Literature DB >> 34051244

Digital PCR assay for the effective detection of COVID-19 patients with SARS-CoV-2 low viral load.

Yong Sun1, Chengchao Ding2, Qingqing Chen1, Jiajia Xie2, Junling Yu1, Yonglin Shi1, Chengcheng Jiang2, Zhuhui Zhang1, Hongliang He2, Yinglu Ge1, Wenting Li2, Jun He3, Yong Gao4.   

Abstract

OBJECTIVE: Viral nucleic acid detection by real-time reverse transcription polymerase chain reaction (qPCR) is the current standard method for diagnosis of SARS-CoV-2 infection. However, due to low viral load in some COVID-19 patients, false negative results from this method have been repeatedly reported.
METHOD: In this study, we compared the sensitivity and specificity of digital PCR (dPCR) in simulated samples and clinical samples with qPCR assay through a series of vigorous tests.
RESULTS: The results showed that dPCR was more sensitive than qPCR especially for samples with low viral load (≤3 copies). In addition, dPCR had similar specificity as qPCR and could effectively distinguish other human coronaviruses and influenza virus from SARS-CoV-2. More importantly, dPCR was more sensitive than qPCR in detecting the virus in the "negative" samples from recurrent COVID-19 patients.
CONCLUSIONS: In summary, dPCR could serve as a powerful complement to the current qPCR method for SARS-CoV-2 detection, especially for the samples with extremely low viral load, such as recurrent COVID-19 patients.
Copyright © 2021. Published by Elsevier B.V.

Entities:  

Keywords:  Digital PCR (dPCR); Recurrent COVID-19 patient; SARS-CoV-2; Sensitivity; Specificity

Mesh:

Substances:

Year:  2021        PMID: 34051244      PMCID: PMC8149472          DOI: 10.1016/j.jviromet.2021.114185

Source DB:  PubMed          Journal:  J Virol Methods        ISSN: 0166-0934            Impact factor:   2.623


Introduction

The ongoing outbreak of 2019 novel coronavirus disease (COVID-19) caused by a new coronavirus, severe acute respiratory syndrome (SARS) coronavirus 2 (SARS-CoV-2), has spread rapidly across the world and became a public health emergency of international concern (Huang et al., 2020; Di Gennaro et al., 2020). Both human-to-human and asymptomatic transmission of SARS-CoV-2 infection through close contacts have been reported (Lai et al., 2020; Rothe et al., 2020). Early diagnosis of COVID-19 is critical for prevention and control of virus infection and can help patients receive early treatment and prevent disease progression (Tahamtan and Ardebili, 2020). To date, viral nucleic acid detection by real-time reverse transcription polymerase chain reaction (qPCR) is regarded as the gold standard method for diagnosis of SARS-CoV-2 infection (Davila et al., 2014; Zhu et al., 2020; Liu et al., 2020). However, the sensitivity and reliability of qPCR has been questioned due to its false negative results in some clinically diagnosed patients who have exposure history, typical clinical symptoms, and typical image of CT scanning (Xiao et al., 2020; Pan et al., 2020). Furthermore, recently, the recurrence of SARS-CoV-2 in discharged COVID-19 patients has also been reported (Xiao et al., 2020; Yuan et al., 2020a). Even though its mechanism is still unclear, the possible explanations might be false negative of detection, and reactivation or re-infection of the virus. Taking into account the fact that recurrence and asymptomatic COVID-19 patients usually have a very low viral load, a more sensitive detection method is urgently needed to improve the accuracy in identifying SARS-CoV-2 infected patients, in order to effectively prevent the virus transmission. Digital PCR (dPCR) is an absolute quantitative method that partitions the PCR reaction into a large number of smaller reactions and collect the intensity of fluorescence signals at the end point of each reaction (Wang et al., 2016; Basu, 2017). Unlike qPCR, dPCR does not depend on standard curves or relative threshold (CT) values for quantitation (Pavsic et al., 2016) and is more tolerant to PCR inhibitors (Dingle et al., 2013). The absolute quantification results come from Poisson statistics by counting the negative reactions followed by calculating the positive wells or the nucleic acid copies in the sample (Strain et al., 2013). The high precision and the low limit of detection of dPCR could be used for detection of small variation of DNA or RNA copy number (Whale et al., 2012), rare allele mutation (Zhou et al., 2018a) and gene fusion (Zhou et al., 2018b). Recently, several studies reported that dPCR showed higher sensitivity and specificity in detecting low viral load of SARS-CoV-2 compared with qPCR (Vasudevan et al., 2021; Alteri et al., 2020; Falzone et al., 2020). However, the detectability for the samples with extremely low viral load, such as obtained from recurrent COVID-19 patients, remains unclear. In this study, we compared the sensitivity and specificity of dPCR with qPCR on simulated and clinical sputum samples. To analyze the detection efficacy of the low viral load samples, dPCR quantitative assay of SARS-CoV-2 were explored in clinically recurrent COVID-19 patients. Our findings demonstrated that in comparison with qPCR, dPCR was more sensitive with low viral load of both simulated and clinical SARS-CoV-2 samples and could accurately reflect the variation in copy number of viral RNA. The dPCR also showed high specificity in distinguishing the SARS-CoV-2 from other respiratory pathogens such as common influenza virus and other human coronaviruses. In addition, dPCR shows superiority over qPCR in detection of “recurrent” COVID-19 patients, which could be used as an important complement method for difficult or confusing samples in the clinic.

