictable manner, but more importantly,
it can also effectively improve surface dynamics (Figure E).
Figure 5
Numerical simulation
of agitation and flow. (A) Schematic diagram
of the magnetic agitator. (B, E) Velocity field distribution on the
parallel surface 0.3 mm away from the electrode surface in the stirred
state and flow state, respectively (unit: m/s). (C, F) i–t and DPV curves of LIPCE30 (4) in 0.4 mg/L
Pb2+ solution, respectively (stirring rate: 300 r/min;
flow rate: 2 mL/min). (D) Schematic diagram of the flow device.
Numerical simulation
of agitation and flow. (A) Schematic diagram
of the magnetic agitator. (B, E) Velocity field distribution on the
parallel surface 0.3 mm away from the electrode surface in the stirred
state and flow state, respectively (unit: m/s). (C, F) i–t and DPV curves of LIPCE30 (4) in 0.4 mg/L
Pb2+ solution, respectively (stirring rate: 300 r/min;
flow rate: 2 mL/min). (D) Schematic diagram of the flow device.It can be seen from Figure C that the i–t curve
in the flowing state is much smoother than that in the stirring state,
which indicates that the stirring of the solution will cause current
fluctuations and unstable deposition of lead ions. However, the solution’s
flow keeps the analyte on the electrode surface always in a new state,
and lead ions can be stably deposited on the electrode surface and
generate a stable current. It can be seen from Figure F that the detection sensitivity of the electrode
to lead ions in the flow state is higher than that in the stirred
and static states. Therefore, flow detection was adopted in the following
experiments.The deposition time from 100 to 1000 s was studied
(Figure A). The peak
current increases
with the increase in electrodeposition time, which indicates that
extending the electrodeposition time can improve the sensitivity of
lead ion detection. But with the increase in electrodeposition time,
the linear relationship will become worse. Therefore, we choose 200
s as the best electrodeposition time. The deposition potential from
−1.8 to −0.8 V was studied (Figure B). The peak current increases significantly
from −1.8 to −1.2 V and reaches a maximum at −1.2
V, and then, the current decreases. So, −1.2 V was selected
as the optimal deposition potential. Similarly, the flow rate of the
analyte from 1.4 to 2.4 mL/min was studied (Figure C). The DPV response peak current increases
with increasing flow rate and reaches a maximum at 2 mL/min. Then,
the current drops significantly. When the flow rate is less than 2
mL/min, the deposition amount of lead ions increases with the increase
in the flow rate. When the flow rate is more than 2 mL/min, the lead
ion cannot be deposited effectively due to the high flow rate. So,
the flow rate of 2 mL/min was selected for this work. Then, the pH
of the buffer was optimized. Figure D shows the effect of NaAc-HAc buffer pH on electrochemical
behavior. The response peak current of the DPV curve increases with
the increase in the pH value and reaches the maximum at pH = 4. Then,
the current gradually decreases. When the pH of the buffer increases,
the concentration of OH– in the solution also increases.
OH– will react with Pb2+, leading to
a decrease in the concentration of lead ions in the solution. So,
the DPV response peak current goes down. Therefore, the buffer solution
with pH = 4 was selected for the experiment.
Figure 6
Optimization of electrochemical
analytical conditions. Plot of
the DPV peak current curve versus (A) deposition time, (B) deposition
potential, (C) analyte flow rate, and (D) buffer pH.
Optimization of electrochemical
analytical conditions. Plot of
the DPV peak current curve versus (A) deposition time, (B) deposition
potential, (C) analyte flow rate, and (D) buffer pH.
Analytical Performance of the Sensor toward
Pb2+ Detection
To realize the quantitative detection
of Pb2+, the sensor detected different concentrations of
Pb2+ through DPV measurement. LIPCE30 (4) was used as the
electrode, and the experiment was performed under the optimized conditions. Figure A shows the DPV curves
for the concentrations from 0.05 to 0.35 mg/L. Figure B shows that peak currents are linearly related
to the concentrations of Pb2+, and the corresponding calibration
curve is y = 6.4440x – 0.2041
(R2 = 0.9849). The limit of detection
(LOD) is estimated to be 0.0330 mg/L for Pb2+.
Figure 7
Analytical
performance of the sensor toward Pb2+ detection.
(A) DPV recordings obtained for increasing concentrations of Pb2+ (0.05–0.35 mg/L) at LIPCE30 (4) in 0.1 M NaAc-HAc
containing 0.1 M NaCl (pH = 4). Deposition potential: −1.2
V; deposition time: 200 s; flow rate of the analyte: 2 mL/min. (B)
The calibration curve corresponded to (A).
Analytical
performance of the sensor toward Pb2+ detection.
(A) DPV recordings obtained for increasing concentrations of Pb2+ (0.05–0.35 mg/L) at LIPCE30 (4) in 0.1 M NaAc-HAc
containing 0.1 M NaCl (pH = 4). Deposition potential: −1.2
V; deposition time: 200 s; flow rate of the analyte: 2 mL/min. (B)
The calibration curve corresponded to (A).
