Puhong Zhang1,2, Xi Li3, Zhe Fang1, Yanling Lu3, Jingchen Cui4, Xin Du1, Rong Hu3.
Abstract
BACKGROUND: Accurate measurement of blood pressure (BP) is crucial to hypertension control and prevention of future stroke and heart attack. All BP measuring devices must be validated independently in the clinical setting.Entities:
Mesh:
Year: 2021 PMID: 34001755 PMCID: PMC8568324 DOI: 10.1097/MBP.0000000000000547
Source DB: PubMed Journal: Blood Press Monit ISSN: 1359-5237 Impact factor: 1.430
Participant recruitment details of three devices
| Screening and recruitment |
| Recruitment ranges | ||||
|---|---|---|---|---|---|---|
| BP range and level (mmHg) |
| n on Rx | ||||
| Omron HEM-7120 | ||||||
| Total screened | 44 | SBP | Low | <90 | 0 | 4 |
| Total excluded | 11 | 90–129 | 10 | |||
| Ranges complete | 1 | Medium | 130–160 | 12 | 9 | |
| Range adjustment | 0 | High | 161–180 | 8 | 8 | |
| Arrhythmias | 0 | >180 | 3 | |||
| Device failure | 0 | |||||
| Poor quality sounds | 0 | Low | <40 | 0 | 4 | |
| Cuff size unavailable | 0 | 40–79 | 10 | |||
| Observer disagreement | 2 | DBP | Medium | 80–100 | 12 | 8 |
| Distribution | 6 | High | 101–130 | 9 | 7 | |
| Other reasons | 2 | >130 | 2 | |||
| Total recruited | 33 | |||||
| Yuwell YE680A | ||||||
| Total screened | 44 | SBP | Low | <90 | 0 | 4 |
| Total excluded | 11 | 90–129 | 11 | |||
| Ranges complete | 1 | Medium | 130–160 | 10 | 7 | |
| Range adjustment | 0 | High | 161–180 | 9 | 10 | |
| Arrhythmias | 0 | >180 | 3 | |||
| Device failure | 0 | |||||
| Poor quality sounds | 1 | DBP | Low | <40 | 0 | 6 |
| Cuff size unavailable | 2 | 40–79 | 11 | |||
| Observer disagreement | 3 | Medium | 80–100 | 11 | 5 | |
| Distribution | 2 | High | 101–130 | 11 | 10 | |
| Other reasons | 2 | >130 | 0 | |||
| Total recruited | 33 | |||||
| Cofoe KF-65B | ||||||
| Total screened | 41 | SBP | Low | <90 | 0 | 0 |
| Total excluded | 8 | 90–129 | 12 | |||
| Ranges complete | 1 | Medium | 130–160 | 10 | 8 | |
| Range adjustment | 0 | High | 161–180 | 9 | 10 | |
| Arrhythmias | 1 | >180 | 2 | |||
| Device failure | 0 | |||||
| Poor quality sounds | 1 | DBP | Low | <40 | 0 | 2 |
| Cuff size unavailable | 2 | 40–79 | 10 | |||
| Observer disagreement | 2 | Medium | 80–100 | 12 | 6 | |
| Distribution | 1 | High | 101–130 | 9 | 10 | |
| Other reasons | 0 | >130 | 2 | |||
| Total recruited | 33 | |||||
Subject details for device validation
| Characteristics | Omron HEM-7120 | Yuwell YE680A | Cofoe KF-65B |
|---|---|---|---|
| Sex | |||
| Male:female | 19:14 | 18:15 | 18:15 |
| Age (years) | |||
| Range (low: high) | 30:80 | 26:83 | 25:80 |
| Mean ± SD | 50.0 ± 11.2 | 49.5 ± 13.7 | 48.8 ± 15.0 |
| Arm circumference (cm) | |||
| Range (low:high) | 22:31 | 24:32 | 22:32 |
| Mean (SD) | 27.0 ± 2.2 | 27.6 ± 2.6 | 26.9 ± 2.8 |
| Recruitment SBP (mmHg) | |||
| Range (low:high) | 100:199 | 91:196 | 94:216 |
| Mean (SD) | 144.5 ± 25.0 | 143.6 ± 26.2 | 143.3 ± 31.6 |
| Recruitment DBP (mmHg) | |||
| Range (low:high) | 49:140 | 40:129 | 57:145 |
| Mean (SD) | 92.2 ± 19.6 | 88.3 ± 21.1 | 91.2 ± 21.3 |
| Antihypertensive treatment | |||
|
| 19 | 21 | 18 |
| % | 57.6 | 63.7 | 54.6 |
No statistical significance (P < 0.05) was found among the three device groups. Comparison of binary data was conducted using the chi-square test, and ANOVA was used for continuous data.
Observer measurements in each recruitment range
| BP range | Omron HEM-7120 | Yuwell YE680A | Cofoe KF-65B |
|---|---|---|---|
| SBP (mmHg) | |||
| Overall range (low:high) | 96:191 | 88:200 | 90:203 |
| Low (<130) | 38 | 34 | 33 |
| Medium (130–160) | 39 | 34 | 35 |
| High (>160) | 22 | 31 | 31 |
| Maximum difference | 17 | 3 | 4 |
| DBP (mmHg) | |||
| Overall range (low:high) | 49:132 | 54:129 | 49:124 |
| Low (<80) | 30 | 33 | 39 |
| Medium (80–100) | 38 | 42 | 36 |
| High (>100) | 31 | 24 | 24 |
| Maximum difference | 8 | 18 | 15 |
Validation results for the three devices
| Part 1 | ≤5 mmHg | ≤10 mmHg | ≤15 mmHg | Grade 1 | Mean (mmHg) | SD (mmHg) |
|---|---|---|---|---|---|---|
| Pass required | ||||||
| Two of | 73 | 87 | 96 | |||
| All of | 65 | 81 | 93 | |||
| Omron HEM-7120 achieved | ||||||
| SBP | 73 | 94 | 96 | Pass | 1.3 | 5.3 |
| DBP | 73 | 91 | 96 | Pass | -2.0 | 5.6 |
| Yuwell YE680A achieved | ||||||
| SBP | 50 | 79 | 91 | Fail | 1.0 | 9.0 |
| DBP | 67 | 87 | 95 | Fail | -1.1 | 6.5 |
| Cofoe KF-65B achieved | ||||||
| SBP | 53 | 79 | 91 | Fail | 4.1 | 7.0 |
| DBP | 65 | 89 | 96 | Pass | -3.3 | 5.3 |
| Part 2 | 2/3 ≤5 mmHg | 0/3 ≤5 mmHg | Grade 2 | Grade 3 | ||
| Pass required | ≥24 | ≤3 | ||||
| Omron HEM-7120 achieved | ||||||
| SBP | 27 | 2 | Pass | Pass | ||
| DBP | 24 | 2 | Pass | Pass | ||
| Yuwell YE680A achieved | ||||||
| SBP | 16 | 4 | Fail | Fail | ||
| DBP | 24 | 2 | Pass | Fail | ||
| Cofoe KF-65B achieved | ||||||
| SBP | 16 | 4 | Fail | Fail | ||
| DBP | 24 | 2 | Pass | Pass | ||
| Part 3 | Result | |||||
| Omron HEM-7120 | Pass | |||||
| Yuwell YE680A | Fail | |||||
| Cofoe KF-65B | Fail | |||||
Fig. 1Bland–Altman plots of the differences among three devices’ readings and the observer measurements for SBP.
Fig. 2Bland–Altman plots of the differences among three devices’ readings and the observer measurements for DBP.