Materials and methods

Patients and preparation of simulated and clinical samples

We enrolled 21 COVID-19 patients with confirmed SARS-CoV-2 infection including 17 recurrent patients and 4 non-recurrent patients from Jan 21 to Feb 29, 2020 at the First Affiliated Hospital of University of Science and Technology of China (USTC), Huainan First Hospital, Zongyang Hospital, Lixin Hospital, Bozhou Hospital, Fuyang Second Hospital, and Guoyang Hospital in Anhui Province, China. The throat swab, sputum, and/or anal swab were collected. In the meantime, six negative lower respiratory tract samples from 6 healthy donors were also collected and mixed together to serve as a negative matrix for preparation of different simulated virus samples with various LOD. In addition, 13 sputum samples containing influenza virus and other human coronavirus were collected from the corresponding patients to serve as negative controls. Following throat and anal swabs and sputum samples being collected and the simulated samples being prepared, the viral RNAs were extracted using the Nucleic acid purification kit (TIANLONG Technology CO. LTD., Cat. No: T014) according to the manufacturer’s instructions and were subsequently subjected to the detection by qPCR and dPCR. The reference SARS-CoV-2 RNA [GBW(E)091112] was obtained from Shanghai Institute of Measurement and Testing Technology, in which the target gene covers the full length of N gene (GenBank No.MT027064.1) with the RNA concentration of 3.0 107 copies/μL. The normalized RNA concentration was confirmed by dPCR method for authentic SARS-CoV-2. The study protocol was approved by the medical ethics board of Anhui Center for Disease Control and Prevention (CDC). Please note that the study was performed on existing samples collected during standard diagnostic tests, posing no extra burden to patients.

Digital PCR (dPCR)

The dPCR assay was performed via the digital PCR chips (Thermofisher, cat#A26317) following the manufacturer’s protocols, and QuantStudio 3D Digital PCR System (Thermo Fisher, Waltham, MA) using the Gnomegen COVID-19 RT-PCR Detection Kit (Gnomegen, San Diego, CA). The kit allowed the detection of the N gene and a positive RNase P reference gene, which was approved for Emergency Use Authorization (EUA) by FDA, pursuant to Section 564 of the Federal Food, Drug, and Cosmetic Act (the Act) (21 U.S.C. §360bbb-3). Briefly, 14.5 μl of reaction mix was prepared for each reaction, comprised of 7.25 μl of digital PCR master mix (Thermofisher, A26358), 0.20 μl Superscript II (Thermofisher, cat# 18064014), 0.20 μl RNaseOUT (Thermofisher, cat# 10777019), 0.725 μl COVID-19 Assay, 2.125 μl Molecular Grade Nuclease Free Water and 4 μl RNA samples. The reaction mixture was then loaded on the chips using the QuantStudio™ 3D Digital PCR Chip Loader (Thermofisher, cat# 4482592), and cycled in ProFlex™ 2×Flat Block Thermal Cycler (Thermofisher, cat#4484078) under the following protocol: step 1, 42 °C for 20 min (reverse transcription); step 2, 96 °C for 10 min (DNA polymerase activation, denaturation) and 60 °C for 2 min (annealing); step 3, 39 cycles of 98 °C for 30 s (denaturation) and 60 °C for 2 min (annealing); step 4, hold at 20 °C. Finally, the chips were imaged by QuantStudio™ 3D Digital PCR Instrument with Power Cord (Thermofisher, cat# 4489084) and analyzed using the QuantStudio 3D Analysis Suite Cloud Software (Current version: 3.1.4-PRC-build1) to measures the concentration of N gene and RNase P gene, respectively. The result was considered positive when the FAM signals of the SARS-CoV-2 target ≥ 3 with the VIC signals of the RNase P reference target ≥ 0, and negative when the FAM signals ≤ 2 with the VIC signals ≥ 60. If the FAM signals ≤ 2 with the VIC signals <60, the results were considered inconclusive and excluded. Concentrations of the target RNA sequences, along with their Poisson-based 95 % confidence intervals were calculated by using the QuantStudio 3D Analysis Suite Cloud.