Reproducibility, Stability, and Anti-interference
Studies
The reproducibility of LIPCE30 (4) was evaluated
by 5 times DPV measurements of 0.4 mg/L Pb2+ (Figure A). The relative
standard deviation (RSD) is 4.79% (Figure B). It indicates that the reproducibility
of LIPCE30 (4) is good. The stability of LIPCE30 (4) was evaluated
by CV measurement in a 5 mM [Fe(CN)6]3–/4- solution containing 0.1 M KCl. Figure C shows that the CV oxidation peak current
changes little within 5000 sweep segments, indicating that LIPCE30
(4) has high stability. To study the anti-interference of LIPCE30
(4), some possible interfering metal ions were added into 0.4 mg/L
Pb2+. These ions were 10 times more concentrated than lead
ions. It can be seen from Figure D that these metal ions have little interference with
Pb2+ except for Cu2+. When the interference
of Cu2+ to lead ion detection was studied, the oxidation
peak of Cu2+ would also appear. It indicated that Cu2+ and Pb2+ were deposited together on the electrode
surface when the deposition potential was −1.2 V. Then, Cu2+ would be dissolved together with Pb2+. So, the
lead ion has a lower response peak current. However, lead ions still
can be detected in the presence of Cu2+. These characteristics
make LIPCE30 (4) have a broad application prospect in electrochemical
detection.
Figure 8
Reproducibility, stability, and anti-interference studies. (A)
Five DPV recordings of Pb2+ (0.4 mg/L) at the same LIPCE30
(4) in 0.1 M NaAc-HAc containing 0.1 M NaCl (pH = 4). (B) Plot of
peak current versus the cycle number corresponding to (A). (C) Plot
of CV oxidation peak current versus sweep segments in 0.1 M KCl containing
5 mM [Fe(CN)6]3–/4–. (D) Anti-interference
study of LIPCE30 (4) to 0.4 mg/L Pb2+ in the presence of
4 mg/L Ba2+, Co2+, Cu2+, Hg2+, Zn2+, Mg2+, Ca2+, and Cd2+.
Reproducibility, stability, and anti-interference studies. (A)
Five DPV recordings of Pb2+ (0.4 mg/L) at the same LIPCE30
(4) in 0.1 M NaAc-HAc containing 0.1 M NaCl (pH = 4). (B) Plot of
peak current versus the cycle number corresponding to (A). (C) Plot
of CV oxidation peak current versus sweep segments in 0.1 M KCl containing
5 mM [Fe(CN)6]3–/4–. (D) Anti-interference
study of LIPCE30 (4) to 0.4 mg/L Pb2+ in the presence of
4 mg/L Ba2+, Co2+, Cu2+, Hg2+, Zn2+, Mg2+, Ca2+, and Cd2+.
Conclusions
In conclusion, we reported a low-cost, direct laser sintering method
for electrode preparation. The electrode can detect lead ions quickly
and efficiently. The laser irradiation power and path significantly
affect the electrochemical performance of LIPCE. The best laser irradiation
power and path were obtained through optimization. According to the
electrochemical characterization, the electrode process was controlled
by diffusion. The electrochemical deposition was usually carried out
in the state of agitation. To compare the effects of agitation and
flow on electrochemical detection, the velocity distribution near
the electrode surface under stirring and flowing conditions was simulated
numerically. Simulation results showed that the laminar flow in a
continuous flow reactor was more beneficial to the improvement of
electrode surface dynamics than stirring in a batch reactor, and electrochemical
experiments also showed that flow detection had higher sensitivity.
Under the optimal experimental conditions, the sensor was used for
the quantitative analysis of Pb2+. The estimated LOD was
0.0330 mg/L. The study also proved that LIPCE30 (4) had good stability
and repeatability. Therefore, the sensor was expected to be widely
used for more electrochemical detections.
Experimental
Section
Materials and Reagents
PI was obtained
from Changchun Gao Qi Polyimide Material Co., Ltd. Co(NO3)2·6H2O, K3[Fe(CN)6], K4Fe(CN)6·3H2O, CH3COONa, CH3COOH, HCl, and KCl were bought from Tianjin
Kemiou Chemical Reagent Corp. Mg(NO3)2·6H2O, Ba(NO3)2, and Pb(NO3)2 were received from the Shanghai Aladdin Biochem Technology
Corp. Zn(NO3)2·6H2O and Ca(NO3)2·4H2O were purchased from Damao
Chemical Reagent Factory (Tianjin, China). NaCl and Cd(NO3)2·4H2O were obtained from Shanghai Macleans
Biochemical Technology Co., Ltd. (CH3COO)2Cu·H2O was bought from Sinopharm Chemical Reagent Co., Ltd. Hg(NO3)2·H2O was received from West Asia
Chemical Technology (Shandong) Co., Ltd. The copper sheet was bought
from Shanghai Lujiang Hardware. Silver wire (1.0 mm in diameter, annealed,
99.9%, metal basis) was obtained from Alfa Aesar Chemical Co., Ltd.
