| Literature DB >> 34047701 |
Akira-Sebastian Poncette1,2, Maximilian Markus Wunderlich2, Claudia Spies1, Patrick Heeren1,2, Gerald Vorderwülbecke1, Eduardo Salgado1,2, Marc Kastrup1, Markus A Feufel3, Felix Balzer1,2.
Abstract
BACKGROUND: As one of the most essential technical components of the intensive care unit (ICU), continuous monitoring of patients' vital parameters has significantly improved patient safety by alerting staff through an alarm when a parameter deviates from the normal range. However, the vast number of alarms regularly overwhelms staff and may induce alarm fatigue, a condition recently exacerbated by COVID-19 and potentially endangering patients.Entities:
Keywords: ICU; alarm fatigue; alarm management; alarm system; alarm system quality; clinical alarms; data science; digital health; intensive care unit; medical devices; patient monitoring; patient safety; technological innovation
Mesh:
Year: 2021 PMID: 34047701 PMCID: PMC8196351 DOI: 10.2196/26494
Source DB: PubMed Journal: J Med Internet Res ISSN: 1438-8871 Impact factor: 5.428
Data analysis framework applied in this study in line with the quality dimensions introduced by Hüske-Kraus et al [12] and including metrics suggested by Hüske-Kraus et al [12] as well as metrics suggested by other sources for each dimension, wherever possible.
| Quality dimension | Definition | Metrics used in this study |
| Alarm load | Metrics related to the number of alarms | Alarms per bed per day, frequency of individual alarms, alarms per device, alarms per criticality (red, yellow, and blue; ie, alarm at high criticality, alarm at medium criticality, and technical alarm at low criticality, respectively), average temporal distribution of alarms and alarm flood conditions (10 or more alarms occurring within 10 minutes) [ |
| Avoidable alarms | False-positive alarms, nonactionable alarms, and technical alarms | Technical alarms per bed per day, technical alarms per device |
| Responsiveness and alarm handling | Alarm duration, response time, muting of alarms, and corrective actions | Duration of alarms |
| Sensing | The quality of the technical infrastructure, such as consumable, overmonitoring, and undermonitoring | Average usage of the alarm pause function per bed per day, proper pause-to-pause ratio [ |
| Exposure | How alarms are distributed in the unit | Average alarm frequencies per room and per bed per room type, number of beds issuing more alarms than the average |
Figure 1Feedback loop regarding do-it-yourself (DIY) instructions for self-analysis of patient monitoring alarm data in the intensive care unit.
Figure 2Frequency of individual alarm parameters within 93 days. The colors correspond to the alarm criticalities (red, yellow, and blue). *: ventilator arm; ABPs: systolic arterial blood pressure; ECG: electrocardiogram; FREQUENCY: ventilator alarm indicating that the upper respiratory rate threshold has been exceeded; HR: heart rate; RR: respiratory rate derived from the ECG (see Multimedia Appendix 2 for all abbreviations).
Figure 3Alarms from medical devices within 93 days subdivided into the criticality levels (red, yellow). ECG: electrocardiogram; IBP: invasive blood pressure; ICP: intracranial pressure; NIBP: noninvasive blood pressure; SpO2: oxygen saturation.
Figure 4Average distribution of alarms across 24 hours. The white spaces between the grey bars (ie, shifts) visualize handover periods. Each dot shows the average alarm frequency of 1 minute for the specified device. The line for each device is calculated by ggplot2’s smoothing function and represents a generalized additive model of the distribution (with the formula y ~ s(x, bs = "cs"). It serves to aid in detecting trends in the data. ECG: electrocardiogram; IBP: invasive blood pressure.
Figure 5Temporal distribution of alarm flood conditions over 24 hours. Each dot indicates the sum of all alarm flood conditions that were initiated at the respective time of day in 10-minute intervals. For example, the first dot on the far left indicates that 43 alarm floods occurred between 7:10 and 7:20 AM across all days in the data. The blue line is a local regression, calculated by ggplot2’s smoothing function (formula: y ~ x). The white spaces between the grey bars (ie, shifts) visualize handover periods.
Figure 6Median alarm duration of the 3 medical devices that issue most alarms over 24 hours. Each dot represents the median alarm duration for each minute of the day of the respective device. The line for each device is based on ggplot2’s smoothing function and represents a generalized additive model of the distribution (with the formula y ~ s(x, bs = "cs"). The white spaces between the grey bars (ie, shifts) visualize handover periods. ECG: electrocardiogram; IBP: invasive blood pressure.
Figure 7The median alarm duration from 8 medical devices plotted against the total number of alarms issued by the respective device. The colors correspond to the alarm criticalities (red, yellow, and blue). ECG: electrocardiogram; IBP: invasive blood pressure; ICP: intracranial pressure; NIBP: noninvasive blood pressure; SpO2: oxygen saturation.