Fluorescent quantitative PCR (qPCR)

qPCR was conducted with primers and probes targeting the ORF1ab (FAM) and N gene (VIC). Reaction system and amplification conditions were carried out according to the manufacturer's specifications (BioGerm, China). A 25μl of reaction mixture contained 12 μl of reaction buffer, 4 μl of mixed enzymes, 4 μl of ORF1ab/N reaction mixture and 5 μl of nucleic acid from each sample. Thermal cycling was performed at 50 °C for 10 min for reverse transcription, followed by 95 °C for 5 min (pre-denaturation) and then 40 cycles of 95 °C for 10 s (denaturation) and 55 °C for 40 s (extension) in Applied Biosystems ABI 7500 real time PCR system (Thermo Fisher, Waltham, MA). The baseline setting was 3–10 or 6–15 cycles of fluorescence signal, and the threshold setting principle was that the threshold line just exceeded the highest point of the amplification curve of the negative control substance, which could also be adjusted according to the noise of the instrument. The result was considered positive when the CT value of ORF1ab or N gene <35, and negative when there was no CT value or the CT value of ORF1ab or N gene ≥ 40. If the CT value of ORF1ab or N gene ≥ 35 and < 40, the result was inconclusive and the sample was retested.

Results

Both dPCR and qPCR could detect low and moderate viral load of simulated SARS-CoV-2 samples

In the present study, we first investigated whether dPCR and qPCR assays could equally effectively detect SARS-CoV-2 virus in low and moderate concentrations. To do so, low and moderate positive simulated samples were prepared by diluting the authenticSARS-CoV-2 into the negative matrix to 3–6 copies per reaction as low viral load samples (20 samples) and 9–15 copies per reaction as moderate viral load samples (10 samples) which were subjected to both dPCR and qPCR assays. As shown in Fig. 1 A, the positive rates of the low and moderate positive samples were 100 % for either dPCR or qPCR assays. When analyzing the values of the low and moderate positive samples detected by dPCR assay, it showed they were largely in the range of the copy number of the virus in the respective samples (Fig. 1B).
Fig. 1

Comparison of dPCR assay with qPCR assay on low and moderate viral load of simulated SARS-CoV-2 samples. (A) The positive rate of low (20 samples) and moderate (10 samples) viral load samples detected by dPCR and qPCR assays. (B) The FAM values of dPCR assay targeting N gene of SARS-CoV-2 of low and moderate viral load samples.

Comparison of dPCR assay with qPCR assay on low and moderate viral load of simulated SARS-CoV-2 samples. (A) The positive rate of low (20 samples) and moderate (10 samples) viral load samples detected by dPCR and qPCR assays. (B) The FAM values of dPCR assay targeting N gene of SARS-CoV-2 of low and moderate viral load samples.

dPCR was more sensitive than qPCR in detecting very low viral load of simulated SARS-CoV-2 samples

We next compared the dPCR and qPCR for their limit of sensitivity of detection of SARS-CoV-2 by using serial dilution of authentic SARS-CoV-2 spiked into the negative matrix. A two-phase approach was used to determine the limit of detection (LOD). In phase I, the preliminary LOD was established through diluting the authentic SARS-CoV-2 into different concentrations. To do so, a total of 5 dilutions were created at 0.375, 0.75, 1.5, 3 and 6 copies of viral RNA with three replicates per concentration. As shown in Fig. 2 A and B, when the virus was diluted between 3 and 6 copies per reaction, the positive rate was 100 % for both dPCR and qPCR. At concentration of 1.5, 0.75, and0.375 copies per reaction, dPCR was still able to obtain the positive results (66.67 % for 1.5 copies, 33.33 % for 0.75 copies, and 66.7 % for 0.375 copies) while qPCR failed to detect any positive signals at any concentration of the samples, indicating that dPCR is more sensitive in detecting SARS-CoV-2 when the viral copy number is below 3. To further elaborate the sensitivity of the assays, both dPCR and qPCR were used to test 20 replicates on simulated samples of 3 copies of SARS-CoV-2. The dPCR assay showed positive rate at 95 % (19/20) while qPCR at 90 % (18/20) (Fig. 2C), indicating that the dPCR was more sensitive than qPCR for samples with viral load around 3 copies.
Fig. 2

Comparison of sensitivity between qPCR and dPCR assays in detecting simulated SARS-CoV-2 samples. (A) Positive rate at each concentration of 6.0, 3.0, 1.5, 0.75, 0.375 copies per reaction. (B) dPCR detection of serial dilution of authentic SARS-CoV-2 spiked into negative matrix. Expected copies of viral RNA were plotted on the X axis versus measured FAM values on the Y axis from dPCR assay targeting N gene of SRAS-CoV-2. Data are representative of 5 independent experiments with 3 replicates for each concentration of 6, 3, 1.5, 0.75, and 0.375 copies of viral RNA. (C) Further confirmation of positive rates detected by qPCR and dPCR assays on 20 replicates of low viral load (3 copies) simulated samples.

