Caleb M Yeung1, Jonathan Lans, Joseph B Kuechle, Zachary Wright, Connie Y Chang, Santiago A Lozano-Calderón.
Abstract
INTRODUCTION: Elbow prosthetic reconstruction after distal humeral tumor resection is challenging. We identify the value of the proximal ulna dorsal angulation (PUDA) as an easily-measured radiographic parameter that can help inform ulnar component sizing in the Solar Elbow System (SES) and the Modular Universal Tumor and Revision System (MUTARS), two modular prosthetic systems that are commonly used after tumor resection in this anatomic location. We hypothesized that a larger PUDA measurement would require smaller ulnar stems.Entities:
Year: 2020 PMID: 33970583 PMCID: PMC7434037 DOI: 10.5435/JAAOSGlobal-D-20-00062
Source DB: PubMed Journal: J Am Acad Orthop Surg Glob Res Rev ISSN: 2474-7661
Figure 1A, Measurement of the proximal ulnar dorsal angulation. B, Templated MUTARS total elbow implant. C, Templated SES total elbow implant. D, Template showing a MUTARS total elbow implant which does not fit the elbow given the large PUDA measurement. MUTRS = Modular Universal Tumor and Revision System, PUDA = proximal ulna dorsal angulation, SES = Solar Elbow System
Descriptive Characteristics and the Proximal Ulna Dorsal Angulation Across Age Groups
| Age Group | Average Height (m) | Average Weight (kg) | Average BMI (kg/m2) | Mean PUDA (°) (95% CI) | SD PUDA (°) | Mean PUDA (°) (Q1, Q3) |
| Younger than 40 | 1.70 | 79.38 | 27.23 | 4.41 (3.71-4.57) | 2.06 | 3.97 (1.9, 6.15) |
| 41-50 | 1.67 | 80.74 | 28.32 | 4.80 (4.31-5.29) | 2.32 | 5.10 (3, 6.5) |
| 51-60 | 1.67 | 85.91 | 29.91 | 4.60 (4.10-5.10) | 2.38 | 4.63 (3, 6.3) |
| 61-70 | 1.63 | 77.93 | 28.74 | 4.43 (3.89-4.97) | 2.54 | 4.58 (2.4, 6.2) |
| 71-80 | 1.67 | 81.74 | 29.34 | 5.22 (4.71-5.73) | 2.48 | 5.10 (3.65, 6.9) |
| Older than 80 | 1.62 | 70.04 | 26.52 | 5.17 (4.58-5.75) | 2.85 | 5.23 (3.5, 6.85) |
BMI = body mass index, CI = confidence Interval, PUDA = proximal ulna dorsal angulation
Templating of Ulnar Component Prostheses and Fit of Systems Across Angle Groups
| Prosthetic System | Angle Group, <5° | Angle Group, 5–10° | Angle Group, >10° | PUDA (°), Median (Q1, Q3) | ||
| MUTARS, n (%) | ||||||
| Fit | 75 (97.4) | 97 (91.5) | 35 (92.1) | 0.26[ | 5.4 (1.7, 9.1) | 0.34[ |
| No fit | 2 (2.6) | 9 (8.5) | 3 (7.9) | 5.8 (5, 9.8) | ||
| SES, n (%) | ||||||
| Large | 77 (100) | 106 (100) | 36 (94.7) | 0.029[ | 5.4 (1.8, 8.9) | 0.034[ |
| Other | — | — | 2 (5.3) | 11.7 (10.5, 12.8) |
MUTARS = Modular Universal Tumor and Revision System, PUDA = proximal ulna dorsal angulation, SES = Solar Elbow System
Using the Fisher exact test.
Using the Mann-Whitney U test.