Literature DB >> 3400852

Cuff failure in polyvinyl chloride tracheal tubes sprayed with lignocaine.

A J Walmsley1, L M Burville, T P Davis.   

Abstract

The incidence of tracheal tube cuff rupture was noted in 30 polyvinyl chloride tracheal tubes lubricated with three different solutions. All cuffs moistened with water were intact after 2 hours of cuff inflation whereas two of 10 lubricated with 4% lignocaine solution had burst. Both of these had leaked at the site of cuff attachment to the tube. Fifty percent of tubes lubricated with Astra lignocaine spray burst during the study. Four of the five had developed pinholes in the cuffs themselves. The remaining 50% of this group showed marked distortion and thinning of their intact cuff walls. The implications of these findings are discussed in view of the widespread use of PVC tracheal tubes.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3400852     DOI: 10.1111/j.1365-2044.1988.tb09024.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

1.  Intracuff alkalized lidocaine reduces sedative/analgesic requirements for mechanically ventilated patients.

Authors:  Ahmed Sobhy Basuni
Journal:  Saudi J Anaesth       Date:  2014-10

Review 2.  Effect of Intracuff Lidocaine on Postoperative Sore Throat and the Emergence Phenomenon: A Systematic Review and Meta-Analysis of Randomized Controlled Trials.

Authors:  Fai Lam; Yu-Cih Lin; Hsiao-Chien Tsai; Ta-Liang Chen; Ka-Wai Tam; Chien-Yu Chen
Journal:  PLoS One       Date:  2015-08-19       Impact factor: 3.240

  2 in total