The electrocardiographic forward problem: A benchmark study.

Jake A Bergquist1, Wilson W Good1, Brian Zenger2, Jess D Tate3, Lindsay C Rupp1, Rob S MacLeod4.   

Abstract

BACKGROUND: Electrocardiographic forward problems are crucial components for noninvasive electrocardiographic imaging (ECGI) that compute torso potentials from cardiac source measurements. Forward problems have few sources of error as they are physically well posed and supported by mature numerical and computational techniques. However, the residual errors reported from experimental validation studies between forward computed and measured torso signals remain surprisingly high.
OBJECTIVE: To test the hypothesis that incomplete cardiac source sampling, especially above the atrioventricular (AV) plane is a major contributor to forward solution errors.
METHODS: We used a modified Langendorff preparation suspended in a human-shaped electrolytic torso-tank and a novel pericardiac-cage recording array to thoroughly sample the cardiac potentials. With this carefully controlled experimental preparation, we minimized possible sources of error, including geometric error and torso inhomogeneities. We progressively removed recorded signals from above the atrioventricular plane to determine how the forward-computed torso-tank potentials were affected by incomplete source sampling.
RESULTS: We studied 240 beats total recorded from three different activation sequence types (sinus, and posterior and anterior left-ventricular free-wall pacing) in each of two experiments. With complete sampling by the cage electrodes, all correlation metrics between computed and measured torso-tank potentials were above 0.93 (maximum 0.99). The mean root-mean-squared error across all beat types was also low, less than or equal to 0.10 mV. A precipitous drop in forward solution accuracy was observed when we included only cage measurements below the AV plane.
CONCLUSION: First, our forward computed potentials using complete cardiac source measurements set a benchmark for similar studies. Second, this study validates the importance of complete cardiac source sampling above the AV plane to produce accurate forward computed torso potentials. Testing ECGI systems and techniques with these more complete and highly accurate datasets will improve inverse techniques and noninvasive detection of cardiac electrical abnormalities.
Copyright © 2021 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Cardiac bioelectricity; Cardiac electrophysiology; Electrocardiographic forward problem; Electrocardiographic imaging; Experimental cardiac electrophysiology; Experimental validation

Year:  2021        PMID: 34051453     DOI: 10.1016/j.compbiomed.2021.104476

Source DB:  PubMed          Journal:  Comput Biol Med        ISSN: 0010-4825            Impact factor:   4.589


  2 in total

1.  Body Surface Potential Mapping: Contemporary Applications and Future Perspectives.

Authors:  Jake Bergquist; Lindsay Rupp; Brian Zenger; James Brundage; Anna Busatto; Rob S MacLeod
Journal:  Hearts (Basel)       Date:  2021-11-05

2.  Reconstruction of cardiac position using body surface potentials.

Authors:  Jake A Bergquist; Jaume Coll-Font; Brian Zenger; Lindsay C Rupp; Wilson W Good; Dana H Brooks; Rob S MacLeod
Journal:  Comput Biol Med       Date:  2022-01-20       Impact factor: 4.589

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.