Comparison of sensitivity between qPCR and dPCR assays in detecting simulated SARS-CoV-2 samples. (A) Positive rate at each concentration of 6.0, 3.0, 1.5, 0.75, 0.375 copies per reaction. (B) dPCR detection of serial dilution of authentic SARS-CoV-2 spiked into negative matrix. Expected copies of viral RNA were plotted on the X axis versus measured FAM values on the Y axis from dPCR assay targeting N gene of SRAS-CoV-2. Data are representative of 5 independent experiments with 3 replicates for each concentration of 6, 3, 1.5, 0.75, and 0.375 copies of viral RNA. (C) Further confirmation of positive rates detected by qPCR and dPCR assays on 20 replicates of low viral load (3 copies) simulated samples.

dPCR was specific in detecting simulated SARS-CoV-2 samples

To test the specificity of dPCR and qPCR assays, six negative lower respiratory tract clinical samples were mixed together to act as a matrix of negative samples in which the quantified reference N gene RNA was added to prepare the low positive simulated samples. The matrix negative samples (20 replicates for dPCR and 21 replicates for qPCR) and low positive simulated samples (3–6 copies per reaction, 20 replicates for both dPCR and qPCR) were used to compare the specificity between qPCR and dPCR. The negative samples tested by qPCR and dPCR were all “negative”. Among the positive samples, 85.71 % (18/21) were tested positive (≥3 copies) by dPCR while only 75 % (15/20) were tested positive (CT < 35) by qPCR (Fig. 3 A). As shown in Fig. 3B, the FAM signals of dPCR were all below the threshold of 3. Thus, the dPCR assay showed a consistent specificity while performing higher sensitivity than qPCR.
Fig. 3

Comparison of specificity between qPCR and dPCR assays. (A) The positive rate of the negative and positive samples detected by dPCR and qPCR. (B) The FAM values of the negative and low positive samples detected by dPCR.

Comparison of specificity between qPCR and dPCR assays. (A) The positive rate of the negative and positive samples detected by dPCR and qPCR. (B) The FAM values of the negative and low positive samples detected by dPCR.

dPCR was more sensitive than qPCR in detecting clinical SARS-CoV-2 samples

To investigate the possible difference of dPCR and qPCR assays in clinical use, we performed a parallel analysis using positive clinical sputum specimens with CT value at 20 from previous qPCR assay of four confirmed COVID-19 patients. The samples were serially diluted at 1:10 dilution, and a total of 5 dilutions for each specimen were prepared and subjected to dPCR and qPCR assays. The results showed that the dilutions ranged from 10−1 to 10-5 were all detected positive by both dPCR and qPCR. However, for the dilutions of 10-6, there were 4/4 (100 %) samples were detected positive by dPCR, while 2/4 (50 %) positive by qPCR (Table 1 ).
Table 1

Results on gradient dilution samples of confirmed COVID-19 patients by qPCR and dPCR.

SampleCTqPCRChip# of Neg (FAM)# of Neg (VIC)FAMVICdPCR
26210−119.9841+*200418_074027_D02NNC.eds973413315,42712,267+
10−223.3934+200418_074139_D040ID.eds559611,34499704222+
10−327.3696+200418_074241_D0415Z.eds11,80112,200661262+
10−430.4212+200418_074431_D03ZXH.eds16,64816,6919148+
10−534.5454+200418_074532_D03ZXE.eds15,09415,086614+
10−635.6356+200418_074634_D040HO.eds15,68415,664323+
26510−119.3003+200418_074814_D02NM5.eds11414,12615,6291617+
10−223.0511+200418_074945_D0415 G.eds477014,4389855187+
10−327.0236+200418_075046_D03ZYD.eds13,80115,025125026+
10−430.2578+200418_155345_D03JQP.eds(补)16,04416,0431516+
10−533.4663+200418_075522_D040C0.eds16,19016,180616+
10−635.6604+200418_075638_D040A9.eds15,87015,86978+
34110−122.7873+200418_075734_D03NQJ.eds484313,04510,2612059+
10−226.5013+200418_075845_D02PET.eds14,13115,6751793249+
10−330.0149+200418_080012_D03Z9U.eds15,46815,56213339+
10−433.4111+200418_080133_D04169.eds15,36615,21830178+
10−535.3413+200418_080234_D02NNZ.eds15,38115,378710+
10−6undetected_*200418_080402_D02PCR.eds15,94415,905140_
34410−118.5843+200418_080526_D02NON.eds5316515,18615,074+
10−222.1338+200418_080626_D02NP7.eds481711,98011,6704507+
10−326.1415+200418_080741_D02PCD.eds13,58914,3881121322+
10−429.7896+200418_080915_D02PDF.eds15,08815,17515063+
10−532.833+200418_081047_D02LBX.eds14,56714,5621419+
10−6undetected_200418_081157_D02NPA.eds16,32016,305318+

Note: * The “+” is positive, the “-” is negative.

