Literature DB >> 34007233

Trichromatic Enhanced Dynamic Color Screening on the PDI Check Nintendo 3DS Game.

Andrew Arnold1, Kyle Smith2, Aaron Molina2, Alex Damarjian2, Tarah Desatoff3, Robert Arnold4.   

Abstract

INTRODUCTION: Classification of color deficiency has required substantial time and expense with the static Farnsworth-Munsell and Innova Rabin tests. Therefore, dynamic color tests were developed for the Nintendo 3DS.
METHODS: Fifteen color deficient patients and 17 age-matched normals performed Rabin color test in addition to PDI Check dynamic color games resembling Farnsworth-Munsell presentation (version 0.2.8) and 3-color iso-luminance gray (version 0.2.13).
RESULTS: Tests of red, green and blue cone-deficient with the v0.2.8 had sensitivity/specificity/PPV of 92%/86%/92% protanopes, 78%/90%/88% deutanopes and 87%/50%/93% tritanopes. Version 0.2.13 had sens/spec/PPV of 78%/83%/78% red-cone, 100%/85%/80% green cone and 67%/78%/33% blue cone. Corresponding IntraClass Correlation (ICC) utilizing v0.2.8 were red-cone 0.22 (-0.02-0.60), green-cone 0.34 (-0.10-0.67) and blue-cone 0.38 (0.12-0.75). ICC for v0.2.13 was higher with protanope 0.62 (-0.07-0.87), deuteranope 0.64 (-0.09-0.88) and tritanope 0.31 (-0.07-0.70). The PDI Check color game took 65 seconds compared to 197 seconds for Innova Rabin.
CONCLUSION: The PDI Check color game quickly identifies patients with inherited color deficiencies.
© 2021 Arnold et al.

Entities:  

Keywords:  Farnsworth-Munsell; Rabin color cone test; color blindness; deuteranomaly; dynamic; protanomaly; vision screening

Year:  2021        PMID: 34007233      PMCID: PMC8124021          DOI: 10.2147/OPTO.S305042

Source DB:  PubMed          Journal:  Clin Optom (Auckl)        ISSN: 1179-2752


Plain Language Summary

In addition to visual acuity, peripheral vision and depth perception, human eyes can detect a range of color. Three different sets of cones in the retina sense red, green and blue tones to provide a full spectrum. Individuals, mainly males, with inherited color blindness have decreased numbers or performance of either their green, their red or rarely their blue cones. Formal tests to identify and quantify color blindness take substantial time and money. An inexpensive game on the Nintendo 3DS console presents certain colored objects to either right or left eye gradually changing from subtle to obvious allowing quick identification of the three main types of color blindness. Compared Color Vision Tests on the PDI Check Game Against the Innova Rabin Trichromatic Standard Test Abbreviations: ICC, intraclass correlation; F-M, Farnsworth-Munsell.

Introduction

Accurate characterization of color defects is important in the diagnosis and management of inherited and acquired retinal and neuro-ophthalmic conditions. Adult and pre-literate matching cards are often used clinically, however more complex tests are required for detailed diagnosis and quantification of color deficiency.1 Many tests are expensive and require significant time, understanding2 and cooperation. Smartphone presentation can substitute for flip-book Ishihara color plates.3 Utilizing the autostereoscopic barrier screen of the Nintendo 3DS XL console, near vision screening testing games have been developed.4 Near monocular visual acuity, stereopsis and color vision can be screened in normal military staff and also in remote, non-English speaking villagers.5 PDI Check correlates well with conventional near vision tests and helps identify ocular suppression both through stereo deficiency, but also with autostereoscopically presented, rivalry near optotypes.6 Champion shotgun shooters demonstrated extraordinary red-cone and blue cone detection with PDI Check compared to age-matched normal subjects.7 We undertook this study to analyze two types of dynamic, quick trichromatic separated color tests on the PDI Check and then compare them with industry standard, static, calibrated, computer delivered Innova Rabin system.1,8

