| Literature DB >> 34040659 |
Roy Gigi1, Jacob Mor2, Inbar Lidor3, Dror Ovadia1, Eitan Segev1.
Abstract
PURPOSE: Several hexapod external fixators are used in the treatment of bone fracture and deformity corrections. One characteristic of all of them is the requirement for manual adjustment of the fixator struts. The purpose of this study was to introduce a novel robotic system that executes automatic adjustment of the struts.Entities:
Keywords: OrthoSpin; automatic struts; bone deformities; hexapod external fixation
Year: 2021 PMID: 34040659 PMCID: PMC8138786 DOI: 10.1302/1863-2548.15.210063
Source DB: PubMed Journal: J Child Orthop ISSN: 1863-2521 Impact factor: 1.548
Fig. 1.Auto Strut device assembled on a circular fixator: a) Auto Strut device; b) assembled on patient’s leg.
Fig. 2.Schematic diagram of the Auto Strut system. A control unit is placed onto the frame of the fixator. It is connected by a Universal Serial Bus cable to the external computer for downloading the treatment plan, prior to treatment initiation.
Demographics
| Patient | Age, yrs | Sex | Weight, kg | Height, cm | Aetiology |
|---|---|---|---|---|---|
| 01 | 14 | M | N/A | N/A | Idiopathic tibia vara |
| 02 | 12 | F | 47.5 | N/A | Hemihypertrophy |
| 03 | 17 | F | 60 | 161 | Fibular hemimelia (Paley classification 3B2) with congenital short femur |
| 04 | 14 | F | 68 | 175 | Infantile Blount’s disease |
| 05 | 16 | M | 85 | 165 | Idiopathic tibia vara |
| 06 | 14 | M | 86.4 | 180 | Idiopathic tibia vara |
| 07 | 20 | F | 60 | 164 | Post-traumatic equinus due to degloving injury around the ankle (soft-tissue distraction) |
| 08 | 14 | M | 70 | 167 | Ollier syndrome - tibia valga |
| 09 | 19 | F | 68 | 120 | Achondroplasia – tibia vara bilateral |
| 10 | 18 | M | 100 | 175 | Idiopathic tibia vara |
All patients were treated for tibia bone deformity correction except for patient 07 who underwent soft-tissue distraction due to fixed ankle equinus deformity
N/A, not available
Planned treatment and outcomes
| Patient | Treated limb | Treatment period | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Deformity correction | Distraction phase | Consolidation phase | ||||||||
| Axial – lengthening, mm short | Axial – angulation, internal/ external | AP – angulation, varus/valgus | Lateral – angulation, posterior/ anterior | Duration, days | Index, days/cm | Length, cm | Time to consolidation, days | Consolidation index, days/cm | ||
| 01 | Left | 30 | - | 15° varus | - | 24 | 8 | 3 | 93 | 31 |
| 02 | Left | 40 | - | 5° varus | - | N/A | N/A | 4 | N/A | N/A |
| 03 | Left | 60 | - | - | 90 | 15 | 6 | 114 | 19 | |
| 04 | Left | 25 | 5° internal | 17° varus | - | 33 | 13.2 | 2.5 | 62 | 24.8 |
| 05 | Left | 35 | - | 15° varus | - | 35 | 10 | 3.5 | 64 | 18.3 |
| 06 | Left | 10 | 15° internal | 20° varus | - | 10 | 10 | 1 | 108 | 108 |
| 07 | Right | 10 | - | - | 50° plantar flexion | 49 | - | 1 | N/A | N/A |
| 08 | Left | 60 | - | 16° valgus | 5° apex anterior | 63 | 10.5 | 6 | 135 | 22.5 |
| 09 | Right | 60 | 5° internal | 15° varus | 6° apex anterior | 60 | 10 | 6 | 128 | 21.3 |
| Left | 60 | 5° internal | 15° varus | 6° apex anterior | 60 | 10 | 6 | 128 | 21.33 | |
| 10 | Right | 20 | 5° internal | 15° varus | - | 24 | 12 | 2 | - | - |
AP, anteroposterior; N/A, not available
Fig. 3.Automatic strut adjustment versus planned readings of a representative patient (patient 08). The actual automatic adjustment (blue) is presented against the office visits readings (orange) for each of the six struts of the hexapod external fixator. Readings of the scale numbers of the struts were carried out on six time points during the treatment course (x-axis). Y-axis represents number on the struts.
Fig. 4.Radiographs before surgery and during distraction of bilateral tibia vara with Auto Strut Hexapod external fixation system in 19-year-old achondroplasia patient: a) preoperative leg-length radiograph of 19-year-old achondroplasia patient for deformity correction and lengthening procedure; b) radiograph after 60 days of distraction for bilateral tibial genuvarum correction and lengthening, using Auto Struts over an circular cexapod external fixator system.