Results on gradient dilution samples of confirmed COVID-19 patients by qPCR and dPCR. Note: * The “+” is positive, the “-” is negative. The CT value of qPCR was strongly correlated with the serial dilutions from 10−1 to 10-5 for the 4 samples (R2 = 0.99102 ± 0.00552). In the meantime, the FAM value of dPCR was also strongly correlated, but to a less degree than qPCR, with the serial dilutions (R2 = 0.9498 ± 0.04167) (Fig. 4 ). These results indicated that both dPCR and qPCR were able to accurately reflect the variation in concentration of SARS-CoV-2 samples.
Fig. 4

Evaluation of linearity using clinical samples by dPCR and qPCR assay. The linearity of CT value of qPCR, FAM positive of dPCR with the serial dilution point ranges from 10−1 to 10−6 for the #262, #265, #341 and #344 COVID-19 patient samples.

Evaluation of linearity using clinical samples by dPCR and qPCR assay. The linearity of CT value of qPCR, FAM positive of dPCR with the serial dilution point ranges from 10−1 to 10−6 for the #262, #265, #341 and #344 COVID-19 patient samples.

The specificity of dPCR assay in detecting non-SARS-CoV-2 specimens

To further validate the specificity of dPCR assay, the samples of influenza virus and other human coronaviruses of respiratory pathogen were collected and detected by dPCR. As shown in Table 2 , none of 13 tested samples containing influenza virus or other human coronavirus demonstrated positive results, indicating that dPCR possessed high specificity to distinguish the SARS-CoV-2 from other respiratory pathogens such as common influenza virus and other human coronaviruses.
Table 2

The test results of clinical samples with respiratory pathogen by dPCR.

Name of samplePathogensChipFAMResult
SVIRUS1H1pdm2009200327_015214_D02NP5.eds0-*
SVIRUS2Influenza-B200327_015325_D02LDH.eds2
SVIRUS3Influenza-H9N2200327_015425_D02PDB.eds1
S19105Rhinovirus200418_155601_D0409Q.eds2
SHAD1Parainfluenza-Ш200418_160706_D040W9.eds0
S19144Adenoviridae200418_160157_D0409O.eds1
SVIRUS4Measles virus200327_015525_D02PF0.eds0
SSAM1Rubella virus200327_015620_D02LCC.eds1
S19060hCoV-HKU1200327_015718_D02NNQ.eds0
S19111hCoV-NL63200418_155850_D0408 J.eds0
S19116hCoV-OC43200418_160059_D03ZT4.eds1
S19158HMPV200418_160608_D0409B.eds0
S19201SFTSV200418_160816_D0407 N.eds2

Note: *The “-” is negative.

The test results of clinical samples with respiratory pathogen by dPCR. Note: *The “-” is negative.

dPCR and qPCR assays on the “negative” samples from recurrent COVID-19 patients

Finally, to test whether dPCR can be applied to effectively detect the samples from suspected recurrent COVID-19 patients, a total of 18 “negative” samples from 17 such patients including 9 throat swabs, 6 sputum samples, and 3 anal swabs, were tested by dPCR and re-tested by qPCR. Two positive samples were used as controls (Number 1 and 2). The results of dPCR assay were contrastively analyzed with the qPCR results. As shown in Table 3 , 7 samples, including 1 (1/6, 16.7 %) sputum samples, 4 (4/9, 44.4 %) throat swab samples and 2 (2/3, 66.7 %) anal swab samples, were identified positive by dPCR assay (FAM ≥ 3) but not by qPCR assay while 2 throat swab samples (2/9, 22.2 %) were tested positive by qPCR assay (CT value < 35) but not by dPCR assay. These results demonstrated that the dPCR assay has an advantage in identifying the clinical samples with extremely low viral load of SARS-CoV-2 virus especially the samples from the recurrent patients and will be better ruling out the false negative results when used together with the qPCR assay in the clinic.
Table 3

The test results of samples from suspicious recurrent COVID-19 patients by qPCR and dPCR.

NumberSample typeqPCR (ORF1ab)qPCR (N)dPCR (FAM)
1#Sputum27.7525.8320
2#Sputum23.1923.21871
3Sputum-*0
4Sputum2
5Sputum1
6Sputum13
7Sputum37.1035.640
8Sputum0
9Throat swab1
10Throat swab34.8834.912
11Throat swab37.433
12Throat swab3
13Throat swab34.9933.042
14Throat swab35.2736.546
15Throat swab0
16Throat swab37.276
17Throat swab38.1062
18Anal swab6
19Anal swab35.6735.965
20Anal swab36.8537.411

Note: # Positive control samples; * The “-” is negative.