Methods

This study has IRB approval through Pacific University [109-17; 1098158-5]. It complies with the Declarations of Helsinki and the Health Insurance Portability and Accountability Act. Patients and/or their guardians provided written informed consent. De-identified data can be found: . Color deficient individuals were identified through Ishihara testing during eye examinations. Normal subjects were recruited from clinical and research staff. Each performed Innova Rabin’s Cone Contrast color test8 following standard procedure monocularly with the non-tested eye patched. PDI Check has two different dynamic color testing games; two different dynamic color tests were programmed for the Nintendo 3DS XL. One (“FM,” version 0.2.8) presents four forced choice identical discs on the upper screen starting at the cone confusion regions of an FM 100 test, then one random quadrant gradually deviates around the spectrum for the subject to identify by clicking the lower screen. The second type of test (“gray disc,” version 0.2.13) starts with four iso-luminance gray discs against a neutral gray background. Then the pixels directed at one eye, in one random quadrant transform into a gradually intensifying teal, pink or a purple color for the subject to select as soon as the difference is perceived. Some subjects performed comparison of both PDI Check color versions. The Nintendo 3DS console automatically adjusts background screen luminance to an aesthetically pleasing viewing level. The PDI Check game on Nintendo 3DS and the Innova Rabin tests were viewed from 30 cm. Given a mean of 95 and standard deviation of 10, the sample size to detect a difference of 15 given alpha 0.05 and Power 0.80 is 7.

Results

Fifteen color deficient individuals were identified through Ishihara testing during comprehensive eye examinations. Ages ranged from 7 to 50 years. According to Innova Rabin scores, the hue was predominantly red deficient in 6, green deficient in 8 and blue deficient in one. Thirteen of the fifteen color deficient were males. Seventeen normal individuals aged 11 to 60 served as controls of which 7 were males. All subjects scored at least 20/25 on near testing. The mean and standard deviation combined right and left color score in Innova Rabin for normals was 97±1 (range 96–98) compared to 81.4±9 (range 61–88) for PDI Check. The corresponding values for any color deficiency Innova Rabin was 65 ± 10 (range 47–77) and for PDI Check 66±16 (range 31–80). For n=7 green deficient, Innova Rabin was 67±8 (range 58–77) and for PDI Check 68±13 (range 41–80). The time to perform patched Innova Rabin on both eyes averaged 197 seconds (range 162 to 260 seconds) while time to complete un-patched, monocular PDI Check dynamic gray-game for both eyes averaged 65 seconds (range 40–120 seconds). The test re-Test reliability for version 0.2.8 color was 54±7 for red, 60±6 for green and 63±7 for blue; essentially the deviation was 10% of the mean. In comparison, PDI stereo can perform with 3% deviation compared to the mean. Bland Altman plots for normals and color deficient with version 0.2.13 are shown in Figure 1. The mean differences for normals was 16 ± 3 (SEM) with PDI Check normals reporting the lower value. Corresponding IntraClass Correlation (ICC) utilizing v0.2.8 were red-cone 0.22 (−0.02–0.60), green-cone 0.34 (−0.10–0.67) and blue-cone 0.38 (0.12–0.75; Table 1)Table 1. ICC for v0.2.13 were higher with protanope 0.62 (−0.07–0.87), deuteranope 0.64 (−0.09–0.88).
Figure 1

Bland Altman analysis of color scores comparing PDI Check gray disc with Innova Rabin trichromatic scores averaged with both eyes for age-matched normal and patients with inherited color deficiencies.

Table 1

Compared Color Vision Tests on the PDI Check Game Against the Innova Rabin Trichromatic Standard Test

VersionICCUp ICCLow ICCMean DiffLowUpSESensitivitySpecificityPPV
0.2.8F-M red0.22−0.020.67.66−48.964.26.692%86%92%
F-M green0.34−0.10.678.89−51.669.47.178%90%88%
F-M blue0.380.120.75−33.7−77.410.15.387%50%93%
0.2.13Iso-gray red0.62−0.070.87−27%−81%28%4%78%83%78%
Iso-gray green0.64−0.090.8828%−77%21%4%100%85%80%
Iso-gray blue0.31−0.070.7−22%−40%−4%2%67%78%33%

Abbreviations: ICC, intraclass correlation; F-M, Farnsworth-Munsell.

Bland Altman analysis of color scores comparing PDI Check gray disc with Innova Rabin trichromatic scores averaged with both eyes for age-matched normal and patients with inherited color deficiencies. Tests of red, green and blue cone-deficient with the v0.2.8 had sensitivity/specificity/PPV of 92%/86%/92% for protanopes, 78%/90%/88% deutanopes and 87%/50%/93% for the tritanope. Version 0.2.13 had sensitivity/specificity/PPV of 78%/83%/78% red-cone and 100%/85%/80% green cone.