Epoxy adhesive (Devcon 14250) was obtained from Shenzhen Darbond Technology
Co., Ltd. Deionized water was used for all of the experiments.
Preparation of the Working Electrode (WE),
Conference Electrode (CE), and Ag/AgCl Reference Electrode (RE)
Both WE and CE were manufactured by laser sintering on PI films
(Scheme A). Using
a CO2 laser (marking speed: 400 mm/s; frequency: 15 kHz;
resolution: 0.001 mm; repeated positioning accuracy: 0.003 mm), the
PI film was converted into a porous carbon electrode with a design
pattern (7.48 × 28.05 mm). The RE was prepared by an electroplating
method. First, the silver wire was ultrasonically cleaned in alcohol
for 30 min and rinsed with deionized water. It was then electroplated
for 30 min in a 0.1 M HCl solution by chronopotentiometry, with commercial
Ag/AgCl as the RE and Pt as the CE. The silver wire was coated with
AgCl at last, and we obtained the Ag/AgCl RE.
Scheme 1
Procedure of Electrochemical
Analysis
(A) Laser sintering process;
(B, C) schematic and digital photographs of the electrochemical sensor,
respectively.
Procedure of Electrochemical
Analysis
(A) Laser sintering process;
(B, C) schematic and digital photographs of the electrochemical sensor,
respectively.
Assembly
of the 3D-Printed Flow Reactor and
Electrodes
We used Solidworks to design the reactor structure
and imported the STL files into a 3D printer. The reactor can be accurately
manufactured with the 3D printer at a high resolution of 50 μm.
The flow reactor structure with the reaction channel is schematically
illustrated in Scheme B. The reaction channel is 14 mm in length, 1 mm in depth, and 1
mm in width. A hole (r = 0.65 mm) for the RE was
made on the longer side of the channel.The Ag/AgCl RE was fixed
in the reactor with epoxy adhesive. The PI film with the WE and CE
was fixed between the silicon gasket and the higher flange. The whole
device was then fixed with screws. The copper sheets, which were fixed
on the edge of both WE and CE with conductive silver glue, were used
to ensure good contact between the sensor and the electrochemical
workstation. Scheme B,C shows schematic and digital photographs of the electrochemical
sensor, respectively. The inlet and outlet of the channel were connected
by hoses. The inlet hose was then connected to a syringe pump.
Electrochemical Measurements
A three-electrode
system was used in all experiments. The electrodes were characterized
by CV in 0.1 M KCl containing 5 mM [Fe(CN)6]3–/4–. Except for special declarations, the scanning rate of the CV test
was 0.1 V/s. The lead ions were determined by DPV. The pulse width
was 0.05 s, the pulse period was 0.5 s, and the pulse amplitude was
0.05 V.
Calibration Experiments
The RSD between
the electrodes can be solved by calibration experiments. First, we
use electrode A to obtain the standard curve and its peak current
(IPA) in 0.4 mg/L Pb2+ solution.
For the other electrode B, we can also obtain its peak current (IPB) in 0.4 mg/L Pb2+ solution. IPA/IPB is the correction
factor (σ). When using the B electrode to detect the sample,
the corresponding peak current (IPI) is
obtained. IPI × σ is the peak
current of the A electrode for the sample, and then the concentration
of the sample is obtained from the standard curve.
Characterization
SEM images were
obtained by a QUANTA 450 scanning electron microscope at 20 kV. The
pore properties were surveyed by an ASAP 2010 analysis instrument
at 77 K. The specific surface areas were calculated by the BET method,
and the pore size was calculated using the Barrett–Joyner–Halenda
(BJH) model. XRD patterns were measured on a Rigaku D/MAX-2400 X-ray
powder diffractometer (Japan) operating at 40 kV and 100 mA using
Cu Kα radiation. The Raman spectra were recorded with a Nicolet
Almega XR Raman system with a 532 nm laser. LIPC was obtained by a
CO2 laser (Coherent C series 30 W, 10.6 μm wavelength).
The injector containing the analyte was controlled by a syringe pump
(Leifu, TYD03-01). The electrochemical performance of the LIPCE was
studied using CV and DPV on a CHI model 660D electrochemical workstation
(CH Instrument, Inc.). The flow reactor was produced by 3D printing
from stereolithography resin (methacrylate photopolymer resin, Formlabs
Form 2, 0.05 mm layer resolution).
Authors: Olja Simoska; Marta Sans; Mignon D Fitzpatrick; Christopher M Crittenden; Livia S Eberlin; Jason B Shear; Keith J Stevenson Journal: ACS Sens Date: 2018-12-20 Impact factor: 7.711
Authors: Jian Lin; Zhiwei Peng; Yuanyue Liu; Francisco Ruiz-Zepeda; Ruquan Ye; Errol L G Samuel; Miguel Jose Yacaman; Boris I Yakobson; James M Tour Journal: Nat Commun Date: 2014-12-10 Impact factor: 14.919