The test results of samples from suspicious recurrent COVID-19 patients by qPCR and dPCR. Note: # Positive control samples; * The “-” is negative.

Discussion

It was reported that around 60 % of SARS-CoV-2 infections were asymptomatic (Qiu, 2020). Moreover, due to the limitations of RT-PCR technology, only 30 %−60 % positive results could be obtained in COVID-19 patients that further confirmed by chest CT and other diagnostic aid (Wu and McGoogan, 2020). In addition, it is very important to acquire the true negative results from the convalescent COVID-19 patients who are about to discharge and out of quarantine, in order to avoid virus transmission and recurrence. Therefore, it is essential to find more sensitive method(s) to ensure the accuracy of detection of SARS-CoV-2 virus. Compared with qPCR, digital PCR quantification is less affected by poor amplification efficiency and inhibitors of amplification that may present in samples (Vasudevan et al., 2021; Falzone et al., 2020; Jiang et al., 2020). The process of sample partitioning also effectively concentrates template molecules within the micro reactions, improving analytical sensitivity for rare species by reducing competition between different targets for amplification reagents in the reaction mixture (Sanders et al., 2013; White et al., 2009; Pohl and Shih, 2004). In this study, we first tested the detection efficacy of dPCR and qPCR assays on low (3–6 copies) and middle (9–15 copies) simulated SARS-CoV-2 samples, and found that the two assays gave consistent results for both samples and dPCR was able to accurately reflect the copy number in the respective samples. We then compared the sensitivity of dPCR with qPCR on simulated and clinical sputum samples and the results showed that dPCR was more sensitive when the viral copy number is below 3 (lower threshold than qPCR) in the simulated or clinical samples. In the meantime, the serial dilutions with CT value and FAM value are both strongly correlated with the former slightly higher, indicating that both dPCR and qPCR could accurately reflect the variation in the concentration of SARS-CoV-2 virus in the samples. In regard to the testing specificity, dPCR obtained absolute negative signals (FAM value = 0) for 14/20 (70 %) of negative samples. However, it did give certain positive signals (1–2 copies per reaction) in 6/20 (30 %) negative samples (Fig. 3B). Nevertheless, when using 2 copies as the cutoff value, dPCR showed 100 % specificity, same as qPCR, in testing negative clinical samples (Fig. 3A). By using this threshold, dPCR could effectively distinguish the SARS-CoV-2 from other respiratory pathogens such as common influenza virus and other human coronaviruses (Table 3). In addition, three of 21 positive samples were not detected by dPCR while 5 of them were not detected by qPCR, which indicated that the sensitivity of dPCR is higher than qPCR in detecting SARS-CoV-2 (Fig. 3B). Notably, the FAM value of 3 “negative samples” was not absolute negative signals. It is suggested that we should attaches more importance to “negative samples” with FAM value signals. Recently, the recurrence of positive SARS-CoV-2 viral RNA in recovered COVID-19 patients is receiving more attention. Yuan B., et al. reported that 20 (10.99 %) out of 182 recovered COVID-19 patients were detected to be re-positive of SARS-CoV-2 RNA, although none showed any recurrent clinical symptoms (Yuan et al., 2020b). In the meantime, there is an argument on the possible reason for recurrence of COVID-19 patients. At present, it is believed that RNA negative conversion generally takes 2–3 weeks. A recent study indicated that SARS-CoV-2 nucleic acid existed in feces for nearly 50 days (Wu et al., 2020). Another recent study found that SARS-CoV-2 viral particles remained in the lungs of patients whose nasopharyngeal swab test results were negative at three consecutive times (Yao et al., 2020). These studies suggest that during the convalescent stage of the infection, the virus might remain in the patient at a very low level. In this study, we compared the difference of the two assays in detecting the “negative” samples from clinically recurrent COVID-19 patients. Seven “negative” samples, including 1 sputum samples, 4 throat swab samples and 2 anal swab samples, were identified positive by dPCR assay but not by qPCR assay. On the contrary, only 2 throat swab samples were tested positive by qPCR assay but not by dPCR assay. More importantly, most of the dPCR identified positive samples could not be detected even subjected to the second qPCR test. Therefore, our study clearly demonstrated that dPCR was superior over qPCR in detecting difficult or confusing samples with extremely low viral load from “recurrent” or convalescent COVID-19 patients, suggesting that the more sensitive assay, e.g. dPCR, is a powerful tool to help rule out the false negative results to avoid discharge of COVID-19 patients who might still have the virus becoming active source of the disease.

