Investigation of Elekta AQUA software for kilovoltage to megavoltage radiation isocenter coincidence.

Craig Norvill1, Felix Bockelmann2, Patrick Blumenberg3, Simon Goodall3,4.   

Abstract

Elekta AQUA v2.02 software (Gantry Runout isocenter test) was investigated as a tool for verification of kilovoltage to megavoltage gantry radiation isocenter coincidence. AQUA reported megavoltage (6 MV) isocenter was independent of field size over the range 5 cm × 5 cm to 20 cm × 20 cm. For the 10 cm × 10 cm field size, standard deviation in AQUA reported 3D megavoltage (6MV) isocenter over ten consecutive deliveries was less than 0.04 mm for any axis. Compared to the full AQUA test delivery (Gantry Runout), the shorter AQUA test version (Gantry Runout short) gave a root mean square MV isocenter (± 1 SD) difference of 0.18 mm ± 0.08. Across 7 machines, root mean square differences between AQUA and PIPS PRO reported MV isocenter (for 6 MV and 6 MV FFF beams) was 0.1 mm ± 0.1 mm, with most of the difference observed in the gun-target (Y-axis). AQUA 6 MV isocentre position was offset to gantry relative to Elekta XVI customer acceptance test (CAT) workflow by 0.25 mm to 0.52 mm. For 6 MV FFF beams, AQUA reported an MV isocenter of between 0.27 and 0.39 mm offset to target relative to Elekta CAT.

Keywords:  AQUA; Gantry; Isocenter; Quality assurance

Year:  2021        PMID: 34033014     DOI: 10.1007/s13246-021-01014-1

Source DB:  PubMed          Journal:  Phys Eng Sci Med        ISSN: 2662-4729


  1 in total

1.  Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi-leaf collimator quality assurance.

Authors:  Simon K Goodall; Craig Norvill
Journal:  J Appl Clin Med Phys       Date:  2022-06-06       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.