Discussion

After gradual recruitment from a pediatric eye clinic, patients with inherited color deficiency and age-matched normal subjects had compared digital trichromatic evaluation on the same day. Each had industry- and US Military standard INNOVA Rabin computerized, timed static color quantification with non-tested eye patched compared to non-patched, dynamic, autostereoscopic monocular presentation on a portable, Nintendo 3DS XL game console. There was a good correlation and the dynamic presentation was quicker. Speed was not the issue when multiple digital color tests were compared in 68 color defective individuals without color-normal controls.9 Insufficient time can be an issue with the D15 type color tests.10 A limitation of this study was relatively small sample size. It is not simple to collect large numbers of color deficient individuals due to the relative rarity of the condition.5 We did not yet include patients with acquired color defects. We defined color cone deficiency by Innova Rabin method that may not exactly represent FM 100 or anomaloscope. Not all our subjects were able to complete both PDI Check versions on the different test phases, but an advantage of our study was head-to-head comparison of two digital tests one the same day with identical lighting. The scores for normal color in PDI Check under-estimated Innova Rabin values therefore further mathematical modeling is underway. PDI Check continues to refine the output and presentation of the game to better provide useful clinical data. The autostereoscopic screen on the Nintendo 3DS XL allows monocular rapid screening for color defects in addition to monocular near dynamic acuity and stereopsis.

Conclusion

PDI Check in a quick, dynamic, forced-choice monocular presentation was able to quickly discriminate color deficient from normal vision individuals.
  10 in total

1.  Quantification of color vision using a tablet display.

Authors:  Alicia Chacon; Jeff Rabin; Dennis Yu; Shawn Johnston; Timothy Bradshaw
Journal:  Aerosp Med Hum Perform       Date:  2015-01       Impact factor: 1.053

2.  Limitations and Precautions in the Use of the Farnsworth-Munsell Dichotomous D-15 Test.

Authors:  Stephen J Dain; David A Atchison; Jeffery K Hovis
Journal:  Optom Vis Sci       Date:  2019-09       Impact factor: 1.973

3.  Validity and Acceptance of Color Vision Testing on Smartphones.

Authors:  Omar K Ozgur; Trisha S Emborgo; Mark B Vieyra; Rebecca F Huselid; Rudrani Banik
Journal:  J Neuroophthalmol       Date:  2018-03       Impact factor: 3.042

4.  Predicting the Farnsworth-Munsell D15 and Holmes-Wright-A lantern outcomes with computer-based color vision tests.

Authors:  Ali Almustanyir; Jeffery Hovis; Mackenzie G Glaholt
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2020-04-01       Impact factor: 2.129

5.  Performance of a Quick Screening Version of the Nintendo 3DS PDI Check Game in Patients With Ocular Suppression.

Authors:  Kyle A Smith; Andrew W Arnold; Jacob H Sprano; Stephanie L Arnold; Robert W Arnold
Journal:  J Pediatr Ophthalmol Strabismus       Date:  2019-07-01       Impact factor: 1.402

6.  Color vision testing in young children: a review.

Authors:  K R Alexander
Journal:  Am J Optom Physiol Opt       Date:  1975-05

Review 7.  Review of the main colour vision clinical assessment tests.

Authors:  A Fanlo Zarazaga; J Gutiérrez Vásquez; V Pueyo Royo
Journal:  Arch Soc Esp Oftalmol (Engl Ed)       Date:  2018-10-22

8.  Calibrated measurement of acuity, color and stereopsis on a Nintendo® 3DS™ game console.

Authors:  Kyle A Smith; Alex G Damarjian; Aaron Molina; Robert W Arnold
Journal:  Clin Optom (Auckl)       Date:  2019-04-30

9.  Compared Near-Vision Testing With the Nintendo 3DS PDI Check Game on the Thai-Burma Border.

Authors:  Samuel J Martin; Kayla S Rowe; Nay Hser; Hser Eh Htoo; Ray Khin; Kyle A Smith; Robert W Arnold
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2019 Jul-Aug

10.  Extraordinary Dynamic Near Vision in Champion Shotgun Shooters: PDI Check Evaluation of Stereo and Color without Floor Effect.

Authors:  Steven Brown; Clayton Rue; Kyle Smith; Robert Arnold
Journal:  Clin Ophthalmol       Date:  2021-02-16
  10 in total
  2 in total

1.  Pediatric Non-Refractive Vision Screening with EyeSwift, PDI Check and Blinq: Non-Refractive Vision Screening with Two Binocular Video Games and Birefringent Scanning.

Authors:  Tsaina Mahlen; Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2022-02-11

2.  Multifaceted Amblyopia Screening with blinq, 2WIN, and PDI Check.

Authors:  Robert Arnold; Mario Angi
Journal:  Clin Ophthalmol       Date:  2022-02-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.