Conclusion

Our study demonstrated that in comparison with qPCR, dPCR is more sensitive in detecting both simulated and clinical SARS-CoV-2 samples with low viral load and can accurately reflect the variation in the copy number of viral RNA. The dPCR is also highly specific in distinguishing SARS-CoV-2 from other respiratory pathogens such as common influenza virus and other human coronaviruses. More importantly, dPCR shows superiority over qPCR in detecting samples from “recurrent” COVID-19 patients, which can be used as an important complement method for difficult or confusing samples in the clinical settings.

Funding

This work is funded by Special Project for Emergency Scientific and Technological Research on New Coronavirus Infection (YG, No. YD9110002001) supported by “The Fundamental Research Funds for the Central Universities” and Emergency Scientific and Technological Research Project for Novel Coronavirus Infection supported by Anhui Science and Technology Bureau (202004a07020002; 202004a0702000)

Authors' contributions

YS, CD, QC, JX, JY and CJ collected the epidemiological, and performed the experiments. YS, ZZ, HH, YG and WL contributed to clinical and laboratory data acquisition and analysis. JH, CD and YG drafted the manuscript. YS, JH and YG had the idea for the study and take responsibility for the integrity of the data and the accuracy of the data analysis.

Ethics approval and consent to participate

The study was approved by the Institutional Review Board Ethics Committee of University of Science and Technology of China. Written informed consent was obtained from each patient.

Declaration of Competing Interest

The authors have no conflicts of interests to declare.
  29 in total

Review 1.  Digital Assays Part I: Partitioning Statistics and Digital PCR.

Authors:  Amar S Basu
Journal:  SLAS Technol       Date:  2017-04-27       Impact factor: 3.047

2.  Efficacy and toxicity management of 19-28z CAR T cell therapy in B cell acute lymphoblastic leukemia.

Authors:  Marco L Davila; Isabelle Riviere; Xiuyan Wang; Shirley Bartido; Jae Park; Kevin Curran; Stephen S Chung; Jolanta Stefanski; Oriana Borquez-Ojeda; Malgorzata Olszewska; Jinrong Qu; Teresa Wasielewska; Qing He; Mitsu Fink; Himaly Shinglot; Maher Youssif; Mark Satter; Yongzeng Wang; James Hosey; Hilda Quintanilla; Elizabeth Halton; Yvette Bernal; Diana C G Bouhassira; Maria E Arcila; Mithat Gonen; Gail J Roboz; Peter Maslak; Dan Douer; Mark G Frattini; Sergio Giralt; Michel Sadelain; Renier Brentjens
Journal:  Sci Transl Med       Date:  2014-02-19       Impact factor: 17.956

3.  Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection.

Authors:  Luca Falzone; Nicolò Musso; Giuseppe Gattuso; Dafne Bongiorno; Concetta Ilenia Palermo; Guido Scalia; Massimo Libra; Stefania Stefani
Journal:  Int J Mol Med       Date:  2020-07-13       Impact factor: 4.101

4.  Recurrence of positive SARS-CoV-2 viral RNA in recovered COVID-19 patients during medical isolation observation.

Authors:  Bo Yuan; Han-Qing Liu; Zheng-Rong Yang; Yong-Xin Chen; Zhi-Yong Liu; Kai Zhang; Cheng Wang; Wei-Xin Li; Ya-Wen An; Jian-Chun Wang; Shuo Song
Journal:  Sci Rep       Date:  2020-07-17       Impact factor: 4.379

5.  Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China.

Authors:  Chaolin Huang; Yeming Wang; Xingwang Li; Lili Ren; Jianping Zhao; Yi Hu; Li Zhang; Guohui Fan; Jiuyang Xu; Xiaoying Gu; Zhenshun Cheng; Ting Yu; Jiaan Xia; Yuan Wei; Wenjuan Wu; Xuelei Xie; Wen Yin; Hui Li; Min Liu; Yan Xiao; Hong Gao; Li Guo; Jungang Xie; Guangfa Wang; Rongmeng Jiang; Zhancheng Gao; Qi Jin; Jianwei Wang; Bin Cao
Journal:  Lancet       Date:  2020-01-24       Impact factor: 79.321

6.  Pathological evidence for residual SARS-CoV-2 in pulmonary tissues of a ready-for-discharge patient.

Authors:  Xiao-Hong Yao; Zhi-Cheng He; Ting-Yuan Li; Hua-Rong Zhang; Yan Wang; Huaming Mou; Qiaonan Guo; Shi-Cang Yu; Yanqing Ding; Xindong Liu; Yi-Fang Ping; Xiu-Wu Bian
Journal:  Cell Res       Date:  2020-04-28       Impact factor: 25.617

