Yang Bai1,2, Connie Tompkins2, Nancy Gell2, Dakota Dione3, Tao Zhang4, Wonwoo Byun1.
Abstract
OBJECTIVES: The aim of this study was to evaluate the accuracy of three consumer-based activity monitors, Fitbit Charge 2, Fitbit Alta, and the Apple Watch 2, all worn on the wrist, in estimating step counts, moderate-to-vigorous minutes (MVPA), and heart rate in a free-living setting.Entities:
Year: 2021 PMID: 34038458 PMCID: PMC8153432 DOI: 10.1371/journal.pone.0251975
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Descriptive demographics of the participants (N = 48).
| Mean | Standard Deviation | Minimum | Maximum | |
|---|---|---|---|---|
| Age (years) | 26.8 | 11.9 | 18.0 | 59.0 |
| Stature (cm) | 168.4 | 10.4 | 147.0 | 190.5 |
| Body mass (kg) | 66.3 | 13.0 | 42.6 | 105.4 |
| Body Mass Index (kg/m2) | 23.2 | 3.0 | 17.9 | 32.9 |
| Body Fat Percentage | 20.5 | 8.5 | 7.7 | 36.1 |
| Systolic Pressure (mmHg) | 112.1 | 13.5 | 87.7 | 147.0 |
| Diastolic Pressure (mmHg) | 70.6 | 9.6 | 53.7 | 95.0 |
| Heart Rate (bpm) | 78.8 | 49.3 | 49.0 | 93.7 |
Validity of steps and moderate-to-vigorous minutes estimation from three consumer monitors and two research monitors (N = 48).
| Mean (SD) | MPE (SD) | MAPE (SD) | RMSE | Correlation | |
|---|---|---|---|---|---|
| Steps/day | |||||
| Yamax SW-200 | 10906 (5166) | ||||
| Fitbit Alta | 11922 (5348) | -15.2% (35.2%) | 20.7% (32.2%) | 2141 | 0.94 |
| Fitbit Charge 2 | 11734 (5413) | -10.3% (23.4%) | 17.1% (18.9%) | 1989 | 0.94 |
| Apple Watch 2 | 11550 (5248) | -11.8% (33.2%) | 20.3% (28.7%) | 2055 | 0.94 |
| MVPA without 10 minutes bout in minutes/day | |||||
| ActiGraph | 87.1 (52.2) | ||||
| Fitbit Alta | 64.0 (46.9) | -26.0% (366.7%) | 92.3% (355.5%) | 32 | 0.90 |
| Fitbit Charge 2 | 72.2 (56.0) | -35.8% (319.0%) | 91.0% (307.5%) | 42 | 0.74 |
| Apple Watch 2 | 53.3 (33.5) | -21.4% (120.3%) | 58.9% (106.7%) | 48 | 0.67 |
| MVPA with 10 minutes bout in minutes/day | |||||
| ActiGraph | 47.8 (41.1) | ||||
| Fitbit Alta | 64.0 (46.9) | 32.0% (50.0%) | 44.7% (38.7%) | 26 | 0.91 |
| Fitbit Charge 2 | 72.2 (56.0) | 50.1% (88.9%) | 67.3% (76.3%) | 45 | 0.77 |
| Apple Watch 2 | 53.3 (33.5) | 26.2% (74.6%) | 55.4% (55.7%) | 31 | 0.66 |
Note: MPE = Mean Percent Error; MAPE = Mean Absolute Percent Error; RMSE = Root Mean Square Error; SD = Standard Deviation
** Correlation is significant at an alpha level of 0.01 (2-tailed)
* Correlation is significant at an alpha level of 0.05 (2-tailed).
Equivalence Zone from criterion measure was underlined.
Validity of heart rate estimation from Apple Watch 2 and Fitbit Charge 2.
| No. of minutes | Polar Mean (SD) | Monitor Mean (SD) | MPE (SD) | MAPE (SD) | RMSE | Correlation | |
|---|---|---|---|---|---|---|---|
| Sedentary | 7811 | 61.5 (11.5) | 59.6 (10.6) | 2.7% (7.1%) | 4% (6.1%) | 8.94 | 0.90 |
| Light PA | 6559 | 74.8 (12.2) | 70.7 (13.7) | 4.9% (14%) | 10% (10%) | 11.34 | 0.70 |
| MVPA | 618 | 125.6 (18.0) | 108.9 (17.7) | 13% (13%) | 14% (12%) | 22.33 | 0.49 |
| Sedentary | 2444 | 66.6 (12.3) | 66.7 (11.8) | -1.1% (13%) | 7% (11%) | 5.33 | 0.73 |
| Light PA | 2075 | 75.6 (12.0) | 73.5 (11.8) | 1.8% (15%) | 10% (11%) | 10.97 | 0.56 |
| MVPA | 590 | 94.3 (19.3) | 83.9 (19.7) | 9.6% (19%) | 16% (14%) | 24.54 | 0.49 |
Note: PA = Physical Activity; MVPA = Moderate to Vigorous Physical Activity; MPE = Mean Percent Error; MAPE = Mean Absolute Percent Error; RMSE = Root Mean Square Error; SD = Standard Deviation.
** Correlation is significant at an alpha level of 0.01 (2-tailed).
*Correlation is significant at an alpha level of 0.05 (2-tailed).
Fig 1Agreement between criterion and comparison monitors on 95% equivalence testing.
Fig 2Bland-Altman plots comparing steps between Yamax and research and consumer activity monitors.
Fig 4Bland-Altman plots comparing heart rate between Polar H7 and Apple Watch 2 and Fitbit Charge 2.
Fig 3Bland-Altman plots comparing moderate-to-vigorous physical activity between ActiGraph and three consumer activity monitors with- and without bout criterion.