7.  A Novel Coronavirus from Patients with Pneumonia in China, 2019.

Authors:  Na Zhu; Dingyu Zhang; Wenling Wang; Xingwang Li; Bo Yang; Jingdong Song; Xiang Zhao; Baoying Huang; Weifeng Shi; Roujian Lu; Peihua Niu; Faxian Zhan; Xuejun Ma; Dayan Wang; Wenbo Xu; Guizhen Wu; George F Gao; Wenjie Tan
Journal:  N Engl J Med       Date:  2020-01-24       Impact factor: 91.245

8.  Polymerase Chain Reaction Assays Reverted to Positive in 25 Discharged Patients With COVID-19.

Authors:  Jing Yuan; Shanglong Kou; Yanhua Liang; Jianfeng Zeng; Yanchao Pan; Lei Liu
Journal:  Clin Infect Dis       Date:  2020-11-19       Impact factor: 9.079

9.  Detection and quantification of SARS-CoV-2 by droplet digital PCR in real-time PCR negative nasopharyngeal swabs from suspected COVID-19 patients.

Authors:  Claudia Alteri; Valeria Cento; Maria Antonello; Luna Colagrossi; Marco Merli; Nicola Ughi; Silvia Renica; Elisa Matarazzo; Federica Di Ruscio; Livia Tartaglione; Jacopo Colombo; Chiara Grimaldi; Stefania Carta; Alice Nava; Valentino Costabile; Chiara Baiguera; Daniela Campisi; Diana Fanti; Chiara Vismara; Roberto Fumagalli; Francesco Scaglione; Oscar Massimiliano Epis; Massimo Puoti; Carlo Federico Perno
Journal:  PLoS One       Date:  2020-09-08       Impact factor: 3.240

10.  Digital PCR is a sensitive new technique for SARS-CoV-2 detection in clinical applications.

Authors:  Yanfang Jiang; Haifeng Wang; Sijia Hao; Yukun Chen; Jiaxue He; Yong Liu; Liguo Chen; Yuanhua Yu; Shucheng Hua
Journal:  Clin Chim Acta       Date:  2020-11-04       Impact factor: 3.786

View more
  5 in total

1.  High concordance in SARSCoV-2 detection between automated (Abbott m2000) and manual (DaAn gene) RT-PCR systems: The EDCTP PERFECT-Study in Cameroon.

Authors:  Nadine Nguendjoung Fainguem; Joseph Fokam; Ezechiel Ngoufack Jagni Semengue; Alex Durand Nka; Désiré Takou; Joshua Ageboh Nkembi-Leke; Claudia Alteri; Luna Colagrossi; Roméo Bouba Yagai; Collins Ambe Chenwi; Michel Carlos Tchouaket Tommo; Grace Angong Beloumou; Aude Christelle Ka'e; Sandrine Claire Ndjeyep Djupsa; Aissatou Abba; Laeticia Grace Heunko Yatchou; Krystel Nnomo Zam; Rachel Kamgaing; Samuel Martin Sosso; Lucien Mama; Nicaise Ndembi; Vittorio Colizzi; Carlo-Federico Perno; Giulia Cappelli; Alexis Ndjolo
Journal:  J Public Health Afr       Date:  2022-05-24

Review 2.  Diagnostic, Prognostic, and Therapeutic Value of Droplet Digital PCR (ddPCR) in COVID-19 Patients: A Systematic Review.

Authors:  Angela Ishak; Mousa M AlRawashdeh; Stepan M Esagian; Ilias P Nikas
Journal:  J Clin Med       Date:  2021-12-06       Impact factor: 4.241

Review 3.  Recent advances in point of care testing for COVID-19 detection.

Authors:  Renata Salgado Fernandes; Juliana de Oliveira Silva; Karina Braga Gomes; Ricardo Bentes Azevedo; Danyelle M Townsend; Adriano de Paula Sabino; Andre Luis Branco de Barros
Journal:  Biomed Pharmacother       Date:  2022-08-12       Impact factor: 7.419

Review 4.  Digital PCR Applications in the SARS-CoV-2/COVID-19 Era: a Roadmap for Future Outbreaks.

Authors:  Raphael Nyaruaba; Caroline Mwaliko; David Dobnik; Pavel Neužil; Patrick Amoth; Matilu Mwau; Junping Yu; Hang Yang; Hongping Wei
Journal:  Clin Microbiol Rev       Date:  2022-03-08       Impact factor: 50.129

5.  Micro-injection Molded Droplet Generation System for Digital PCR Application.

Authors:  Daae Jo; So Young Kim; Hyeon Woo Kang; Sung Han Pyo; Nam Kyu Jeong; Nam Ho Bae; Seok Jae Lee; Yong Tae Kim; Kyoung G Lee
Journal:  Biochip J       Date:  2022-09-06       Impact factor: 4.229

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.